From: Trueness of 12 intraoral scanners in the full-arch implant impression: a comparative in vitro study
 | CS 3600® | CS 3700® | DWIO® | EMERALD® | EMERALD S® | ITERO ELEMENTS 5D® | MEDIT I-500® | OMNICAM® | PRIMESCAN® | RUNEYES® | TRIOS 3® | VIRTUO VIVO® |
---|---|---|---|---|---|---|---|---|---|---|---|---|
CS 3600® |  | 0.08 (0.05) | −0.43 (0.06) | − 0.32 (0.05) | − 0.16 (0.05) | 0.06 (0.05) | 0.05 (0.05) | − 0.34 (0.06) | − 0.02 (0.05) | − 0.09 (0.07) | 0.00 (0.05) | − 0.08 (0.07) |
CS 3700® | 0.9 |  | −0.51 (0.04) | −0.40 (0.04) | − 0.24 (0.04) | −0.01 (0.03) | − 0.03 (0.04) | −0.42 (0.05) | − 0.10 (0.03) | −0.16 (0.06) | − 0.08 (0.03) | − 0.16 (0.06) |
DWIO® | < 0.0001 | < 0.0001 |  | 0.11 (0.04) | 0.27 (0.04) | 0.50 (0.04) | 0.49 (0.04) | 0.09 (0.05) | 0.41 (0.04) | 0.35 (0.06) | 0.43 (0.04) | 0.35 (0.07) |
EMERALD® | < 0.0001 | < 0.0001 | 0.2 |  | 0.16 (0.04) | 0.38 (0.03) | 0.37 (0.04) | −0.02 (0.05) | 0.30 (0.03) | 0.23 (0.06) | 0.32 (0.03) | 0.24 (0.06) |
EMERALD S® | 0.09 | < 0.0001 | < 0.0001 | 0.002 |  | 0.23 (0.03) | 0.22 (0.04) | −0.18 (0.05) | 0.14 (0.03) | 0.08 (0.06) | 0.16 (0.03) | 0.08 (0.06) |
ITERO ELEMENTS 5D® | 1.0 | 1.0 | < 0.0001 | < 0.0001 | < 0.0001 |  | −0.01 (0.03) | −0.40 (0.04) | − 0.09 (0.02) | −0.15 (0.06) | − 0.06 (0.02) | −0.14 (0.06) |
MEDIT I-500® | 1.0 | 1.0 | < 0.0001 | < 0.0001 | < 0.0001 | 1.0 |  | −0.39 (0.05) | −0.08 (0.03) | − 0.14 (0.06) | −0.05 (0.03) | − 0.13 (0.06) |
OMNICAM® | < 0.0001 | < 0.0001 | 0.8 | 1.0 | 0.01 | < 0.0001 | < 0.0001 |  | 0.32 (0.04) | 0.25 (0.07) | 0.34 (0.04) | 0.26 (0.07) |
PRIMESCAN® | 1.0 | 0.08 | < 0.0001 | < 0.0001 | 0.0009 | 0.007 | 0.4 | < 0.0001 |  | −0.06 (0.06) | 0.02 (0.02) | −0.06 (0.06) |
RUNEYES® | 1.0 | 0.2 | < 0.0001 | 0.006 | 1.0 | 0.2 | 0.5 | 0.009 | 1.0 |  | 0.09 (0.06) | 0.01 (0.08) |
TRIOS 3® | 1.0 | 0.4 | < 0.0001 | < 0.0001 | < 0.0001 | 0.2 | 0.9 | < 0.0001 | 1.0 | 0.9 |  | −0.08 (0.06) |
VIRTUO VIVO® | 1.0 | 0.4 | < 0.0001 | 0.01 | 1.0 | 0.4 | 0.6 | 0.02 | 1.0 | 1.0 | 1.0 |  |