From: Trueness of 12 intraoral scanners in the full-arch implant impression: a comparative in vitro study
 | CS 3600® | CS 3700® | DWIO® | EMERALD® | EMERALD S® | ITERO ELEMENTS 5D® | MEDIT I-500® | OMNICAM® | PRIMESCAN® | RUNEYES® | TRIOS 3® | VIRTUO VIVO® |
---|---|---|---|---|---|---|---|---|---|---|---|---|
CS 3600® |  | 0.05 (0.07) | −0.46 (0.09) | −0.33 (0.08) | − 0.18 (0.08) | 0.18 (0.08) | 0.07 (0.08) | −0.28 (0.10) | 0.10 (0.08) | −0.14 (0.09) | 0.08 (0.07) | −0.12 (0.09) |
CS 3700® | 1.0 |  | −0.50 (0.06) | −0.37 (0.05) | − 0.23 (0.05) | 0.13 (0.06) | 0.02 (0.05) | −0.33 (0.08) | 0.05 (0.05) | −0.19 (0.06) | 0.03 (0.04) | −0.16 (0.07) |
DWIO® | < 0.0001 | < 0.0001 |  | 0.13 (0.07) | 0.28 (0.06) | 0.64 (0.07) | 0.53 (0.07) | 0.17 (0.09) | 0.56 (0.06) | 0.31 (0.08) | 0.54 (0.06) | 0.34 (0.08) |
EMERALD® | 0.003 | < 0.0001 | 0.7 |  | 0.15 (0.05) | 0.51 (0.06) | 0.40 (0.06) | 0.04 (0.08) | 0.43 (0.05) | 0.18 (0.07) | 0.41 (0.05) | 0.21 (0.07) |
EMERALD S® | 0.5 | 0.0003 | 0.002 | 0.2 |  | 0.36 (0.06) | 0.25 (0.06) | −0.11 (0.08) | 0.28 (0.05) | 0.04 (0.07) | 0.26 (0.04) | 0.06 (0.07) |
ITERO ELEMENTS 5D® | 0.6 | 0.4 | < 0.0001 | < 0.0001 | < 0.0001 |  | −0.11 (0.07) | −0.47 (0.09) | − 0.08 (0.06) | −0.32 (0.07) | − 0.10 (0.05) | −0.30 (0.08) |
MEDIT I-500® | 1.0 | 1.0 | < 0.0001 | < 0.0001 | 0.002 | 0.9 |  | −0.35 (0.09) | 0.03 (0.06) | −0.21 (0.07) | 0.01 (0.05) | −0.19 (0.07) |
OMNICAM® | 0.2 | 0.003 | 0.8 | 1.0 | 1.0 | < 0.0001 | 0.004 |  | 0.39 (0.08) | 0.14 (0.09) | 0.37 (0.08) | 0.17 (0.09) |
PRIMESCAN® | 1.0 | 1.0 | < 0.0001 | < 0.0001 | < 0.0001 | 1.0 | 1.0 | 0.0005 |  | −0.24 (0.07) | −0.02 (0.05) | −0.22 (0.07) |
RUNEYES® | 0.9 | 0.1 | 0.004 | 0.2 | 1.0 | 0.001 | 0.1 | 0.9 | 0.02 |  | 0.22 (0.06) | 0.03 (0.08) |
TRIOS 3® | 1.0 | 1.0 | < 0.0001 | < 0.0001 | < 0.0001 | 0.8 | 1.0 | 0.0004 | 1.0 | 0.01 |  | −0.20 (0.06) |
VIRTUO VIVO® | 1.0 | 0.4 | 0.002 | 0.1 | 1.0 | 0.009 | 0.3 | 0.8 | 0.09 | 1.0 | 0.10 |  |