From: Trueness of 12 intraoral scanners in the full-arch implant impression: a comparative in vitro study
 | CS 3600® | CS 3700® | DWIO® | EMERALD® | EMERALD S® | ITERO ELEMENTS 5D® | MEDIT I-500® | OMNICAM® | PRIMESCAN® | RUNEYES® | TRIOS 3® | VIRTUO VIVO® |
---|---|---|---|---|---|---|---|---|---|---|---|---|
CS 3600® |  | 55.9 (27.9) | 90.9 (35.7) | 98.4 (4.2) | 226.9 (47.3) | 57.0 (27.2) | 80.5 (36.4) | 18.6 (13.0) | 20.6 (23.2) | −71.6 (18.5) | 154.6 (52.6) | 145.2 (32.6) |
CS 3700® | 0.7 |  | 35.0 (23.4) | 42.6 (29.5) | 171.0 (36.8) | 1.1 (4.6) | 24.6 (9.4) | −37.3 (21.6) | −35.2 (10.2) | − 127.4 (26.9) | 98.8 (25.4) | 89.4 (12.1) |
DWIO® | 0.3 | 0.9 |  | 7.5 (36.8) | 136.0 (58.1) | −33.9 (26.3) | −10.4 (28.0) | −72.3 (30.2) | −70.3 (16.6) | − 162.5 (23.8) | 63.7 (39.2) | 54.3 (34.8) |
EMERALD® | < 0.0001 | 1.0 | 1.0 |  | 128.5 (49.9) | −41.4 (29.3) | −18.0 (38.3) | −79.9 (11.5) | − 77.8 (24.4) | − 170.0 (17.3) | 56.2 (54.4) | 46.8 (34.6) |
EMERALD S® | 0.0001 | 0.0003 | 0.5 | 0.3 |  | −169.9 (32.8) | − 146.4 (33.2) | − 208.3 (48.4) | − 206.3 (45.4) | − 298.5 (58.3) | −72.3 (36.1) | −81.7 (25.0) |
ITERO ELEMENTS 5D® | 0.6 | 1.0 | 1.0 | 1.0 | < 0.0001 |  | 23.5 (9.3) | −38.4 (22.8) | −36.4 (13.0) | −128.5 (29.0) | 97.6 (25.4) | 88.3 (9.1) |
MEDIT I-500® | 0.5 | 0.3 | 1.0 | 1.0 | 0.0008 | 0.3 |  | −61.9 (30.8) | −59.8 (18.8) | − 152.0 (36.0) | 74.2 (16.3) | 64.8 (10.1) |
OMNICAM® | 1.0 | 0.9 | 0.4 | < 0.0001 | 0.001 | 0.9 | 0.7 |  | 2.1 (16.4) | −90.1 (12.9) | 136.1 (46.5) | 126.7 (28.7) |
PRIMESCAN® | 1.0 | 0.03 | 0.002 | 0.07 | 0.0004 | 0.2 | 0.07 | 1.0 |  | −92.2 (17.4) | 134.0 (34.2) | 124.6 (21.8) |
RUNEYES® | 0.007 | 0.0002 | < 0.0001 | < 0.0001 | < 0.0001 | 0.0007 | 0.002 | < 0.0001 | < 0.0001 |  | 226.2 (51.5) | 216.8 (37.0) |
TRIOS 3® | 0.1 | 0.007 | 0.9 | 1.0 | 0.7 | 0.008 | 0.0004 | 0.1 | 0.006 | 0.0008 |  | −9.4 (22.1) |
VIRTUO VIVO® | 0.0006 | < 0.0001 | 0.9 | 1.0 | 0.05 | < 0.0001 | < 0.0001 | 0.0008 | < 0.0001 | < 0.0001 | 1.0 |  |