Table 3 Measured flexural forces before and after sterilization

From: Effects of disinfection and sterilization on the dimensional changes and mechanical properties of 3D printed surgical guides for implant therapy – pilot study

Sterilization type

no. 1

no. 2

Comparison

Strength (N)

Decreasing (%)

Strength (N)

Decreasing (%)

CV

Difference (%)

Control

412.9

0

171.5

0

0.58

141

Disinfection

373.5

10

158.8

7

0.57

135

Plasma sterilized

169.7

59

158.3

8

0.05

7

Autoclave 121 °C

384.3

7

171.3

0

0.54

124

Autoclave 134 °C

232.8

44

195.9

−14

0.12

19

Mean

314.7

 

171.1

 

85.3

SD

106.9

15.2

66.4

CV

0.34

0.09

0.78