From: Effects of disinfection and sterilization on the dimensional changes and mechanical properties of 3D printed surgical guides for implant therapy – pilot study
Sterilisation type
Strength (N)
Decreasing (%)
Control
88.6
0
Disinfection
56.4
36
Plasma sterilized
71.9
19
Autoclave 121 °C
85.3
4
Autoclave 134 °C
101.8
−15
Mean
80.8
SD
17.3
CV
0.21