Skip to main content

Table 2 Photobiomodulation parameters

From: Complementarity of surgical therapy, photobiomodulation, A-PRF and L-PRF for management of medication-related osteonecrosis of the jaw (MRONJ): an animal study

Parameter Value/characteristics
Manufacturer Hamerz, Tehran, Iran
Model identifier 3L-IR
Year produced 2019
Number and type of emitters One Diode laser
Wavelength 808 nm
Pulse mode Continuous wave (CW)
Power 500 mW
Beam spot size at the target 0.28 cm2
Irradiance at target 1785.71 mW/cm2
Exposure duration 120 s
Radiant exposure 5 J/cm2
Radiant energy 1.4 J
Number of points irradiated One
Area irradiated 0.28 cm2
Application technique Noncontact mode 0.5–1 cm distance from the target
Number and frequency of treatment sessions Five sessions on postoperative 1st, 3rd, 5th, 7th, and 10th days
Total radiant energy over the entire treatment course 7 J