Skip to main content

Table 3 Shows area under (ROC) curve (AUC), standard error (SE), confidence Interval (CI) for dehiscence group

From: Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study

 

AUC

SE

95% CI of AUC

HD-CBCT without MAR

1.00

0.00

0.839, 1.00

LD-CBCT without MAR

1.00

0.00

0.839, 1.00

HD-CBCT with MAR

0.818

0.0761

0.591, 0.950

LD-CBCT with MAR

0.682

0.0761

0.445, 0.865

p-value

HD-CBCT without MAR versus LD-CBCT without MAR: 1.00

HD-CBCT without MAR versus HD-CBCT with MAR: 0.02*

HD-CBCT without MAR versus LD-CBCT with MAR: < 0.001*

HD-CBCT with MAR versus LD-CBCT without MAR: 0.02*

HD-CBCT with Mar versus LD-CBCT with MAR: 0.21

LD-CBCT without MAR versus LD-CBCT with MAR: < 0.001*

  1. *Statistically significant at p-value ≤ 0.05