Skip to main content

Table 1 Mean and standard deviation of retention forces between straight and angled screw access channel abutments

From: Retention of zirconia crowns to titanium bases with straight versus angled screw access channels: an invitro study

Groups

Straight screw access channel Ti-base (n = 7)

Angled screw access channel Ti-base (n = 7)

Mean (SD)

1731.57 (63.68) N

1032.29 (89.82) N

Mean difference (95% CI)

-699.29 (-608.62, -789.95)

Effect size (d)

8.982

Test (P value)

16.804 (< 0.0001*)

  1. *Statistically significant at P ≤ .05, CI: Confidence Interval