Retention of zirconia crowns to titanium bases with straight versus angled screw access channels: an invitro study

Abstract

Background

The aim of this study was to assess the impact of abutments with angled screw access channel on the retention of zirconia crowns.

Methods

Seven implant replicas were inserted in epoxy resin blocks. Fourteen zirconia crowns for central incisor tooth were digitally fabricated and cemented to titanium bases (Ti-bases) with resin cement. Titanium bases were categorized into 2 groups (n = 7). Control group (Group STA) included straight screw access channel abutments. Study group (Group ASC) included angled screw access channel abutments. Following aging (5 °C-55 °C, 60 s; 250,000 cycles, 100 N, 1.67 Hz), the pull-off forces (N) were recorded by using retention test (1 mm/min). Types of failure were defined as (Type 1; Adhesive failure when luting agent predominantly remained on the Ti-base surface (> 90%); Type 2; Cohesive failure when luting agent remained on both Ti-base and crown surfaces; and Type 3; Adhesive failure when luting agent predominantly remained on the crown (> 90%). Statistical analysis was conducted by using (IBM SPSS version 28). Normality was checked by using Shapiro Wilk test and Q-Q plots. Independent t-test was then used to analogize the groups.

Results

Mean ± standard deviation of retention force records ranged from 1731.57 (63.68) N (group STA) to 1032.29 (89.82) N (group ASC), and there was a statistically significant variation between the 2 groups (P < .05). Failure modes were Type 2 for group STA and Type 3 for group ASC.

Conclusions

The retention of zirconia crowns to abutments with a straight screw access channel is significantly higher than abutments with angled screw access channel.

Peer Review reports

Background

The management of dental situations with screw-retained restorations has been applicable with the spread of abutment with Angled Screw access Channel (ASC) [1]. In the esthetic zone and tilted anterior maxilla, it may be required to place the crown at an inclination to the implant long axis. This leads to the placement of screw access channel on the labial side of the restoration [2]. However, systems that have been recently presented, can permit correction of unfavourable angulation.

The screw access channel of the ASC might be located anywhere between 0 and 25° from the implant long axis [3, 4] Ball hexagon driver can engage the specially designed abutment screw of the ASC to apply the recommended torque values [4].

The ASC abutment has larger screw access opening and fewer axial walls than the straight abutment as a reason of the greater angulation. Therefore, the surface area between the abutment and the crown is reduced.

Recently, titanium bases (Ti-bases) have become a popular part in the digital workflow of implant supported restorations. One advantage is that the crown and the Ti-base can be extra-orally bonded together and then that one piece restoration can be screw retained to the implant [5]. Therefore, the remaining cement is restrained, and the risk of biological complications is reduced [6] Ti-bases encompassing ASC are now available [7].

Moreover, restorations with angled screw channel can be digitally designed and fabricated through the advancements of the software programs and implant systems [7].

Retention force is directly related to the abutment’s surface area. The geometry of the abutment may be a priority to chemical bonding as it provides more retention [5]. The retentiveness of zirconia can be significantly improved by increasing height of the abutment as well as the number of axial walls as they provide more surface area [8,9,10] It has been stated that, retention of zirconia crowns is directly correlated with Ti-base height, Ti-base blasting or silica coating and chemical bonding [11]. The successful outcome of the Ti-base concept is based on the fact that Ti-base and the ceramic superstructure have a stable bonded interface [12].

Therefore, this in vitro study aimed to assess the impact of abutments with angled screw access channel on the retention force applied on zirconia crowns. The null hypothesis was that the angled screw channel Ti-bases have no effect on the retention of adhesively bonded zirconia crowns.

Methods

Preparation of sample

Seven implant analogues were embedded into epoxy resin blocks. Implant replicas were maintained in position with a guide pin and connected to a laboratory surveyor. The surveyor was used to permit the fixation of the implant replicas into the resin blocks [3]. Implants were positioned at a distance of 3 mm between the cervical portion of the implant and upper surface of the resin.

The sample size was calculated presuming 5% alpha error and 80% study power. Sample size was based on Rosner’s method [13] calculated by G* power 3.1 [14]. The total sample size was determined as (n = 14) samples that were grouped into 2 groups based on the angulation of the screw-access channel. Group STA (control group) included (n = 7) straight screw access channel Ti bases (Fig. 1). Group ASC (study group) included (n = 7) angled screw access channel Ti bases [15] (Fig. 2).

Fig. 1
figure 1

Group STA (control group) included straight screw access channel Ti-bases

Fig. 2
figure 2

Group ASC (study group) included angled screw access channel Ti-bases

Single zirconia crown for central incisor teeth was digitized for each sample. For Group STA, straight screw access channel Ti-base (GM exact titanium base with removable screw, Neodent) of size 4.5 × 6 × 2.5 mm was selected from the library on (Ceramill, Amann Girrbach) CAD software. The screw access hole of the straight Ti-base is located at the incisal edge. The screw access hole of the restoration was designed to be located near the incisal edge (Fig. 3).

Fig. 3
figure 3

Screw access hole of restoration for group STA was located on incisal edge

For Group ASC, angled screw access channel Ti-base (GM titanium base AS, Neodent) of the same dimensions was selected from the library on (Ceramill, Amann Girrbach) CAD software. The screw access channel was situated on the palatal surface with angulation 24 degrees (Fig. 4 A, B).

Fig. 4
figure 4

A, B Screw access hole of restoration for group ASC was located on palatal surface with angulation 24 degrees

The crowns of both groups were designed precisely for easier pull off the retention test.5 The STL files for the digitized crowns were exported to 5-axis milling machine (CORiTEC 150i dry, imes icore) for the milling process of zirconia crowns (Chengdu Besmile Biotechnology Co.,Ltd; yttrium oxide 4.5%).

Polytetrafluoroethylene tape was used to close the screw access hole. Aluminium oxide particles (120 µ) were used to sandblast the inner surface of the crown. The restoration was cleaned in an ultrasonic unit for approximately 1 min. The restoration was then thoroughly washed with water spray and dried with water/air followed by cleaning with Ivoclean. A single coat of primer was applied with a brush to the pre-treated surfaces. The primer was left for 180 s then was dissipated with a strong stream of air. Self-adhesive resin sealant (Multilink speed, Ivoclar Vivadent) was used as a bonding agent to bind zirconia crowns to their corresponding Ti-bases [16]. Spot curing for the margins for one second per quarter surface. After the elimination of excess cement, curing for the entire surface was carried out for 20 s. This procedure was conducted with firm finger pressure for 5 s and a 25 N load which was vertically applied for a period of 10 min [17] (Fig. 5). Subsequently, the excess cement was cleaned, and the one-piece restorations were screwed onto the implants with the torque recommended by the manufacturer (20 N). Retightening was done 10 min after the application of the first torque to compensate for preload loss because of settling effect of screw [18, 19]. Polytetrafluoroethylene tape and a composite filling material were used to close the screw access channels [16].

Fig. 5
figure 5

Load (25 N) was vertically applied for period of 10 min.Caption: Central vertical arm can be only raised for short distance to accommodate height of crown without epoxy resin block

Artificial aging

The samples were exposed to artificial aging through thermal cycling (5 °C to 55 °C, dwell time 60 s) according to ISO TR11405 (1994) and mechanical loading (250,000 cycles, 100 N, 1.67 Hz) in a mastication simulator, which is correspondent to relatively 1 year of masticatory cycles [20, 21]. An antagonist indenter was used in the shape of steatite ball of diameter 4 mm to load a force of 100 N at an angle on 2 mm from the incisal edge of the crowns.

Pull-off retention test

Universal testing machine (5ST, Tinius Olsen) was used to implement the retention testing with a customized device for axially separating crowns from abutment surface at 1 mm/min crosshead speed [5, 17, 22]. The customized device was formed of two parts screwed to each other, having a split ring imbedded in between to support the neck of the crown (Fig. 6 A, B).

Fig. 6
figure 6

(A, B) A. Custom tensile device for axially seperating crowns from abutment surface. B Universal testing machine

A software program (Horizon, version 12.2.4.0.) was used to register the records for the maximum pull-off force in newtons (N), for each sample. After the retention test, the abutment surfaces and the inner surfaces of the restorations were visually inspected with a Stereomicroscope (B016, Olympus, Japan; Software: Toup view, version 3.7) under × 20 magnification for remaining cement, and classification of failure modes (Fig. 7 A-D). Three categories for modes of failure were identified; Type 1, Adhesive failure when luting agent predominantly remained on the Ti-base surface (> 90%); Type 2, cohesive failure when luting agent remained on both Ti-base and crown surfaces; and Type 3, Adhesive failure when luting agent predominantly remained on the crown (> 90%) [23].

Fig. 7
figure 7

(A, B, C, D) Representative Stereomicroscope under × 20 magnification of remaining cement after retention test, (A,B) STA group, (C,D) ASC group

Statistical software (IBM SPSS version 28) was used for the statistical analysis to estimate the differences between the groups. Normality was checked by using Shapiro Wilk test and Q-Q plots. Values were normally distributed; therefore, groups were compared by using independent t-test. The tests were two-tailed, and the level of significance was set at P ≤ 0.05.

Results

Mean value of retention forces in group STA (1731.57 N) was higher than that of group ASC (1032.29 N). Therefore, there is statistically significant difference between pull-off retention forces of straight and angled screw access channel abutments (P < 0.05) as illustrated in Table 1.

Table 1 Mean and standard deviation of retention forces between straight and angled screw access channel abutments

Visual inspection of all samples under × 20 magnification of stereomicroscope revealed Type 2 failure mode for all the samples in group STA along with Type 3 for group ASC.

Discussion

The study revealed that angled screw channel abutment affected the pull-off retention forces on zirconia crown. In fact, the angled abutments caused reduction in the retention of zirconia restorations. Therefore, the null hypothesis was rejected.

The explanation to this result is the fact that, ASC abutment has larger screw access opening and fewer axial walls than the straight abutment as result of the greater angulation. Therefore, the surface area between the abutment and the crown is reduced causing further reduction in retention force.

The retention of prosthetic restorations is affected by many factors such as height of abutment, surface area, preparation geometry, surface roughness, and bonding agent [10, 24,25,26]. This comes in agreement with Tan et al. [9], who reported that extra surface area can be provided by increasing the number of axial walls. In addition, Muller L et al. [27], stated that abutment height is positively correlated to retention of permanent cements luting zirconia crowns to Ti bases and higher retention forces were reported for tall abutments compared to short abutments [8, 10, 11]

Bergamo ET et al. [28], stated that sandblasting the titanium surface can increase surface roughness, which provides an improved micromechanical interlocking by increasing surface area available for bonding, and such surface modifications enhanced the wettability of the bonding system.

In this study, two specimens of the straight abutments showed abutment screw fracture without crown-abutment debonding at 1350 N and 1590 N. One specimen of the straight abutment showed fracture of zirconia crown during retention test at 1810 N.

The retention forces, in this study, were of high values in comparison to other studies [17, 24, 28, 29] but were consistent with others [27, 30]. The reason may include the use of abutments with greater heights, titanium abutments with retentive areas and the type of self-adhesive resin sealant in addition to using primer which incorporate 10-methacryloxydecyl dihydrogen phosphate (MDP) [11, 17].

The forces needed to dislodge the crown were higher in the straight abutment than the angled abutment and the difference was found to be statistically significant. Further invivo studies are needed to identify the clinical significance of the recorded values.

Failure modes that occurred after the pull-off test showed mixed cement deposition on both of Ti-base and crown surfaces (Type 2) for the straight abutment group, indicating cohesive failure. While, in the angled screw access channel abutment, most cement remained on the intaglio of the crowns (Type 3), which identified an adhesive failure between the Ti-base surface and the luting agent. These observations assist the benefit of increasing means of retention to improve the adhesion of angled screw channel abutments to titanium surfaces.

In this study, the methodology was used to mimic the standard clinical situations. Thermocycling and mechanical loading were performed to replicate the oral environment [28]. Thermocycling is an identified form of aging, because hydrolytic degradation, repeated expansion/contraction, temperature changes, as well as other stresses within a sample have a significant impact on tensile strength [30]. Replicating the masticatory cycle is an essential factor in laboratory simulation [20]. Increased number of cycles in cyclic loading may also have an impact on monolithic zirconia crowns [31].

Conclusions

The retention of zirconia crowns to abutments with a straight screw access channels is significantly higher than abutments with angled screw access channel. Angled screw access channel abutment causes reduction in the pull-off retention force of zirconia crown. It is recommended to perform further invivo studies to prove the invitro findings of this study.

Availability of data and materials

The datasets used and analyzed during the current RCT are available from the corresponding author upon reasonable request.

Abbreviations

ASC:

Angled screw access channel abutment

STA:

Straight screw access channel abutment

Ti-bases:

Titanium bases

N:

Newtons

MDP:

Methacryloxydecyl dihydrogen phosphate

References

  1. Friberg B, Ahmadzai M. A prospective study on single tooth reconstructions using parallel walled implants with internal connection (NobelParallel CC) and abutments with angulated screw channels (ASC). Clin Implant Dent Relat Res. 2019;21(2):226–31. https://doi.org/10.1111/cid.12709.

    Article  PubMed  Google Scholar 

  2. Gjelvold B, Sohrabi MM, Chrcanovic BR. Angled Screw Channel: An Alternative to Cemented Single-Implant Restorations-Three Clinical Examples. Int J Prosthodont. 2016;29(1):74–6. https://doi.org/10.11607/ijp.4686.

    Article  PubMed  Google Scholar 

  3. Garcia-Hammaker S, Saglik B, Sierraalta M, Razzoog M. Influence of Screw Channel Angulation on the Fracture Resistance of Zirconia Abutments: An In Vitro Study. J Prosthodont. 2021;30(4):329–34. https://doi.org/10.1111/jopr.13275.

    Article  PubMed  Google Scholar 

  4. Garcia-Gazaui S, Razzoog M, Sierraalta M, Saglik B. Fabrication of a screw-retained restoration avoiding the facial access hole: a clinical report. J Prosthet Dent. 2015;114(5):621–4. https://doi.org/10.1016/j.prosdent.2015.06.007.

    Article  PubMed  Google Scholar 

  5. Pitta J, Burkhardt F, Mekki M, Fehmer V, Mojon P, Sailer I. Effect of airborne-particle abrasion of a titanium base abutment on the stability of the bonded interface and retention forces of crowns after artificial aging. J Prosthet Dent. 2021;126(2):214–21. https://doi.org/10.1016/j.prosdent.2020.06.013.

    Article  PubMed  Google Scholar 

  6. Korsch M, Obst U, Walther W. Cement-associated peri-implantitis: a retrospective clinical observational study of fixed implant-supported restorations using a methacrylate cement. Clin Oral Implants Res. 2014;25(7):797–802. https://doi.org/10.1111/clr.12173.

    Article  PubMed  Google Scholar 

  7. Rasaie V, Abduo J, Falahchai M. Clinical and Laboratory Outcomes of Angled Screw Channel Implant Prostheses: A Systematic Review. Eur J Dent. 2022;16(3):488–99. https://doi.org/10.1055/s-0041-1740298.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Abbo B, Razzoog ME, Vivas J, Sierraalta M. Resistance to dislodgement of zirconia copings cemented onto titanium abutments of different heights. J Prosthet Dent. 2008;99(1):25–9. https://doi.org/10.1016/S0022-3913(08)60005-0.

    Article  PubMed  Google Scholar 

  9. Tan KM, Masri R, Driscoll CF, Limkangwalmongkol P, Romberg E. Effect of axial wall modification on the retention of cement-retained, implant-supported crowns. J Prosthet Dent. 2012;107(2):80–5. https://doi.org/10.1016/S0022-3913(12)60028-6.

    Article  PubMed  Google Scholar 

  10. Rodiger M, Kloss J, Gersdorff N, Burgers R, Rinke S. Removal forces of adhesively and self-adhesively luted implant-supported zirconia copings depend on abutment geometry. J Mech Behav Biomed Mater. 2018;87:119–23. https://doi.org/10.1016/j.jmbbm.2018.07.028.

    Article  PubMed  Google Scholar 

  11. Zahoui A, Bergamo ET, Marun MM, Silva KP, Coelho PG, Bonfante EA. Cementation Protocol for Bonding Zirconia Crowns to Titanium Base CAD/CAM Abutments. Int J Prosthodont. 2020;33(5):527–35. https://doi.org/10.11607/ijp.6696.

    Article  PubMed  Google Scholar 

  12. Wiedenmann F, Liebermann A, Spintzyk S, Eichberger M, Stawarczyk B. Influence of Different Cleaning Procedures on Tensile Bond Strength Between Zirconia Abutment and Titanium Base. Int J Oral Maxillofac Implants. 2019;34(6):1318–27. https://doi.org/10.11607/jomi.7638.

    Article  PubMed  Google Scholar 

  13. Rosner B. Hypothesis Testing: Two-Sample Inference. In: Fundamentals of biostatistics. 7th ed. Boston:Brooks/Cole: Nelson Education. 2015:269–301.

  14. Universitӓt Düsseldorf. G*Power. 2019. Retrieved from http://www.gpower.hhu.de/.

  15. Mulla SH, Seghi RR, Johnston WM, Yilmaz B. Effect of cyclic loading on reverse torque values of angled screw channel systems. J Prosthet Dent. 2022;128(3):458–66. https://doi.org/10.1016/j.prosdent.2020.12.020.

    Article  PubMed  Google Scholar 

  16. Swamidass RS, Kan JYK, Kattadiyil MT, Goodacre CJ, Lozada J. Abutment screw torque changes with straight and angled screw-access channels. J Prosthet Dent. 2021;125(4):675–81. https://doi.org/10.1016/j.prosdent.2020.01.018.

    Article  PubMed  Google Scholar 

  17. Rohr N, Brunner S, Martin S, Fischer J. Influence of cement type and ceramic primer on retention of polymer-infiltrated ceramic crowns to a one-piece zirconia implant. J Prosthet Dent. 2018;119(1):138–45. https://doi.org/10.1016/j.prosdent.2017.02.002.

    Article  PubMed  Google Scholar 

  18. Pardal-Peláez B, Montero J. Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review. J Clin Exp Dent. 2017;9(11):1355–61. https://doi.org/10.4317/jced.54374.

    Article  Google Scholar 

  19. Shin HMHJ, Yun MJ, Jeon YC, Chang BM, Jeong CM. Influence of the implant-abutment connection design and diameter on the screw joint stability. J Adv Prosthodont. 2014;6(2):126–32. https://doi.org/10.4047/jap.2014.6.2.126.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Nawafleh N, Hatamleh M, Elshiyab S, Mack F. Lithium disilicate restorations fatigue testing parameters: a systematic review. J Prosthodont. 2016;25(2):116–26. https://doi.org/10.1111/jopr.12376.

    Article  PubMed  Google Scholar 

  21. Drew A, Randi A, DiPede L, Luke A. Fracture Strength of Implant Screw-Retained All-Ceramic Crowns with the Use of the Angulated Screw Channel: A Pilot Study. Int J Periodontics Restorative Dent. 2020;40(2):245–52. https://doi.org/10.11607/prd.3952.

    Article  PubMed  Google Scholar 

  22. Lopes ACO, Machado CM, Bonjardim LR, Bergamo ETP, Ramalho IS, Witek L, et al. The Effect of CAD/CAM Crown Material and Cement Type on Retention to Implant Abutments. J Prosthodont. 2019;28(2):552–6. https://doi.org/10.1111/jopr.12927.

    Article  Google Scholar 

  23. Palacios RP, Johnson GH, Phillips KM, Raigrodski AJ. Retention of zirconium oxide ceramic crowns with three types of cement. J Prosthet Dent. 2006;96(2):104–14. https://doi.org/10.1016/j.prosdent.2006.06.001.

    Article  PubMed  Google Scholar 

  24. Silva CE, Soares S, Machado CM, Bergamo ET, Coelho PG, et al. Effect of CAD/CAM Abutment Height and Cement Type on the Retention of Zirconia Crowns. Impalnt dentistry. 2018;27(5):582–7. https://doi.org/10.1097/ID.0000000000000811.

    Article  Google Scholar 

  25. de Campos TN, Adachi LK, Miashiro K, Yoshida H, Shinkai RS, et al. Effect of Surface Topography of Implant Abutments on Retention of Cemented Single-Tooth Crowns. Int J Periodontics Restorative Dent. 2010;30(4):409–13.

    PubMed  Google Scholar 

  26. Kappel S, Chepura T, Schmitter M, Rammelsberg P, Rues S. Effects of Cement, Abutment Surface Pretreatment, and Artificial Aging on the Force Required to Detach Cantilever Fixed Dental Prostheses from Dental Implants. Int J Prosthodont. 2017;30(6):545–52. https://doi.org/10.11607/ijp.5281.

    Article  PubMed  Google Scholar 

  27. Muller L, Rauch A, Reissmann DR, Schierz O. Impact of cement type and abutment height on pull-off force of zirconia reinforced lithium silicate crowns on titanium implant stock abutments: an in vitro study. BMC Oral Health. 2021;21(1):592. https://doi.org/10.1186/s12903-021-01958-6.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Bergamo ET, Zahoui A, Ikejiri LL, Marun M, da Silva KP, Coelho PG, et al. Retention of zirconia crowns to Ti-base abutments: effect of luting protocol, abutment treatment and autoclave sterilization. J Prosthodont Res. 2021;65(2):171–5. https://doi.org/10.2186/jpr.JPOR_2019_537.

    Article  Google Scholar 

  29. Turker N, Ozarslan MM, Buyukkaplan US, Basar EK. Effect of Different Surface Treatments Applied to Short Zirconia and Titanium Abutments. Int J Oral Maxillofac Implants. 2020;35(5):948–54. https://doi.org/10.11607/jomi.8224.

    Article  PubMed  Google Scholar 

  30. Santos-Neto OS, Gonçalves LM, Maia-Filho EM, Malheiros AS, Firoozmand LM, Villis P, et al. Retention of cemented zirconia copings on TiBase abutments. Acta Odontol Latinoam. 2021;34(3):214–20. https://doi.org/10.54589/aol.34/3/214.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Nakamura K, Harada A, Kanno T, Inagaki R, Niwano Y, Milleding P, et al. The influence of low-temperature degradation and cyclic loading on the fracture resistance of monolithic zirconia molar crowns. J Mech Behav Biomed Mater. 2015;47:49–56. https://doi.org/10.1016/j.jmbbm.2015.03.007.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors thank department of biomaterials, faculty of dentistry, Alexandria University for contribution of testing and technical work.

Funding

Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).

Author information

Authors and Affiliations

Authors

Contributions

MMK: Conception, Methodology, Interpretation of data, Supervision, Writing. AMA: Conception, Methodology, Interpretation of data, Supervision, review and editing, Project administration. SE: Supervision, Writing—review & editing. All authors revised and approved the final manuscript.

Corresponding author

Correspondence to Eman Assem Ibrahim.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ibrahim, E.A., Khamis, M.M., Ezzelarab, S. et al. Retention of zirconia crowns to titanium bases with straight versus angled screw access channels: an invitro study. BMC Oral Health 23, 458 (2023). https://doi.org/10.1186/s12903-023-03177-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-023-03177-7

Keywords