Skip to main content

Table 1 The values and statistical analyses of three types of initial implant stabilities of various lengths of implants

From: Effects of implant length and 3D bone-to-implant contact on initial stabilities of dental implant: a microcomputed tomography study

Length (mm)

ITV (N.cm)

PTV

ISQ

Median*

IQR

Max

Min

Median*

IQR

Max

Min

Median*

IQR

Max

Min

8.5

32.70a

1.8

34.8

31.2

−3.2 a

0.45

−2.9

−3.85

78.00 a

0.5

80

75

10

39.60b

5.3

43

35.2

−4.2 ab

0.9

−3.4

−4.8

82.00ab

3

83

80

11.5

44.80bc

11.1

52.9

39.3

−4.5ac

0.75

−3.8

−5.05

84.00b

2

85

83

13

56.40c

5

66.7

50.7

−4.6bc

0.35

−4.35

−5

85.00b

1

85.5

83

P

<0.001

   

0.019

   

0.002

   
  1. †Kruskal-Wallis test
  2. * Post hoc pairwise comparisons were conducted by exact Mann-Whitney U-Test with the Bonferroni adjustment; medians with the same letter a,b,c,d are not significantly different at the 0.00833 (0.05/6) level