Table 2 Volumetric measurement of different CBCT voxel sizes with 5 minimum thresholds compared with laser scan

From: Accuracy of in vitro mandibular volumetric measurements from CBCT of different voxel sizes with different segmentation threshold settings

Voxel size (mm)

HU

Mean vol (mm 3 )

Deviation%

P

.125

0

24,436.83

1.91

.004*

 

100

24,007.69

.12

.794

 

200

23,690.75

−1.2

.018*

 

300

23,350.4

−2.62

.000*

 

400

23,014.19

−4.02

.000*

.20

0

24,695.64

2.99

.000*

 

100

24,282.46

1.27

.020*

 

200

23,959.73

−.07

.855

 

300

23,601.82

−1.57

.009*

 

400

23,246.79

−3.05

.000*

.30

0

25,494.24

6.32

.000*

 

100

24,968.95

4.13

.000*

 

200

24,506.46

2.2

.002*

 

300

24,018.74

.17

.757

 

400

23,604.68

−1.56

.018*

.40

0

25,817.22

7.67

.000*

 

100

25,275.56

5.41

.000*

 

200

24,771.32

3.31

.000*

 

300

24,291.91

1.31

.044*

 

400

23,898.76

−.33

.52

Laser

 

23,978.13

  
  1. *P < 0.05