From: Experimental study on the factors affecting torque of beam-type implant torque wrenches
Manufacturers | Target torque value (N cm) | Part of the beam aligned | Bias (%) | Minimum | Maximum | Mean | SD | Coefficient of variation (%) |
---|---|---|---|---|---|---|---|---|
Strauman | 15 | Leading edge | − 18.67 | 11.5 | 12.7 | 12.20 | 0.45 | 3.66 |
Center | − 4.00 | 14.0 | 14.9 | 14.40 | 0.37 | 2.60 | ||
Trailing edge | 14.27 | 16.7 | 17.8 | 17.14 | 0.37 | 2.14 | ||
Nobel Biocare | 15 | Leading edge | − 10.27 | 13.0 | 13.9 | 13.46 | 0.38 | 2.85 |
Center | − 2.80 | 14.3 | 15.0 | 14.58 | 0.24 | 1.65 | ||
Trailing edge | 11.87 | 16.5 | 17.0 | 16.78 | 0.19 | 1.16 | ||
Kyocera | 20 | Leading edge | − 6.00 | 18.2 | 19.2 | 18.80 | 0.34 | 1.81 |
Center | 2.60 | 19.9 | 20.8 | 20.52 | 0.32 | 1.55 | ||
Trailing edge | 14.20 | 22.3 | 23.1 | 22.84 | 0.28 | 1.23 | ||
GC | 10 | Leading edge | − 40.60 | 4.0 | 6.8 | 5.94 | 1.00 | 16.80 |
Center | 2.80 | 9.7 | 10.8 | 10.28 | 0.45 | 4.33 | ||
Trailing edge | 35.00 | 12.8 | 13.9 | 13.50 | 0.39 | 2.85 | ||
KTC | 15 | Leading edge | − 10.67 | 12.2 | 13.8 | 13.40 | 0.61 | 4.55 |
Center | − 5.07 | 13.9 | 14.6 | 14.24 | 0.27 | 1.92 | ||
Trailing edge | 3.07 | 15.2 | 15.6 | 15.46 | 0.15 | 0.97 | ||
Nippon Piston Ring | 20 | Leading edge | − 16.60 | 16.1 | 17.1 | 16.68 | 0.44 | 2.61 |
Center | 0.50 | 19.8 | 20.3 | 20.10 | 0.17 | 0.83 | ||
Trailing edge | 10.90 | 20.9 | 24.0 | 22.18 | 1.19 | 5.37 | ||
THOMMEN | 15 | Leading edge | 3.07 | 15.0 | 15.8 | 15.46 | 0.29 | 1.86 |
Center | 10.53 | 16.0 | 16.8 | 16.58 | 0.31 | 1.88 | ||
Trailing edge | 18.53 | 17.5 | 18.0 | 17.78 | 0.20 | 1.15 | ||
SHOFU | 10 | Leading edge | 1.40 | 9.5 | 11.0 | 10.14 | 0.48 | 4.77 |
Center | 35.80 | 12.9 | 14.0 | 13.58 | 0.39 | 2.85 | ||
Trailing edge | 89.40 | 18.2 | 19.9 | 18.94 | 0.59 | 3.11 |