Skip to main content

Table 3 Comparison of the average stress values induced in the implants and screws between the models OP and TP during group functional and canine guided loading scenarios

From: Stresses induced by one piece and two piece dental implants in All-on-4® implant supported prosthesis under simulated lateral occlusal loading: non linear finite element analysis study

Model component

Loading scenario

Model TP

Model OP

  

Anterior (MPa)

Posterior (MPa)

Anterior (MPa)

Posterior (MPa)

Dental Implant

Scenario1: Group functional occlusion

6.32

9.47

3.47

2.62

 

Scenario2: Canine guidance occlusion

0.85

1.01

0.63

0.53

Screw

Scenario1: Group functional occlusion

4.41

5.84

2.91

2.01

 

Scenario2: Canine guidance occlusion

0.68

1.01

0.54

0.46

  1. MPa Mega Pascal unit