From: Digital versus radiographic accuracy evaluation of guided implant surgery: an in vitro study
Entry Point (mm) | Apex Point (mm) | Angle (°) | |
---|---|---|---|
Mean of difference | -0.032 | 0.006 | 0.139 |
95% limits of agreement | (-0.144, + 0.081) | (-0.135, + 0.147) | (-0.451, + 0.729) |
Maximum tolerated difference | (-0.200, + 0.200) | (-0.200, + 0.200) | (-0.881, + 0.881) |