Table 2 UL1 No attachment or pressure column groups comparison by material type

From: Forces and moments generated during extrusion of a maxillary central incisor with clear aligners: an in vitro study

Force

ATMOS®

TC-85

Zendura FLX®

P-value1

P-value2

P-value3

Mean

Std Dev

Mean

Std Dev

Mean

Std Dev

Fx (N)

10.20

2.93

1.40

0.32

6.00

3.17

 < 0.001

 < 0.001

 < 0.001

Fy (N)

-25.22

3.25

-0.67

0.28

-24.61

4.46

 < 0.001

 < 0.001

 < 0.001

Fz (N)

0.94

16.91

0.41

0.17

-24.09

26.58

 < 0.001

 < 0.001

 < 0.001

Mx (Nmm)

-130.83

211.30

4.68

2.09

-436.64

329.79

 < 0.001

 < 0.001

 < 0.001

My (Nmm)

54.99

110.75

6.98

1.52

-112.44

173.27

 < 0.001

 < 0.001

 < 0.001

Mz (Nmm)

-29.12

62.59

-14.04

2.93

33.60

41.49

 < 0.001

 < 0.001

 < 0.001

  1. P-value1 is a comparison between ATMOS® vs TC-85
  2. P-value2 is a comparison between TC-85 vs Zendura FLX®
  3. P-value3 is a comparison between ATMOS® vs Zendura FLX®