Table 1 Verification assessment results of the OCIT prototype

From: Development of occlusal canting identifying tool: verification, reliability and validation assessment

Verification parameters

n

Mean ± SD (°)

Std. Error

t-test

df

 aP-value

95% Confidence Interval of the Difference

Lower Bound

Upper Bound

Accuracy error of bubble level

3

0.316 ± 0.028

0.016

19.00

2

0.003**

0.245

0.388

Accuracy error of angle between bite plate and protractor

3

0.100 ± 0.050

0.028

3.46

2

0.074

−0.024

0.224

  1. aP-Value * = P ≤ 0.05, ** = P ≤ 0.01, *** = P ≤ 0.001