Skip to main content

Table 3 Validation assessment results of the OCIT

From: Development of occlusal canting identifying tool: verification, reliability and validation assessment

Phantom Head

n

Ref. value (°)

Mean ± SD (°)

Std. Error

t-test

df

aP-value

95% Confidence Interval of the Difference

Lower Bound

Upper Bound

1

40

0.00

0.075 ± 0.266

0.042

1.778

39

0.083

−0.010

0.160

2

40

2.00

1.825 ± 0.635

0.100

−1.740

39

0.090

−0.378

0.028

3

40

4.00

3.875 ± 0.515

0.081

−1.533

39

0.133

−0.289

0.039

4

40

6.00

5.950 ± 0.316

0.050

−1.000

39

0.323

−0.151

0.051

5

40

8.00

7.975 ± 0.276

0.043

−0.572

39

0.570

−0.113

0.063

  1. aP-Value * = P ≤ 0.05, ** = P ≤ 0.01, *** = P ≤ 0.001