Skip to main content

Table 4 The maximum von mises stresses of commercial reconstruction plates and customized reconstruction plates under INC and MOL-L

From: Finite element analysis and clinical application of 3D-printed Ti alloy implant for the reconstruction of mandibular defects

the maximum von mises

stresses(/MPa)

commercial

reconstruction plate

customized reconstruction plates

12 mm

16 mm

20 mm

2.0 mm

2.4 mm

2.8 mm

2.0 mm

2.4 mm

2.8 mm

2.0 mm

2.4 mm

2.8 mm

INC

907.16

1113.8

1358.3

971.48

679.69

802.98

635.71

442.54

761.38

543.55

MOL-L

1006.1

990.15

1193.1

863.7

626.01

710.36

546.21

410.91

696.38

478.48