A prospective clinical study to evaluate the performance of zirconium dioxide dental implants in single-tooth edentulous area: 3-year follow-up

Abstract

Background

Traditionally, dental implants have been made from titanium or titanium alloys. Alternatively, zirconia-based ceramic implants have been developed with similar characteristics of functional strength and osseointegration. Ceramic implants offer advantages in certain settings, e.g. in patients who object to metal dental implants. The aim of this study was to investigate the mid-term (36 months) clinical performance of a ceramic monotype implant in single-tooth edentulous area.

Methods

This was a prospective, open-label, single-arm study in patients requiring implant rehabilitation in single-tooth edentulous area. Ceramic implants (PURE Ceramic Implant, Institut Straumann AG, Basel, Switzerland) with a diameter of 4.1 mm were placed following standard procedure and loaded with provisional and final prostheses after 3 and 6 months, respectively. Implant survival rate and implant success rate were evaluated and crestal bone levels were measured by analysing standardized radiographs during implant surgery and at 6, 12, 24 and 36 months.

Results

Forty-four patients received a study implant, of whom one patient withdrew consent after 3 months. With one implant lost during the first 6 months after surgery, the implant survival rate was 97.7% at 6 months. No further implants were lost over the following 30 months, and 3 patients were lost to follow-up during this time frame. This led to a survival rate of 97.5% at 36 months.

Six months after implant surgery 93.0% of the implants were considered “successful”, increasing to 97.6% at 12 months and remaining at this level at 24 months (95.1%) and 36 months (97.5%).

Bone loss was most pronounced in the first half-year after implant surgery (0.88 ± 0.86 mm). By contrast, between 12 and 36 months the mean bone level remained stable (minimal gain of 0.06 [± 0.60] mm). Hence, the overall bone loss from implant surgery to 36 months was 0.97 (± 0.88) mm.

Conclusions

In the follow-up period ceramic implants can achieve favourable clinical outcomes on a par with titanium implants. For instance, these implants can be recommended for patients who object to metal dental implants. However, longer term studies with different edentulous morphology need to confirm the present data.

Trial registration

Registered on www.clinicaltrials.gov: NCT02163395.

Peer Review reports

Background

For over 40 years, dental implants have been used for the replacement of compromised, lost or missing teeth, following the principles of osseointegration as reported by Brånemark and Adell [1, 2]. While clinical studies have proved the longevity and success of titanium dental implants [3,4,5,6,7,8], the aesthetics of titanium implants can be challenging, especially in the maxillary incisor area when the dark grey colour of the metal becomes visible. Furthermore, there is a risk of hypersensitivity reaction in patients who are allergic to metals or patients might demand metal-free restorations [9].

Ceramic (zirconia [ZrO2]) implants were developed to solve the limitations associated with titanium-based alloys. For instance, the ceramic implant is a good option when it comes to aesthetics, as the shade of zirconia is similar to the colour of a tooth. Like titanium, zirconia also has excellent mechanical, biocompatible and osseointegration properties [10,11,12,13,14,15,16,17] with the potential to serve as a successful implant material with high survival and success rates [18, 19] and only small bone losses [20,21,22]. However, it is difficult to make a consensus regarding the efficacy of ceramic implants due to the variations in ceramic implants and study designs [23].

It has already been shown that certain ceramic implants can achieve clinical outcomes like those of titanium implants after 12 months of follow-up [24]. The present study aims to investigate the safety and clinical performance of this type of ceramic implant concerning survival and success rates, as well as hard and soft tissue parameters after 36 months in the same cohort.

Methods

The study was a prospective 36-month follow-up of an open-label, single-arm study. In the meantime, the study has been extended to 10 years, with planned patient visits at 5 and 10 years after implant surgery. The materials and methods of the study were published previously [24] and are briefly set out below. The study is registered under www.clinicaltrials.gov (study no. NCT02163395).

Patients

Patients were selected according to predefined eligibility criteria. The main inclusion criterion was a single-tooth edentulous area in the mandible (Fédération Dentaire Internationale [FDI] tooth positions 36 to 46) or maxilla (FDI tooth positions 16 to 26), and a natural tooth had to be distally and mesially adjacent to the implant site. The implant positions had to have mainly healed (at least 8 weeks after extraction) and any bone augmentation with autogenous bone had to have been performed at least 3 months prior to implant surgery. Exclusion criteria included different systemic diseases (e.g., uncontrolled diabetes, mucosal disease, untreated periodontitis, gingivitis, or endodontic lesions), smoking more than 10 cigarettes a day, probing pocket depth of ≥4 mm on a tooth immediately adjacent to the implant site, severe bruxing or clenching habits, inadequate oral hygiene and history of local irradiation therapy. Additional secondary exclusion criteria were applied at or after surgery: lack of implant primary stability, inappropriate implant position for prosthetic requirements, major simultaneous augmentation procedures at surgery, and X-ray not showing the implant from first bone contact to apical tip. The complete list of the eligibility criteria can be found in the first publication of the study [24].

Study device

Ceramic implants (PURE Ceramic Implant, Institut Straumann AG, Basel, Switzerland) with a regular diameter (4.1 mm) were placed in this study. These ceramic implants represent monotype one-piece zirconia implants with already integrated abutments into the implant design, avoiding a microgap between implant and abutment. They are made of yttria-stabilized zirconia (Y-TZP) and are available in 4 different lengths and 2 different abutment heights. The implant is based on the features of the Straumann® Tissue Level Standard Plus and Straumann® Bone Level Implants (Institut Straumann AG, Basel, Switzerland) and has a sandblasted, large-grit, acid-etched surface like the SLA surface present in titanium implants. Further details are provided in the first publication of the study [24].

Surgical and restoration procedure

The implant surgeries were performed by KHB, HK, SS and MG at baseline following standard procedures. Sutures were removed 7 to 14 days after implant surgery, and provisional and final prostheses were placed 11 to 13 weeks and 24 to 28 weeks, respectively, after the implants had been placed. No particular prosthetic procedure was defined in the study protocol. Due to the monotype design of the ceramic implant, lithium disilicate glass–ceramic or zirconium dioxide ceramic crowns were inserted in most cases. Permanent glass ionomer luting cement was used to cement the crowns directly onto the implant abutment. Dental floss was used to remove eventual cement remnants.

The patients attended follow-up visits 12, 24 and 36 months after surgery. Additional follow-up visits are planned 5 and 10 years after implant surgeries. Further details are provided in the first publication of the study [24].

Efficacy evaluations

This is a report of secondary outcomes at 24 and 36-months after implant surgery of:

  • Implant survival

  • Implant success

  • Mean (distal and mesial) bone levels

Implant survival

Implants that were still in place 6, 12, 24 and 36 months after implant surgery were considered as surviving implants.

Implant success

Implant success was defined according to these previously described criteria [25]:

  • Absence of pain, foreign body discomfort or dysaesthesia

  • Absence of recurrent peri-implant infection with suppuration

  • Absence of implant mobility

  • Absence of continuous peri-implant radiolucency.

Failed implants were considered as not successful.

Bone levels

Standardized radiographs were taken at baseline (day of implant surgery) as well as 6, 12, 24 and 36 months after implant surgery. Vertical bone level assessment was performed by measuring from the implant shoulder to the first visible bone contact on the implant. Measurements were made at the distal and mesial aspects of each implant, and an average was calculated (Fig. 1). Distortions (changes on the radiographs from the true implant dimensions) were accounted for by normalizing the measurements to the known thread pitch of the implants. All measurements were analysed by an independent specialist.

Fig. 1
figure 1

Illustration of the bone level measurements. (1) Implant shoulder to first implant-to-bone contact. (2) Distance between the threads of the implant

Plaque index, sulcus bleeding index and Oral hygiene instructions

Plaque and sulcus bleeding indices were assessed at the implant site and in the adjacent teeth investigating if there was plaque and bleeding, respectively (yes/no).

All patients were given oral hygiene instructions during each study visit. Further details about efficacy evaluations are provided in the first publication of the study [24].

Complications

Adverse events (AEs) and device deficiencies (DDs) monitoring was performed throughout the study. The patients were asked at every visit if they experienced any complication (like pain, bleeding, etc) since the last visit.

Statistical analysis

Data management was performed using DMSys data management software (Version 5.3), Sigmasoft International, and statistical analysis using IBM SPSS statistical software (Version 21, IBM).

In short, descriptive summary statistics were calculated for all documented parameters, including mean values and standard deviations.

For survival and success rates, patients who were lost to follow-up during the 36-month follow-up period were excluded. At each follow-up visit survival and success rates were calculated by dividing implants still in place and patients meeting all the success criteria, respectively, by the patient population at this follow-up visit. 95% confidence intervals (CIs) were calculated using the modified Wald method.

Mean bone level changes were calculated by subtracting the average bone levels at 6, 12, 24 and 36 months from the average bone level at baseline (day of implant surgery). In other words, positive changes represent bone gain between baseline and 6, 12, 24 and 36 months, respectively, whereas negative changes represent bone loss.

The intention to treat (ITT) population includes all patients who received implant treatment and underwent at least one measurement after surgery, regardless of any major protocol deviations or premature termination. The per protocol (PP) population includes all patients who received an implant and completed the 36-month follow-up period without major protocol deviations. The safety analysis set (SAS) includes all patients who received implant treatment.

Further statistical analysis details are provided in the first publication of the study [24].

Results

Patients

In October 2011, the first patient was enrolled. Four years later (in October 2015), the last patient completed the 36-month study period.

Forty-six patients were screened, of whom 44 (17 males and 27 females) subsequently received the study implant and were analysed according to the ITT data set. Patients were on average 48 ± 14 years old (median: 49 years, range: 18 to 78 years). Most of the implants were placed in the maxilla (40 implants; 90.9%), whereas only 4 implants (9.1%) were placed in the mandible (Fig. 2). Bone augmentation was necessary in 31.8% of the cases (14 out of 44 sites); all of them in the aesthetic zone and at the time of surgery. All were considered as non-major augmentations.

Fig. 2
figure 2

Distribution of implants placed in the study. Number of implants placed at each tooth position. N = 44. N: Total number of implants placed

Of these 44 patients, 5 terminated the study early and were therefore excluded from the PP analysis. They were either lost to follow-up (3 patients; after 6, 12 and 24 months, respectively), withdrew consent (1 patient; after 3 months) or experienced an AE leading to the implant loss (1 patient; after 6 months). A further 8 patients were excluded from the PP analysis due to major protocol deviations. One patient, who terminated the study early, experienced a major protocol deviation, as a provisional crown had not been placed.

In addition, time window deviations were observed for many patients. These were considered “minor protocol deviations”, with no consequences for the data set allocation.

Hence, in total, 13 patients in the ITT population had to be excluded from the PP population, and efficacy analysis according to PP analysis was possible in 31 patients.

Clinical performance evaluations

Implant survival and success

The implant survival and success rates were derived from the ITT data set, with patients who were lost to follow-up being excluded from the analysis. The only implant loss during the 36-month follow-up period occurred before final implant loading. Three patients were lost between 6 and 36 months. These led to a survival rate of 97.7%, with 95% CIs of 86.9% to > 99.9% within the first 6 months after implant surgery and stayed at this level until 36 months (Fig. 3).

Fig. 3
figure 3

Survival and success rate analysis. Scaling of the Y-axis 70% to 100%. ITT population (n = 44). n (0 m) = 44; n (6 m) = 43; n (12 m) = 42; n (24 m) = 41; n (36 m) = 40. Patients lost to follow-up during the 36-month follow-up period were excluded. CI: Confidence interval; ITT: Intention to treat; m: Months; n: Number

During the 36-month follow-up period, a total of 4 implants were considered as “not successful”. One patient lost the implant before month 6. At 6 months, 2 patients experienced a mucositis, one of whom also showed continuous radiolucency around the implants. At 24 months, one patient was considered as “not successful” as he/she experienced a mucositis and also pain, foreign body discomfort or dysesthesia. At 12 and 36 months, the only implant considered as “not successful” was the lost implant. Hence, the success rate is equal to the survival rate at these two visits. Overall, the implant success rate was greater than 90% throughout the 36-month follow-up period and, at 12, 24 and 36 months, greater than 95% (Fig. 3).

Bone levels

Bones levels were derived from the ITT data set. Missing data were not calculated. Over the 36 months 54.1% of implant sites (n = 20) showed bone loss of less than 1.0 mm or bone gain (Fig. 4). Bone loss in general was most pronounced in the first half-year after implant surgery (0.88 ± 0.86 mm). By contrast, between 12 and 36 months the mean bone level remained stable (minimal gain of 0.06 [± 0.60] mm). Hence, the overall bone loss between implant surgery and 36 months was 0.97 (± 0.88) mm. The bone levels at 6, 12, 24 and 36 months were significantly different from baseline at implant surgery (Fig. 5). The ranges of bone losses were 0.1 to 4.4 mm, − 0.2 to 4.6 mm, − 0.2 to 4.7 mm and − 0.8 to 3.0 mm 6, 12, 24 and 36 after the implant surgery, respectively. Note, negative bone losses representing bone gains.

Fig. 4
figure 4

Categorized bone level changes. ITT population (n = 44). n (0–6 m) = 39; n (0–12 m) = 39; n (0–24 m) = 38; n (0–36 m) = 37. Missing values were excluded from the analysis. ITT: Intention to treat; m: Months; n: Number

Fig. 5
figure 5

Bone levels (mean ± SD) from implant surgery to 36-month follow up visit. ITT population (n = 44). n (0–6 m) = 39; n (0–12 m) = 39; n (0–24 m) = 38; n (0–36 m) = 37. Missing values were excluded from the analysis. ITT: Intention to treat; m: Months; n: Number; SD: Standard deviation

Plaque index

Plaque was observed in 17 (38.6%) patients during the study, mostly at screening (10 occurrences [58.8%]), at month 36 (8 occurrences [47.1%]) and at month 12 (7 occurrences [41.2%]). Until month 24 plaque occurrence was only recorded for the adjacent tooth or teeth, whereas at month 36, 3 of the 8 occurrences were reported for the implant site (Fig. 6).

Fig. 6
figure 6

Number of patients with plaque and sulcus bleeding. ITT population (n = 44). ITT: Intention to treat; n: Number

Sulcus bleeding index

Bleeding was detected in 15 (34.1%) patients during the study, mostly at month 36 (8 occurrences [53.3%]), at month 12 (7 occurrences [46.7%]) and at month 24 (6 occurrences [40.0%]). Bleeding occurred either at the implant site, the adjacent teeth or at both locations (Fig. 6).

Complications

In total, the following 27 (39.7%) AEs out of the 68 AEs that occurred over the 36-month follow-up period were deemed to be related to the study device or study treatment: pain (n = 18), bleeding, swelling, inflammation/infection (n = 6) and unfulfilled implant success criteria (n = 3). None (0%) of the 7 SAEs were deemed to be related to the study device or study treatment.

Discussion

The prospective, open-label, single-arm study investigated ceramic implants for replacing a single missing tooth. The implant survival and success rates 36 months after implant surgery were both 97.5%. The mean bone loss was most pronounced in the first half-year after implant surgery (0.88 mm) and remained relatively stable between 12 and 36 months, which resulted in an overall bone loss of 0.97 mm at 36 months.

The survival and success rates of 97.5% 36 months after implant surgery described here are comparable with the published literature on ceramic and titanium implants [26, 27], confirming the good clinical performance of ceramic implant. This within the limits of this study, as a single-arm study design was selected for this study.

In a recent systematic review that included 14 studies using zirconia implants, an overall survival rate of 92% (95% CIs: 87% to 95%) after one year of functional loading was calculated [26]. Only 4 of these 14 studies examined a monotype ceramic implant system separately, and these had an average observational period of at least 2 years. The survival rate in these 4 studies after an average of 2 to 5.9 years was between 77.3 and 100% [22, 28, 29]. The authors state that the heterogeneity of the selected studies may be behind the differences in survival rates. On the one hand the examined zirconia implants varied considerably in implant design and surface characteristics while, on the other, the studies differed in their surgical protocols and prosthetic superstructures [26].

In a systematic review that analysed 46 studies with a mean follow-up of at least 5 years in function after the insertion of titanium implants, the overall survival rate of titanium implants supporting single crowns was 97.2% (95% CIs: 96.3% to 97.9%) after 5 years [27], which is comparable to the survival rate evaluated in the current study. This strengthens the results of several animal studies showing that zirconia implants undergo osseointegration similar to [10, 14, 15, 30,31,32,33,34,35,36], or even better than [16], titanium implants, and confirms the biocompatibility of zirconia as a dental implant material [37]. Furthermore, various in vivo and in vitro studies investigating soft tissue response around zirconia discovered similar [38] or even superior healing response, less inflammatory infiltrate and reduced plaque adhesion on zirconium oxide discs compared to pure titanium [39,40,41,42]. Excellent gingival conditions around monotype ceramic implants were described by Kniha [43]. In addition, like titanium, zirconia is known to have excellent mechanical properties [11, 13, 17].

In contrast to other ceramic implants, the body of the study device features the ZLA surface, a sandblasted, large-grit, acid-etched surface like the SLA surface of titanium implants [24]. Gahlert compared the removal torque of zirconia implants with either machined or sandblasted surfaces placed in mini-pig jaws [44]. The results revealed that a sandblasted surface can display greater stability in bone than machined surfaces, leading to the conclusion that roughening the zirconia implants improves bone anchoring. Oliva compared zirconia implants with 3 different surfaces in a large study: uncoated, coated and acid-etched. Each group consisted of at least 240 implants [28]. The uncoated implants were roughened by mechanical grinding, while the coated implants were roughened, given a bioactive ceramic coating and subsequently sintered. The survival rate for the acid-etched implants was highest, at 97.6% (n: 333, mean: 3.1 years, range: 1 to 4 years), followed by the coated implants, at 93.6% (n: 249, mean: 3.6 years, range: 2 to 5 years) and the uncoated implants, at 92.8% (n: 249, mean: 3.5 years, range: 2 to 5 years).

In the current study bone loss was 0.97 (± 0.88) mm from implant surgery to month 36, which was similar to [22], or slightly less than [29, 45], results reported in other studies with zirconia implants. The first European Workshop on Periodontology [46] considered a bone loss of less than 1.9 mm after 36 months as “successful.” Hence, the bone loss in the current study may be considered as such, all the more so as most of the implants were placed in the maxilla (90.9%) [24]. Marginal bone loss during the first year of loading has been shown to be more pronounced for implants inserted in the maxilla than for those placed in the mandible [29, 45]. The reported bone loss in two studies on zirconia implants (1.63 mm overall and 1.21 mm for single implants after 48 months, and 1.29 mm after 24 months, respectively) [45] was slightly greater than that observed in the current study. As in the study by Borgonovo [45], in this study also bone loss occurred mainly during the first year and stabilized thereafter. Borgonovo assumed that the relatively large bone loss in the first year could be related to bone maturing after surgery and adapting to withstand functional forces [45]. Furthermore, simultaneous augmentation procedures might result in bone loss during the first year of this study – this might be an explanation for bone losses greater than 3 mm after 6, 12 and 24 months at two implants. In this study, the bone loss of 1.02 (± 0.90) mm during the first 12 months after surgery was below the limit of 1.5 mm and considered to be “successful” [46]. As in the study by Borgonovo [45], even a small bone gain was also observed in this study during the follow-up period. In contrast to the previously reported results, in this study bone gain occurred at a later stage (between month 24 and month 36). Borgonovo assumed that bone gain occurred because of new bone trabeculae forming as a result of bone maturation [45].

Most of the AEs that occurred during the 36 months of this study and that were considered to be “related” to the study treatment were experienced during surgery (visit 2), within one week of surgery, or shortly after/in relation to another study intervention. Hence, these AEs can be considered as expected complications following the insertion of dental implants, confirming that there is no safety issue associated with the placement of the ceramic implants.

Conclusion

The results demonstrate that ceramic implants achieve favourable clinical outcomes in the follow-up; so are survival and success rates as well as bone losses comparable to titanium and other ceramic implants. These implants offer a reliable and successful treatment alternative, especially useful for patients who object to metal and request metal-free implants. However, longer term studies with different edentulous morphology need to confirm the present data.

Abbreviations

AE:

Adverse event

CI:

Confidence interval

DD:

Device deficiency

FDI:

Fédération Dentaire Internationale

ITT:

Intention to treat

m:

Months

n:

Number

PP:

Per protocol

SAS:

Safety analysis set

SD:

Standard deviation

Y-TZP:

Yttria-stabilized zirconia

ZrO2 :

Zirconia

References

  1. Branemark PI, Hansson BO, Adell R, Breine U, Lindstrom J, Hallen O, et al. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg Suppl. 1977;16:1–132.

    PubMed  Google Scholar 

  2. Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg. 1981;10:387–416.

    Article  Google Scholar 

  3. Jemt T. Single implants in the anterior maxilla after 15 years of follow-up: comparison with central implants in the edentulous maxilla. Int J Prosthodont. 2008;21:400–8.

    PubMed  Google Scholar 

  4. Mijiritsky E, Lorean A, Mazor Z, Levin L. Implant tooth-supported removable partial denture with at least 15-year long-term follow-up. Clin Implant Dent Relat Res. 2015;17:917–22.

    Article  Google Scholar 

  5. Astrand P, Ahlqvist J, Gunne J, Nilson H: Implant treatment of patients with edentulous jaws: a 20-year follow-up. Clin Implant Dent Relat Res 2008, 10: 207–217.

  6. Krebs M, Schmenger K, Neumann K, Weigl P, Moser W, Nentwig GH. Long-term evaluation of ANKYLOS(R) dental implants, part i: 20-year life table analysis of a longitudinal study of more than 12,500 implants. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e275–86.

    Article  Google Scholar 

  7. Dhima M, Paulusova V, Lohse C, Salinas TJ, Carr AB. Practice-based evidence from 29-year outcome analysis of management of the edentulous jaw using osseointegrated dental implants. J Prosthodont. 2014;23:173–81.

    Article  Google Scholar 

  8. Chappuis V, Buser R, Bragger U, Bornstein MM, Salvi GE, Buser D. Long-term outcomes of dental implants with a titanium plasma-sprayed surface: a 20-year prospective case series study in partially edentulous patients. Clin Implant Dent Relat Res. 2013;15:780–90.

    Article  Google Scholar 

  9. Sicilia A, Cuesta S, Coma G, Arregui I, Guisasola C, Ruiz E, et al. Titanium allergy in dental implant patients: a clinical study on 1500 consecutive patients. Clin Oral Implants Res. 2008;19:823–35.

    Article  Google Scholar 

  10. Depprich R, Zipprich H, Ommerborn M, Mahn E, Lammers L, Handschel J, et al. Osseointegration of zirconia implants: an SEM observation of the bone-implant interface. Head Face Med. 2008;4:25.

    Article  Google Scholar 

  11. Luthardt RG, Holzhuter M, Sandkuhl O, Herold V, Schnapp JD, Kuhlisch E, et al. Reliability and properties of ground Y-TZP-zirconia ceramics. J Dent Res. 2002;81:487–91.

    Article  Google Scholar 

  12. Moller B, Terheyden H, Acil Y, Purcz NM, Hertrampf K, Tabakov A, et al. A comparison of biocompatibility and osseointegration of ceramic and titanium implants: an in vivo and in vitro study. Int J Oral Maxillofac Surg. 2012;41:638–45.

    Article  Google Scholar 

  13. Ozkurt Z, Kazazoglu E. Zirconia dental implants: a literature review. J Oral Implantol. 2011;37:367–76.

    Article  Google Scholar 

  14. Scarano A, Di CF, Quaranta M, Piattelli A. Bone response to zirconia ceramic implants: an experimental study in rabbits. J Oral Implantol. 2003;29:8–12.

    Article  Google Scholar 

  15. Kohal RJ, Weng D, Bachle M, Strub JR. Loaded custom-made zirconia and titanium implants show similar osseointegration: an animal experiment. J Periodontol. 2004;75:1262–8.

    Article  Google Scholar 

  16. Schultze-Mosgau S, Schliephake H, Radespiel-Troger M, Neukam FW. Osseointegration of endodontic endosseous cones: zirconium oxide vs titanium. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000;89:91–8.

    Article  Google Scholar 

  17. Zhang Y, Pajares A, Lawn BR. Fatigue and damage tolerance of Y-TZP ceramics in layered biomechanical systems. J Biomed Mater Res B Appl Biomater. 2004;71:166–71.

    PubMed  Google Scholar 

  18. Kohal RJ, Knauf M, Larsson B, Sahlin H, Butz F. One-piece zirconia oral implants: one-year results from a prospective cohort study. 1. Single tooth replacement. J Clin Periodontol. 2012;39:590–7.

    Article  Google Scholar 

  19. Andreiotelli M, Wenz HJ, Kohal RJ. Are ceramic implants a viable alternative to titanium implants? A systematic literature review. Clin Oral Implants Res. 2009;20(Suppl 4):32–47.

    Article  Google Scholar 

  20. Borgonovo AE, Fabbri A, Vavassori V, Censi R, Maiorana C. Multiple teeth replacement with endosseous one-piece yttrium-stabilized zirconia dental implants. Med Oral Patol Oral Cir Bucal. 2012;17:e981–7.

    Article  Google Scholar 

  21. Cannizzaro G, Torchio C, Felice P, Leone M, Esposito M. Immediate occlusal versus non-occlusal loading of single zirconia implants. A multicentre pragmatic randomised clinical trial. Eur J Oral Implantol. 2010;3:111–20.

    PubMed  Google Scholar 

  22. Roehling S, Woelfler H, Hicklin S, Kniha H, Gahlert M. A retrospective clinical study with regard to survival and success rates of zirconia implants up to and after 7 years of loading. Clin Implant Dent Relat Res. 2016;18:545–58.

    Article  Google Scholar 

  23. Vohra F, Al-Kheraif AA, Ab Ghani SM, Abu Hassan MI, Alnassar T, Javed F. Crestal bone loss and periimplant inflammatory parameters around zirconia implants: a systematic review. J Prosthet Dent. 2015;114:351–7.

    Article  Google Scholar 

  24. Gahlert M, Kniha H, Weingart D, Schild S, Gellrich NC, Bormann KH. A prospective clinical study to evaluate the performance of zirconium dioxide dental implants in single-tooth gaps. Clin Oral Implants Res. 2016;27:e176–84.

    Article  Google Scholar 

  25. Buser D, Weber HP, Lang NP. Tissue integration of non-submerged implants. 1-year results of a prospective study with 100 ITI hollow-cylinder and hollow-screw implants. Clin Oral Implants Res. 1990;1:33–40.

    Article  Google Scholar 

  26. Hashim D, Cionca N, Courvoisier DS, Mombelli A. A systematic review of the clinical survival of zirconia implants. Clin Oral Investig. 2016;20:1403–17.

    Article  Google Scholar 

  27. Jung RE, Zembic A, Pjetursson BE, Zwahlen M, Thoma DS. Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years. Clin Oral Implants Res. 2012;23(Suppl 6):2–21.

    Article  Google Scholar 

  28. Oliva J, Oliva X, Oliva JD. Five-year success rate of 831 consecutively placed zirconia dental implants in humans: a comparison of three different rough surfaces. Int J Oral Maxillofac Implants. 2010;25:336–44.

    PubMed  Google Scholar 

  29. Payer M, Arnetzl V, Kirmeier R, Koller M, Arnetzl G, Jakse N. Immediate provisional restoration of single-piece zirconia implants: a prospective case series - results after 24 months of clinical function. Clin Oral Implants Res. 2013;24:569–75.

    Article  Google Scholar 

  30. Dubruille JH, Viguier E, Le NG, Dubruille MT, Auriol M, Le CY. Evaluation of combinations of titanium, zirconia, and alumina implants with 2 bone fillers in the dog. Int J Oral Maxillofac Implants. 1999;14:271–7.

    PubMed  Google Scholar 

  31. Hoffmann O, Angelov N, Gallez F, Jung RE, Weber FE. The zirconia implant-bone interface: a preliminary histologic evaluation in rabbits. Int J Oral Maxillofac Implants. 2008;23:691–5.

    PubMed  Google Scholar 

  32. Gahlert M, Roehling S, Sprecher CM, Kniha H, Milz S, Bormann K. In vivo performance of zirconia and titanium implants: a histomorphometric study in mini pig maxillae. Clin Oral Implants Res. 2012;23:281–6.

    Article  Google Scholar 

  33. Wenz HJ, Bartsch J, Wolfart S, Kern M. Osseointegration and clinical success of zirconia dental implants: a systematic review. Int J Prosthodont. 2008;21:27–36.

    PubMed  Google Scholar 

  34. Kohal RJ, Wolkewitz M, Hinze M, Han JS, Bachle M, Butz F. Biomechanical and histological behavior of zirconia implants: an experiment in the rat. Clin Oral Implants Res. 2009;20:333–9.

    Article  Google Scholar 

  35. Assal PA. The osseointegration of zirconia dental implants. Schweiz Monatsschr Zahnmed. 2013;123:644–54.

    PubMed  Google Scholar 

  36. Ichikawa Y, Akagawa Y, Nikai H, Tsuru H. Tissue compatibility and stability of a new zirconia ceramic in vivo. J Prosthet Dent. 1992;68:322–6.

    Article  Google Scholar 

  37. Akagawa Y, Ichikawa Y, Nikai H, Tsuru H. Interface histology of unloaded and early loaded partially stabilized zirconia endosseous implant in initial bone healing. J Prosthet Dent. 1993;69:599–604.

    Article  Google Scholar 

  38. Pae A, Lee H, Kim HS, Kwon YD, Woo YH. Attachment and growth behaviour of human gingival fibroblasts on titanium and zirconia ceramic surfaces. Biomed Mater. 2009;4:025005.

    Article  Google Scholar 

  39. Rimondini L, Cerroni L, Carrassi A, Torricelli P. Bacterial colonization of zirconia ceramic surfaces: an in vitro and in vivo study. Int J Oral Maxillofac Implants. 2002;17:793–8.

    PubMed  Google Scholar 

  40. Scarano A, Piattelli M, Caputi S, Favero GA, Piattelli A. Bacterial adhesion on commercially pure titanium and zirconium oxide disks: an in vivo human study. J Periodontol. 2004;75:292–6.

    Article  Google Scholar 

  41. Degidi M, Artese L, Scarano A, Perrotti V, Gehrke P, Piattelli A. Inflammatory infiltrate, microvessel density, nitric oxide synthase expression, vascular endothelial growth factor expression, and proliferative activity in peri-implant soft tissues around titanium and zirconium oxide healing caps. J Periodontol. 2006;77:73–80.

    Article  Google Scholar 

  42. Grossner-Schreiber B, Herzog M, Hedderich J, Duck A, Hannig M, Griepentrog M. Focal adhesion contact formation by fibroblasts cultured on surface-modified dental implants: an in vitro study. Clin Oral Implants Res. 2006;17:736–45.

    Article  Google Scholar 

  43. Kniha K, Gahlert M, Hicklin S, Bragger U, Kniha H, Milz S. Evaluation of hard and soft tissue dimensions around zirconium oxide implant-supported crowns: a 1-year retrospective study. J Periodontol. 2016;87:511–8.

    Article  Google Scholar 

  44. Gahlert M, Gudehus T, Eichhorn S, Steinhauser E, Kniha H, Erhardt W. Biomechanical and histomorphometric comparison between zirconia implants with varying surface textures and a titanium implant in the maxilla of miniature pigs. Clin Oral Implants Res. 2007;18:662–8.

    Article  Google Scholar 

  45. Borgonovo AE, Censi R, Vavassori V, Dolci M, Calvo-Guirado JL, Delgado Ruiz RA, et al. Evaluation of the success criteria for zirconia dental implants: a four-year clinical and radiological study. Int J Dent. 2013;2013:463073.

    Article  Google Scholar 

  46. Albrektsson T., Isidor F.: Consensus report of session IV. In Proceedings of the First European Workshop on Periodontology. Edited by Lang N.P., Karring T. Quintessence; 1994:365–369.

Download references

Acknowledgments

Not applicable.

Funding

This study was funded by Institut Straumann AG, Basel, Switzerland.

Availability of data and materials

The datasets generated and/or analysed during this study are not publicly available but may be obtained from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

KHB: carried out surgical procedures, data acquisition and data interpretation (advisor), and took part in drafting the manuscript and critically revised it for important intellectual content. HK: carried out surgical procedures, data acquisition and data interpretation, and revised the manuscript critically for important intellectual content. SS: carried out surgical procedures, data acquisition and data interpretation, and revised the manuscript critically for important intellectual content. MG: carried out surgical procedures, data acquisition and data interpretation, and revised the manuscript critically for important intellectual content. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Kai-Hendrik Bormann.

Ethics declarations

Ethics approval and consent to participate

The study was conducted in accordance with the Declaration of Helsinki (Seoul revision 2008), ISO 14155, all local legal and regulatory requirements, and the study protocol, and it was approved by the independent ethics committees of the coordinating investigator and all study sites, namely Freiburg Ethics Commission International (ref: 011/1616), Landesärztekammer Baden-Württemberg (ref: F-2O11–060-z) and the Ethics Committee of the Hannover Medical School (ref: 6015). All participating patients gave their written informed consent.

Consent for publication

Not applicable.

Competing interests

All authors except DW are involved in training and education activities for Institut Straumann AG, Basel, Switzerland. The authors declare that they have no further competing interests. Institut Straumann AG manufactures the implants. Therefore, Institut Straumann AG gains financially, if the PURE Ceramic Implants would be more widely used.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bormann, KH., Gellrich, NC., Kniha, H. et al. A prospective clinical study to evaluate the performance of zirconium dioxide dental implants in single-tooth edentulous area: 3-year follow-up. BMC Oral Health 18, 181 (2018). https://doi.org/10.1186/s12903-018-0636-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-018-0636-x

Keywords