Effects of laser treatments on surface roughness of zirconium oxide ceramics

Abstract

Background

The aim of this study was to analyze the surface roughness of yttrium stabilized tetragonal polycrystalline zirconia (Y-TZP) ceramics after different laser treatments (CO2, ER: YAG).

Methods

5x5x2 mm rectangular prisms of forty eight Y-TZP (Zirkonzahn) ceramic specimens were prepared. In order to standardize surfaces, 600-, 1200- grid silicon carbide papers were used to gradually ground wet on 300 rpm grinding machine for 10 s. Eight groups (n = 6) were randomly formed from the specimens of each ceramic as control (GroupC), sandblasted (GroupS), two different CO2 laser treatments (Group3W: 3 W and 382 w/cal, Group4W: 4 W and 509w/cal) and four different Er: YAG laser treatments (Group150SP: 150 mJ and 10-Hz with 100μS; Group150SSP: 150 mJ and 10-Hz with 300μS; Group300SP: 300 mJ and 10-Hz with 100μS; Group300SSP: 300 mJ and 10-Hz with 300μS). A profilometer was used to conduct surface roughness measurements (Ra). Surface morphologies of the specimens were evaluated under SEM after laser treatment.

Results

To analyze the data one-way ANOVA and to compare the mean values Tukey HSD tests (α = .05) were used. One - way ANOVA results showed that Group S had the highest Ra value and Group150 SP had the lowest. After sandblasting group the highest value was seen in Group4W. There were no statistically significant differences among Group C, Group3W, Group150SSP, Group300SP, and Group300SSP.

Conclusions

The study showed that surface roughness of zirconium oxide ceramics was increased with CO2 laser.

Peer Review reports

Background

Recently, one of the most frequently used all ceramic core material for fixed restorations (crown and bridge), orthodontic brackets and CAD/CAM technology is yttrium-stabilized- tetragonal-zirconia-polycrystal (Y-TZP) [1]. It has many advantages like high aesthetic profile, biocompatibility, chemical stability and exceptional mechanical features including hardness, 700–1200 MPa high flexural strength and 7–10 MPa m ½ fracture toughness [2, 3].

A strong and solid bond between cement and zirconia is extremely important for patients’ satisfaction. Marginal seal, proper retention and sufficient aesthetics are improved characteristics of resin cements over conventional cements [4, 5]. However, for sufficient bonding one of the vital components is micromechanical attachment [5,6,7]. Roughening zirconia restorations inner surfaces causes increases in the area convenient for penetration and in situ polymerization of resin based materials, which in turn enhance the mechanical bond.

There are many surface treatment methods to improve a succesful bonding. Hydrofluoric acid etching, which is one of the most effective methods to increase the bonding mechanism, is not a useful technique due to the fact that the zirconia is not glassy and is densely-sintered. Studies to ensure a good bonding in zirconia have been shown for many years that the surface should be cleaned first and then roughened. Then chemical activations such as airborne particle abrasion using pure alumina or silica coating using silica-coated alumina particles can be carried out. Due to the improvement of lasers in dentistry laser irradiation is thought to be an alternative method to increase surface roughness and improve adhesion between ceramics and resin cements [8,9,10,11,12,13,14,15].

Recently, for different dentistry practices including cavity preparation [16], carious dentin removal, surface conditioning [17,18,19] and also as a surface treatment of indirect restorations [20], erbium:yttrium aluminum garnet (Er: YAG) laser is recommended. Because of the synchronization of its wavelength and the main absorption peak of water and because of its good absorption by OH groups in hydroxyapatite, it is frequently used on dental ceramics [8]. The carbon dioxide laser (CO2) is commonly used intraorally especially in soft tissue and hard tissue applications [21, 22]. Because ceramic nearly totally absorbs CO2 laser wavelength, CO2 laser is very suitable for the surface treatment of ceramic materials [23]. CO2 laser etching may represent an effective method for conditioning zirconia surfaces, enhancing micromechanical retention and improving the bond strength [24]. Conchoidal tears, which result from surface warming, occur with the heat initiation of surfaces of ceramic by focusing CO2 laser. These tears are believed to supply mechanical success between resin composite and ceramics retention [25,26,27].

In order to prevent damage to the zirconia surface, laser settings like pulse, power and duration have great importance. In order to strengthen bonding and durability of the restorations, all the above mentioned procedures stress increasing the surface area [24]. Despite the fact that various surface treatments have been proposed, it is still a challenge to choose the most appropriate method. The purpose of this study was to examine the differences and similarities of surface roughness of untreated and sandblasted and laser applicated surfaces of zirconia. The research hypothesis was that different energy values of Er: YAG and CO2 laser treatments affect surface roughness of the zirconia.

Methods

In the study 5x5x2 mm of rectangular prism of forty eight zirconia core specimens were produced by a copy-milling system (Zirconzahn, Bruneck, Italy) using prefabricated blanks of zirconia (ICE Zircon Translucent; Zirconzahn) and then sintered according to the manufacturer’s instructions. The specimens’surfaces were firstly cleaned with ethanol and then air-dried before surface treatment (Branson 2210; Branson Ultrasonics Corporation, Danbury, CT). Zirconia cores were embedded in the centers of autopolymerizing acrylic resin blocks (Meliodent; Heraeus Kulzer, Armonk, NY). In order to standardize surfaces, 600-, 1200- grid silicon carbide papers (English abrasives, English abrasives Ltd. England) were gradually grounded using water coolant on a 300 rpm grinding machine for 10 s (Beuhler Metaserv, Germanyand ultrasonically cleaned for 3 min in ethanol and deionized water and then air-dried. Subsequently, specimens were randomly divided into eight groups, each containing 6 specimens, for the following different surface treatment methods (Fig. 1), (Table 1).

  • GroupC: Control group, no treatment.

  • GroupS: Sandblasted specimens with 50-μm aluminium oxide powder (Korox 50, Bego, Bremen, Germany) at 2.8 bars from a 10 mm distance for 15 s. After sandblasting, in order to remove the remaining powder, compressed air was used to clean the specimens.

Fig. 1
figure 1

Forty-eight yitriyum stabilized tetragonal polycryistalin zirconia ceramic specimens were prepared

Table 1 Groups and surface treatments

CO2 laser treatments

The specimens were treated by using CO2 laser (Smart US-20D, DEKA, Firenze, Italy) working at 10.6 μm. 3 W- 4 W energy level was applied at a continuous and non-contact mode. The application tip’s diameter was 1 mm and its length was 12 mm. Moving up and down, zirconia surfaces were processed with the application tip in slight contact [28, 29] (Fig. 2). The applied energy levels were:

  • Group 3W: The applied energy level was 382w/cal and 3W

  • Group 4W: The applied energy level was 509w/cal and 4W

Fig. 2
figure 2

The application tip was moved from bottom to top and maintained in slight contact with the zirconia surface

Er YAG laser treatments

The specimens were treated by using Er: YAG laser (Fotona AT Fidelis, Ljubljana, Slovenia) at 2940 nm. A 90°-angled dental hand instrument (R14-C, Fotona) with a sapphire cylindrical (1.3 × 12 mm) fiber-optic tip was used at an incidence angle of 90° with water irrigation. Air pressure and also water pressure were fixed at 2 bars. Moving up and down, zirconia surfaces were processed with the application tip in slight contact [28, 29] . The applied energy levels are given below:

  • Group150SP: The applied energy level was 150 mJ with 10hz frequency for 45 s. The width of the pulse was 100 μS.

  • Group150SSP: The applied energy level was 150 mJ with 10hz frequency for 45 s. The width of the pulse was 300 μS.

  • Group300SP: The applied energy level was 300 mJ with 10hz frequency for 45 s. The width of the pulse was 100 μS.

  • Group300SSP: The applied energy level was 300 mJ with 10hz frequency for 45 s. The width of the pulse was 300 μS.

All of the specimens were fixed into a brass mold by using silicone impression material and on each sample a profilometer was used to asses three roughness measurements (Ra, μm) (Perthometer M2, Mahr GmbH, Göttingen, Germany). A cut-off value of 0.25 mm was found, which allowed the detection of only those irregularities. A diamond stylus (NHT-6), which had a radius of 2 μm and 90° angle, was transversed with a force of 0.7 N at a constant speed opposite each of the finished samples. The profilometer was calibrated before measurements of each group. Care was taken to make profilometer records as close as possible to the sample center. The mean of the surface roughness measurements was evaluated to find out the samples’ surface properties. Ra is an average value which gives information about the surfaces traced by the profilometer. When this value (Ra) is low it means smoother surface.

One-way analysis of variance (ANOVA) and post-hoc multiple comparisons Tukey’s tests were used to analyze the surface roughness measurements (Ra values) at a significance level of p < 0.01 by using statistical software program (SPSS for Windows, Version 12.0.1; SPSS Inc., Chicago, IL, USA).

For the assessment of surface morphology, the samples used profilometry were coated in gold-palladium (Quorum Technologies Polaron SC7620, Newhaven, East Sussex, UK) and observed under SEM (JSM-6610 LV Scanning electron microscope, JEOL USA) with 50X, 100X, and 500X magnifications and 15 kV voltage.

Results

Statistical analyses of surface roughness values treated with different groups found are presented in Table 2. ANOVA test and also Tukey’s test (α = 0.05) showed that all treatments yielded different mean surface roughness values. The highest Ra value was showed in Sandblasting group (GroupS = .876 ± .067). After it CO2 laser treatment group followed (Group4W = .622 ± .177) it. There were statistical differences between the two groups (p < .001). Also there were statistically significant differences rest of the groups. There were no statistically significant differences among control (GroupC), CO2 laser treatment (Group3W) and Er YAG laser treatments (Group150SP, Group150SSP, Group300SP, Group300SSP). Similar morphologic differences were seen between the specimens’ surfaces after different surface treatments in SEM images. When compared with other treatments, irregular surface pattern on Y-TZP surfaces were found only in sandblasting with 50 mμ Al2O3 particles. Perceptible loss of material was caused by Er: YAG laser irradiations and smooth areas confined by small fissures and narrow microcracks on the surfaces appeared due to both energy intensities. SEM examination showed softening, too much loss of mass and existence of deep cracks in these groups (Fig. 3). Only 4 W CO2 laser irradiation caused scaly irregularities on the Y-TZP surfaces. An increase on surface roughness was seen as a result of treatment of the surface of Y-TZP discs with CO2 laser as a result of comparison with Er: YAG laser (p < 0.001) and this result was thought to occur because of the degree and kind of surface irregularities produced on the Y-TZP surface.

Table 2 Mean surface roughness (Ra) values and standard deviations (SD) of the groups
Fig. 3
figure 3

SEM photographs of study groups at magnification of X500 roughened areas for CO2 and micro cracks for ER-YAG laser were shown with arrows

Discussion

The purpose of this study was to investigate the effect of air abrasion with Al2O3 and different outputs and energy levels of Er: YAG and CO2 laser applications on the surface roughness of Y-TZP. Since the tested treatments changed the surface roughness of Y-TZP, it can be said that the null hypothesis could be accepted based on the results of this investigation.

There are a great number of extraoral methods to obtain a robust and permanent bond between the tooth and restoration like sandblasting and mechanical abrasion [30, 31]. While creating micro cracks in zirconia to increase retention also it weakens mechanical properties of zirconia [32, 33]. Therefore, recently other methods such as laser etching have been introduced to create surface roughening and for the enhancement of zirconia–veneer ceramic interfacial bonding and integration this method might prove to be a new method of surface treatment [34]. It should be kept in mind that surface alterations may occur on the zirconia surface due to laser irradiation.

In many studies air abrasion is the most effective surface treatment method than the application of CO2 and Er:YAG lasers and it has been shown that irradiation of Y-TZP ceramic surfaces with CO2 and Er:YAG lasers did not increase the surface roughness significantly [25, 30, 35,36,37]. Conditioning with air-abrasion can cause resin-ceramic bonding through improving surface roughness and bonding surface area [38]. Demir et al. [35] stated that in order to get micromechanical retention before luting, air abrasion or 400 mJ Er: YAG laser energy can be used; however, air abrasion was found be the most useful surface treatment method since it had significantly higher surface roughness values than the control group and different modes of Er: YAG (200, 300, and 400 mJ). In their study they analyzed Y-TZP disks after irradiation of various power settings with Er: YAG, CO2 or diode laser, Stübinger et al. [36] found diode lasers to be the best system giving surface preservation and safety in the treatment of zirconia implant surface. In their study they analyzed different Er: YAG laser energy intensities to find out their influences on surface roughness and morphologic characteristics of Y-TZP, Cavalcanti et al. [27] concluded that excessive material deterioration occurred as a result of greater laser modes (400 and 600 mJ), making them unfit as surface treatments for zirconia surfaces. In a study by Miranda et al. [37], which examined the surface roughness on Y-TZP surface after Er: YAG laser irradiation at 1.5 W/20 hz concluded that laser irradiation caused a decrease on surface roughness. The present study used maximum 300 mJ output power of the Er: YAG laser with various pulse widths and it was found that it did not roughen the Y-TZP surfaces to accept a surface treatment method. Miranda et al. [38] evaluated the surface roughness on Y-TZP surface after Er, Cr: YSGG laser irradiation at 1.5 W/20 Hz and found that laser irradiation decreased the surface roughness.

Ersu et al. [21] compared and assessed the results of CO2 laser and conventional surface treatments on surface roughness of in-ceram zirconia discs to dentin and found that for roughening surfaces of the specimens, sandblasting was an effective surface treatment and surface roughness was not increased by CO2 laser irradiation.

In contrast to above studies, irradiation of zirconia ceramic surfaces with CO2 and Er:YAG lasers resulted with an increase on surface roughness in many studies [31, 39,40,41]

Untreated, sandblasting, laser irradiations (Er: YAG and Nd: YAG laser) and mixtures of these laser applications with sandblasting on pre-sintered ZrO2 was examined by Kirmali et al. [39] and they concluded that the surface roughness values increased significantly through sandblasting and Er: YAG laser applications on pre-sintered ZrO2 substructures.

Liu et al. [40] performed X-ray diffractometric analysis of zirconia ceramic samples after sandblasting and CO2 laser application. In parallel with this study, both sandblasting and CO2 laser irradiation were found to raise the surface roughness of zirconia specimens, while the laser group had lower increase than the sandblasting group.

The results of various lasers and particle abrasion on surface characteristics of zirconia ceramics were examined by Arami et al. [41] and all treated surfaces were found to have higher roughness than the control group. Similar surface roughness was found in surfaces treated by Er: YAG laser and air abrasion showing that this laser can be a suitable substitute for air abrasion.

In order to find out the reliable intensity for roughening the Y-TZP surface, different energy intensities were compared in a great number of studies. The laser energy settings in our study were chosen according to previous study reports and the effectiveness of these parameters was examined [23,24,25,26,27,28,29]. The methodology for laser application on ceramic specimens is based upon a series of pilot studies. However, there is a need for further studies since there is still no clear information about effective laser type and applications modes.

Y-TZP’s surface roughness was described by Ra parameters found with a profilometer. The overall roughness of a surface is described by this parameter and it is the arithmetical average value of all absolute distances of the roughness profile from the center line within the measuring length. Surface roughness (Ra) is the finer irregularities of the surface texture caused by the action of the production process or material condition and is expressed in micrometers (μm).

Besides the type of irradiated ceramic, the range of superficial changes on Y-TZP ceramic surface is based on the energy density of the laser radiation. The main effect of laser energy is the conversion of light energy into heat, with the most significant between the laser and substrate being the absorption of the laser energy by the substrate.

In addition to other surface qualities, pigmentation of the surface and its water content determines the extent of energy absorbed by the irradiated surface. Since Y-TZP ceramic is white opaque and there is no water content, retention of laser energy is difficult on Y-TZP ceramic. In this study, graphite powder is used to raise retention of laser energy by zirconia in all laser groups due to this difficulty. Sandblasted specimens showed significant higher surface roughness than the others despite this method among all groups. Higher surface roughness measurements when compared with other laser treated groups were found only in 4 W CO2 laser irradiated Y-TZP groups and statistically significant differences were found (p < .05). Group C, Group 3 W, Group 150SP, Group 150SSP, Group 300 SP, and Group 300 SSP did not show statistically different surface roughness measurements (p > .05). An uneven surface with a rough characteristic is formed by applying Er YAG laser and the micro cracks found in electron microscope evaluation confirm our results (Fig. 3).

According to the results of our study, the most effective treatment to roughen the surface of Y-TZP ceramics was found to be sandblasting. Different Er Yag and CO2 laser irradiations did not roughen the Y-TZP ceramics. Surface roughness measurements after laser irradiations were almost the same with those of the control group.

Conclusions

The materials used in our study gave different surface roughness, the highest surface roughness values were found with sandblasting. Nevertheless, there were no significant differences in surface roughness between laser treated specimens and the control, except 4 W CO2 laser treatment. As an alternative surface treatment to sandblasting for Y-TZP ceramics, only 4 W CO2 (Group 4 W) laser irradiation technique is recommended (p < .05).

Abbreviations

ANOVA:

One-way analysis of variance

CAD CAM:

Computer aided design computer aided manufacturing

CO2:

Carbon dioxide laser

Er YAG:

Erbium yttrium aluminum garnet

Ra:

Roughness measurements

Y-TZP:

Yttrium stabilized tetragonal polycrystalline zirconia

References

  1. Cavalcanti AN, Foxton RM, Watson TF, Oliveira MT, Giannini M, Marchi GM. Y-TZP ceramics: key concepts for clinical application. Op Dent. 2009;3:344–51.

    Article  Google Scholar 

  2. Vagkopoulou T, Koutayas SO, Koidis P, Strub JR. Zirconia in dentistry: part 1. Discovering the nature of an upcoming bioceramic. J Esthet Dent. 2009;4:130–51.

    Google Scholar 

  3. Sen D, Poyrazoglu E, Tuncelli B, Göller G. Shear bond strength of resin luting cement to glass-infiltrated porous aluminumoxide cores. J Prosthet Dent. 2000;83:210–5.

    Article  Google Scholar 

  4. Piwowarczyk A, Lauer HC, Sorensen JA. The shear bond strength between luting cements and zir- conia ceramics after two pre-treatments. Oper Dent. 2005;30:382–8.

    PubMed  Google Scholar 

  5. Kramer N, Lohbauer U, Frankenberger R. Adhesive luting of indirect restorations. Am J Dent. 2000;13:60–76.

    Google Scholar 

  6. Ozcan M, Pfeiffer P, Nergiz I. A brief history and current status of metal-and-ceramic surface-conditioning concepts for resin bonding in dentistry. Quintessence Int. 1998;29:713–24.

    PubMed  Google Scholar 

  7. Shiu P, De Souza-Zaroni WC, Eduardo C de P, Youssef MN. Effect of feldspathic ceramic surface treatments on bond strength to resin cement. Photomed Laser Surg. 2007;25:291–6.

    Article  Google Scholar 

  8. Akyıl MS, Yilmaz A, Karaalioğlu OF, Duymuş ZY. Shear bond strength of repair composite resin to an acid-etched and a laser-irradiated Feldspathic ceramic surface. Photomed Laser Surg. 2010;28:539–45.

    Article  Google Scholar 

  9. Delme KI, De Moor RJ. Scanning electron microscopic evaluation of enamel and dentin surfaces after Er: YAG laser preparation and laser conditioning. Photomed Laser Surg. 2007;25:393–401.

    Article  Google Scholar 

  10. Esteves-Oliveira M, Zezell DM, Apel C, et al. Bond strength of self-etching primer to bur cut, Er,Cr:YSGG, and Er: YAG lased dental surfaces. Photomed Laser Surg. 2007;25:373–80.

    Article  Google Scholar 

  11. Liu JF, Lai YL, Shu WY, Lee SY. Acceptance and efficiency of Er: YAG laser for cavity preparation in children. Photomed Laser Surg. 2006;24:489–93.

    Article  Google Scholar 

  12. Soares LE, Resende EB, Brugnera A Jr, Zanin FA, Martin AA. Combined FT-Raman and SEM studies of the effects of Er: YAG laser irradiation on dentin. Photomed Laser Surg. 2007;25:239–44.

    Article  Google Scholar 

  13. Raucci-Neto W, De Castro LM, Correa-Afonso AM, Da Silva RS, Pecora JD, Palma-Dibb RG. Assessment of thermal alteration during class V cavity preparation using the Er: YAG laser. Photomed Laser Surg. 2007;25:281–6.

    Article  Google Scholar 

  14. Gokce B, Ozpinar B, Dundar M, Comlekoglu E, Sen BH, Gungor MA. Bond strengths of all-ceramics: acid vs. laser etching. Oper Dent. 2007;32:173–8.

    Article  Google Scholar 

  15. Kern M, Swift EJ Jr. Bonding to zirconia. J Esthet Restor Dent. 2011;23:71–72.15.

    Article  Google Scholar 

  16. Al-Batayneh OB, Seow WK, Walsh LJ. Assessment of Er:YAG laser for cavity preparation in primary and permanent teeth: a scanning electron microscopy and thermographic study. Pediatr Dent. 2014;36(3):90–4.

    PubMed  Google Scholar 

  17. Aizawa K, Kameyama A, Kato J, et al. Resin bonding to dentin irradiated by high repetition rate Er: YAG laser. Photomed Laser Surg. 2006;24:397–401.

    Article  Google Scholar 

  18. Bader C, Krejci I. Indications and limitations of Er: YAG laser applications in dentistry. Am J Dent. 2006;19:178–86.

    PubMed  Google Scholar 

  19. Chinelatti MA, Ramos RP, Chimello DT, Corona SA, Pecora JD, Dibb RG. Influence of Er: YAG laser on cavity preparation and surface treatment in microleakage of composite resin restorations. Photomed Laser Surg. 2006;24:214–8.

    Article  Google Scholar 

  20. Van As G. Erbium lasers in dentistry. Dent Clin N Am. 2004;48:1017–59.

    Article  Google Scholar 

  21. Pick RM, Colvard MD. Current status of lasers in soft tissue dental surgery. J Periodontol. 1993;64:589–602.

    Article  Google Scholar 

  22. Wigdor HA, Walsh JT Jr, Featherstone JD, Visuri SR, Fried D, Waldvogel JL. Lasers in dentistry. Lasers Surg Med. 1995;16:103–33.

    Article  Google Scholar 

  23. Ural Ç, Külünk T, Külünk Ş, Kurt M. The effect of laser treatment on bonding between zirconia ceramic surface and resin cement. Acta Odontol Scand. 2010;68:354–9.

    Article  Google Scholar 

  24. Akyil MS, Uzun IH, Bayindir F. Bond strength of resin cement to yttrium-stabilized tetragonal zirconia ceramic treated with air abrasion, silica coating, and laser irradiation. Photomed Laser Surg. 2010;28:801–8.

    Article  Google Scholar 

  25. Ersu B, Yuzugullu B, Ruya Yazici A, Canay S. Surface roughness and bond strengths of glass-infiltrated alumina ceramics prepared using various surface treatments. J Dent. 2009;37:848–56.

    Article  Google Scholar 

  26. Chen JR, Oka K, Kawano T, Goto T, Ichikawa T. Carbon dioxide laser application enhances the effect of silane primer on the shear bond strength between porcelain and composite resin. Dent Mater J. 2010;29:731–7.

    Article  Google Scholar 

  27. Akova T, Yoldes O, Toroglu MS, Uysal H. Porcelain surface treatment by laser for bracket-porcelain bonding. Am J Orthod Dentofac Orthop. 2005;128:630–7.

    Article  Google Scholar 

  28. Enwemeka CS. Standard parameters in laser phototherapy. Photomed Laser Surg. 2008;26(5):411. https://doi.org/10.1089/pho.2008.9770.

    Article  PubMed  Google Scholar 

  29. Jenkins PA, Carroll JD. How to report low-level laser therapy (LLLT)/ Photomedicine dose and beam parameters in clinical and laboratory studies. Photomed Laser Surg. 2011;29:785–7.

    Article  Google Scholar 

  30. Cavalcanti AN, Pilecki P, Foxton RM, et al. Evaluation of the surface roughness and morphologic features of Y-TZP ceramics after different surface treatments. Photomed Laser Surg. 2009;27:473–9.

    Article  Google Scholar 

  31. Casucci A, Mazzitelli C, Monticelli F, Toledano M, Osorio R, Osorio E, et al. Morphological analysis of three zirconium oxide ceramics: effect of surface treatments. Dent Mater. 2010;26:751–60.

    Article  Google Scholar 

  32. Kasraei S, Atefat M, Beheshti M, Safavi N, Mojtahedi M, Rezaei-Soufi L. Effect of surface treatment with carbon dioxide (CO2) laser on bond strength between cement resin and zirconia. J Lasers Med Sci. 2014;5:115–20.

    PubMed  PubMed Central  Google Scholar 

  33. Karakoca S, Yilmaz H. Influence of surface treatments on surface roughness, phase transformation, and biaxial flexural strength of Y-TZP ceramics. J Biomed Mater Res B Appl Biomater. 2009;91:930–7.

    Article  Google Scholar 

  34. Yun JY, Ha SR, Lee JB, Kim SH. Effect of sandblasting and various metal primers on the shear bond strength of resin cement to Y-TZP ceramic. Dent Mater. 2010;26:650–8.

    Article  Google Scholar 

  35. Demir N, Subaşı MG, Ozturk AN. Surface roughness and morphologic changes of zirconia following different surface treatments. Photomed Laser Surg. 2012;30:339–45.

    Article  Google Scholar 

  36. Stubinger S, Homann F, Etter C, Miskiewicz M, Wieland M, Sader R. Effect of Er: YAG, CO2 and diode laser irradiation on surface properties of zirconia endosseous dental implants. Lasers Surg Med. 2008;40:223–8.

    Article  Google Scholar 

  37. Miranda PV, Rodrigues JA, Blay A, Shibli JA, Cassoni A. Surface alterations of zirconia and titanium substrates after Er, Cr:YSGG irradiation. Lasers Med Sci. 2015;30(1):43–8.

    Article  Google Scholar 

  38. Kern M, Barloi A, Yang B. Surface conditioning influences zirconia ceramic bonding. J Dent Res. 2009;88:817–22.

    Article  Google Scholar 

  39. Kirmali O, Akin H, Ozdemir AK. Shear bond strength of veneering ceramic to zirconia core after different surface treatments. Photomed Laser Surg. 2013;31:261–8.

    Article  Google Scholar 

  40. Liu D, Matinlinna JP, Tsoi JK, Pow EH, Miyazaki T, Shibata Y, Kan CW. A new modified laser pretreatment for porcelain zirconia bonding. Dent Mater. 2013;29:559–65.

    Article  Google Scholar 

  41. Arami S, Tabatabae MH, Namdar SF, Chiniforush N. Effects of different lasers and particle abrasion on surface characteristics of zirconia ceramics. J Dent. 2014;11:233–41.

    Google Scholar 

Download references

Acknowledgments

Not applicable.

Funding

There was no funding association with this work.

Availability of data and materials

All data generated or analysed during this study are included in this published article.

Author information

Authors and Affiliations

Authors

Contributions

Study design: GEK. Performed the study: GEK, ID. Analysed the data: ID. Wrote the paper: GEK, ID. All authors of this study have read and approved the manuscript.

Corresponding author

Correspondence to Goknil Ergun Kunt.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ergun Kunt, G., Duran, I. Effects of laser treatments on surface roughness of zirconium oxide ceramics. BMC Oral Health 18, 222 (2018). https://doi.org/10.1186/s12903-018-0688-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-018-0688-y

Keywords