Dental implant location via surface scanner: a pilot study

Abstract

Purpose

Implant location is performed after placement to verify that the safety of neighboring anatomic structure and the realizability of prosthetic plan. Routine postoperative location is based on radiological scanning and raises the concerns on radiation exposure and inconveniency in practice. In the present study a location method based on surface scanning was introduced and the accuracy of this method was assessed in vitro.

Material and methods

A total of 40 implants were placed in 10 resin mandible models. The models were scanned with intraoral scanner (IS group) and extraoral scanner (ES group). The implant position was located with fusing the images of surface scanning and cone beam computerized tomography (CBCT) after implant placement. Deviations were measured between positions located by surface scanner and postoperative CBCT with the parameters: central deviation at apex (cda), central deviation at hex (cdh), horizontal deviation at apex (hda), horizontal deviation at hex (hdh), vertical deviation at apex (vda), vertical deviation at hex (vdh) and angular deviation (ad).

Results

In IS group, the mean value of cda, cdh, hda, hdh, vda, vdh and ad was 0.27 mm, 0.23 mm, 0.12 mm, 0.10 mm, 0.21 mm, 0.19 mm and 0.72°, respectively. In ES group, the mean value of cda, cdh, hda, hdh, vda, vdh and ad was 0.28 mm, 0.25 mm 0.14 mm, 0.11 mm, 0.22 mm, 0.20 mm and 0.68°, respectively. The implant deviations in IS and ES groups were of no significant difference for any of the measurements.

Conclusions

Dental implant can be located via surface scanner with acceptable accuracy for postoperative verification. Further clinical investigation is needed to assess the feasibility of the method.

Peer Review reports

Backgrounds

Although the position of oral implant could be planned quite perfectly before surgery, implant deviation is inevitable regardless of guidance strategy and operator experience [1, 2]. Postoperative examination is required routinely to confirm that the critical anatomical structures are well preserved, and the implant is placed in the planned position that favors the following prosthetic restoration.

For present, the methods to assess the inserted implant are mostly based on radiology, such as intraoral radiography, panoramic radiography and cone beam computed tomography (CBCT). CBCT offers the three-dimensional view on the implant and adjacent tissue, and is thought to be the most accurate way to locate the implant position [3]. But due to the consideration on radiation safety, dental radiological devices have to be kept in a separate room with protective barrier. For this reason, the examination is normally performed after the whole surgery is completed and the patient leaves the operation room. If unacceptable deviation occurs, it’s would be a difficult decision for the surgeon whether a corrective reentry is necessary. Another drawback of postoperative radiography is that the patient’s acceptability is often limited due to the growing concern on radiation hazard [4].

Alternative methods to locate implant postoperatively included measuring the difference in the angles of virtual planned and definitively used abutment replicas on the working model [5], or fusing the preoperative planned implant with the postoperative CBCT scans of the implant replicas of model cast [6]. These methods are unable to tell the surgeon the exact position of implant immediately after insertion and don’t allow any adjustment if the implant is not in a favorable position, because they generally require conventional impression and cast production, which are quite time-consuming.

A more simplified solution could be the introduction of digital impression. Digital impression is a three-dimensional model reconstructed from the data collected by a surface scanner, which records the geometry of tissue surface by measuring the light reflection times of the subject surface. The whole procedure usually costs within few minutes [7]. With known and mature digital tools, digital impression can be fused with CBCT images by matching fiducial points, such as remaining tooth, temporary implant and fixation pin [8]. Considering that the spatial relation of fiducial points to local anatomical structure is constant in a certain period, it’s possible to assess the position of placed implant to neighboring tissue by measuring the its relation to fiducial points. The digital impression with placed implants can also be directly used in following design and fabrication of immediate dental prosthetics, if the implant is in the ideal position. In this workflow, the surgical and restorative phases of the treatment would be both greatly simplified. Furthermore, intraoperative adjustment of deviated implant becomes possible if a handheld device is used, most mainstream commercial intraoral scanners for example.

However, in order to include fiducial points that are distributed dispersedly, a full arch scan is necessary to assess the position of placed implant. Although the precision of surface scanning devices has improved considerably over these years, and now is acceptable for crowns and short fixed prostheses [9, 10], the devices are still vulnerable to inaccuracies in scanning of increased span [11]. Whether the accuracy via surface scanner fulfills the needs in implant location is questionable.

In this study, the authors attempted to assess the implant position by matching scan body that was virtual-connected to implant with the surface scanning image. The accuracy of location was assessed by comparison with postoperative CBCT examination.

Materials and methods

Implant placement

A mandible model with bilateral second premolar and first molar missing was designed based on the CBCT data of a volunteer’s mandible in the software (Mimics v16.0, Materialise, Leuven, Belgium). A total of ten resin replicas of the designing were fabricated using rapid prototyping (ProJet 3500 HD MAX, 3D System, Rock Hill, SC, USA). Four implants (Straumann Standard Plus RN, 4.1 × 10 mm, Straumann AG, Basel, Switzerland) were placed in the edentulous regions of each model. The models were scanned with CBCT (3D eXam, Kavo, Bismarckring, Germany) postoperatively.

Implant location

A scan body (Regular Neck 4.8/10 mm, Straumann AG) was connected to the implant placed in the model (Fig. 1a). The models were scanned with an intraoral scanner in accordance with the manufacturer’s instruction (DL-100, Launca, Shenzhen, China) (Fig. 1b). The digital models of scan body and implant were connected virtually and simplified to a cylinder model for the convenience of following measurement (Fig. 1c, named as location component in the article) in the software (Mimics v16.0). The location component was fused into the surface scanning images using the geometry of scan body as reference (Fig. 1d). The implant part of location component was considered as the position of implant.

Fig. 1
figure 1

Dental implant location via surface scanner. a Scan bodies were connected to the implants placed in the resin mandible model; b the model was scanned with intraoral and extraoral scanners; c location component: scan body and implant were connected virtually and the implant part was simplified to a cylinder of same diameter and length; d The location component was aligned in the surface scanning image by matching the geometry of scan body; e the resulted model in (d) was aligned in the CBCT images by matching the remaining cusps in the model; f The positions located via surface scanning was compared with the position in CBCT in the software 3-matic

Each model was also scanned with an extraoral scanner (D810, 3Shape, Copenhagen, Denmark). The position of implant was measured using the methods described above. All CBCT and surface scanning procedures were performed by the same operator.

Accuracy analysis

The surface scanning/location component-fusing images was fused with the postoperative CBCT image using the adjacent cusps as reference (Fig. 1e). The fused images of location component, surface scanning and postoperative CBCT was uploaded into the software (3-Matic, Materialise, Leuven, Belgium) (Fig. 1f). The position deviations between implant located with surface scanning and postoperative CBCT were measured using seven parameters: central deviation at apex; central deviation at hex, horizontal deviation at apex, horizontal deviation at hex, vertical deviation at apex, vertical deviation at hex and angular deviation (Fig. 2). The values were standardized by the actual implant dimensions.

Fig. 2
figure 2

Assessment of position deviation. The parameters were defined as: cda, central deviation at apex; cdh, central deviation at hex, hda, horizontal deviation at apex, hdh, horizontal deviation at hex, vda, vertical deviation at apex, vdh, vertical deviation at hex; ad, angular deviation

Statistical analysis

The data was analyzed using GraphPad Prism 5 (GraphPad Software, La Jolla, CA, USA). Deviation parameters were presented as mean, maximal/minimal value (max./min.), standardized deviation (SD) and 95% confidence interval (95% CI). Statistical differences between implant deviations measured with intraoral scanner and extraoral scanner were assessed by paired t test. The level of significance was set at 0.05 in all tests.

Results

In IS group, the value of cda, cdh, hda, hdh, vda, vdh and ad was 0.27 ± 0.14 mm, 0.23 ± 0.13 mm, 0.12 ± 0.10 mm, 0.10 ± 0.07 mm, 0.21 ± 0.15 mm, 0.19 ± 0.14 mm and 0.72 ± 0.52 degrees, respectively. In ES group, the value of cda, cdh, hda, hdh, vda, vdh and ad was 0.28 ± 0.14 mm, 0.25 ± 0.12 mm 0.14 ± 0.11 mm, 0.11 ± 0.10 mm, 0.22 ± 0.14 mm, 0.20 ± 0.12 mm and 0.68 ± 0.54 degrees, respectively. The implant deviations measured with intraoral scanner and extraoral scanner were summarized (Tables 1 and 2). There is no significant difference between the implant deviations measured with intraoral scanner and extraoral scanner (Fig. 3).

Table 1 Position deviations located with intraoral scanner
Table 2 Position deviations located with extraoral scanner
Fig. 3
figure 3

Comparison of position deviation between ES (extraoral scanner) and IS (intraoral scanner) groups. There is no significant difference between the implant deviations measured via intraoral scanner and extraoral scanner (p > 0.05)

Discussion

In this pilot study, the surface scanning was fused with the postoperative CBCT based on manually-selected fiducial points. The virtually-defined implant by scan body was compared with the implant shown in CBCT. The deviation from CBCT measurement was less than 0.3 mm linearly and 0.8° angularly. The accuracy was superior to other reported location methods alternative to postoperative radiology [5, 6].

In the presented method, the implant position was calculated based on the position of scan body. Different to the direct scanning with CBCT, the relation of implant to neighboring structures such as bony envelop, mandibular nerve and Schneiderian membrane could not be viewed directly. The indirect location is of no clinical significance unless the accuracy of the calculated position within the safe zone that is mandatory in implant surgery. A safety distance of 0.5–1 mm between the implant and vital anatomy is recommended in most literatures [12, 13]. The deviations reported in the present study were within these recommendations and should therefore be regarded as clinically acceptable if the recommendations are followed in preoperative planning. The location accuracy of our study suggested the potential application of surface scanner in assessing implant location postoperatively.

Theoretically, the direct scanning using the intraoral scanning device could result in the same and even higher accuracy than the indirect digitalization of the gypsum cast using an extraoral scanner [14]. Indirect digitalization actually collects the errors of conventional impression taking, cast pouring, impression material distortion, and optical scanning as in direct digitalization. In this study, both intraoral and extraoral scanner were used to locate the placed implant in the same in vitro model, both of which resulted in the same accuracy. Our results provided experimental evidence to clear up the doubt that intraoral and extraoral scanners may differ when applied in implant location, and suggested that the digitalization of conventional impression is no longer necessary since intraoral scanner offers accuracy of same level. More importantly, application of handheld device as intraoral scanner opens the possibility to assess implant bed preparation by scanning the indicator inserted, and to perform the implant verification immediately after insertion. The immediacy facilitates necessary intraoperative adjustment if significant deviation is detected.

Although the location accuracy shown in this model study is encouraging, it’s important to notice that the accuracy of surface scanning was generally reported to be better in vitro than that in vivo, due to better access, better space for the scanning unit, better illumination, no patient movement, and absent of saliva and blood that may lead to possible fogging of the optical unit [15]. So, whether the accuracy reported here could be realized in clinical scenarios remains to be investigated. Furthermore, the accuracy in vivo could be influenced by the length of clinical career and the region being scanned [16]. Impressions of angulated implants might also diminish the accuracy of the impression [17]. As a result, further in vivo investigations are compulsory to verify the feasibility of this method, even if acceptable accuracy was achieved in the present study.

When the proposed method used in clinical practice, the main purpose should be to assess the relation of implant and anatomic structure. The surface scanning for location is fused with preoperative planning model depicted with CBCT and surface scanning. Image fusing requires manual processing for selecting the fiducial points, which introduce errors inevitably [18]. It would be quite challenging to reduce image fusing error in clinical scenarios where the condition usually varies greatly in different cases. Beside error, manual operation is also suboptimal in daily clinical operation. Because additional trained person and computer workstation are needed during the surgery. Unfortunately, the accuracy of full-automatic fusing method is still unsatisfactory limited to the current technology [19]. More technical adaptions are in need before the practicality of implant location via surface scanner.

Conclusion

Dental implant can be located via surface scanner. The accuracy of implant location with intraoral scanner could be acceptable for postoperative verification, but the efficiency and efficacy in clinical practice need further investigation in future.

Availability of data and materials

All essential data is presented in the manuscript. The step-by-step datasets and images during the current research are available from the corresponding author on reasonable request.

References

  1. Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implant Res. 2017;28(5):602–12.

    Article  Google Scholar 

  2. D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H, Tahmaseb A. Current state of the art of computer-guided implant surgery. Periodontol 2000. 2017;73(1):121–33.

    Article  Google Scholar 

  3. Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health. 2018;18(1):88.

    Article  Google Scholar 

  4. Stratis A, Zhang G, Jacobs R, Bogaerts R, Bosmans H. The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice. Eur Radiol. 2019;29:7009–18.

    Article  Google Scholar 

  5. Naitoh M, Ariji E, Okumura S, Ohsaki C, Kurita K, Ishigami T. Can implants be correctly angulated based on surgical templates used for osseointegrated dental impla nts? Clin Oral Implant Res. 2000;11(5):409–14.

    Article  Google Scholar 

  6. Nickenig HJ, Eitner S. An alternative method to match planned and achieved positions of implants, after virtual planning using cone-beam CT data and surgical guide templates–a method reducing patient radiation exposure (par t I). J Craniomaxillofac Surg. 2010;38(6):436–40.

    Article  Google Scholar 

  7. Rutkūnas V, Gečiauskaitė A, Jegelevičius D, Vaitiekūnas M. Accuracy of digital implant impressions with intraoral scanners. A systematic review. Eur J Oral Implantol. 2017;10(Suppl 1):101–20.

    Google Scholar 

  8. Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implant Res. 2012;23(Suppl 6):112–23.

    Article  Google Scholar 

  9. Mangano FG, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 2019;19(1):101.

    Article  Google Scholar 

  10. Runkel C, Güth JF, Erdelt K, Keul C. Digital impressions in dentistry-accuracy of impression digitalisation by desktop scanners. Clin Oral Investig. 2020;24(3):1249–57.

    Article  Google Scholar 

  11. Flügge T, van der Meer WJ, Gonzalez BG, Vach K, Wismeijer D, Wang P. The accuracy of different dental impression techniques for implant-supported dental prostheses: a systematic review and meta-analysis. Clin Oral Implant Res. 2018;29(Suppl 16):374–92.

    Article  Google Scholar 

  12. Bover-Ramos F, Viña-Almunia J, Cervera-Ballester J, Peñarrocha-Diago M, García-Mira B. Accuracy of implant placement with computer-guided surgery: a systematic review and meta-analysis comparing cadaver, clinical, and in vitro studies. Int J Oral Maxillofac Implants. 2018;33(1):101–15.

    Article  Google Scholar 

  13. Sicilia A, Botticelli D, Working G. Computer-guided implant therapy and soft- and hard-tissue aspects. The third EAO consensus conference 2012. Clin Oral Implants Research. 2012;23(Suppl 6):157–61.

    Article  Google Scholar 

  14. Bohner LOL, De Luca CG, Marció BS, Laganá DC, Sesma N, Tortamano Neto P. Computer-aided analysis of digital dental impressions obtained from intraoral and extraoral scanners. J Prosthet Dent. 2017;118(5):617–23.

    Google Scholar 

  15. Keul C, Güth JF. Accuracy of full-arch digital impressions: an in vitro and in vivo comparison. Clin Oral Investig. 2020;24(2):735–45.

    Article  Google Scholar 

  16. Giménez B, Özcan M, Martínez-Rus F, Pradíes G. Accuracy of a digital impression system based on active wavefront sampling technology for implants considering operator experience, implant angulation, and depth. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e54-64.

    Article  Google Scholar 

  17. Lim JH, Park JM, Kim M, Heo SJ, Myung JY. Comparison of digital intraoral scanner reproducibility and image trueness considering repetitive exp erience. J Prosthet Dent. 2018;119(2):225–32.

    Article  Google Scholar 

  18. Rangel FA, Maal TJJ, de Koning MJJ, Bronkhorst EM, Bergé SJ, Kuijpers-Jagtman AM. Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study. Clin Oral Investig. 2018;22(3):1215–22.

    Article  Google Scholar 

  19. Woo S, Lee S, Chae J, Rim J, Lee J, Seo J, Lee C. Automatic matching of computed tomography and stereolithography data. Comput Methods Programs Biomed. 2019;175:215–22.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The study was financially supported by the National Science and Technology Major Project of China (No. 2016YFC1102900), National Natural Science Foundation of China (No. 81671029), Guangzhou Science Technology and Innovation Commission (No. 201704030024, No. 201803040008) and International Team for Implantology (No. 881-2012). The funders have no role in the experiment designing, performing, data analyze and manuscript preparing.

Author information

Authors and Affiliations

Authors

Contributions

MZ designed and performed the experiment; HZ performed the experiment, SL performed the experiment, YG designed the experiment, analyzed the data and wrote the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yuan-ming Geng.

Ethics declarations

Ethics approval and consent to participate

Not applicable. The research involved no human participants, human material, or human data.

Consent for publication

Not applicable. The manuscript includes no details, images, or videos relating to an individual person.

Competing interests

The authors declare that they have no competing interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhou, M., Zhou, H., Li, Sy. et al. Dental implant location via surface scanner: a pilot study. BMC Oral Health 20, 306 (2020). https://doi.org/10.1186/s12903-020-01297-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-020-01297-y

Keywords