Effect of filament types and loops number on the force degradation of elastomeric chains used for orthodontic treatment: an in-vitro study

Abstract

Background

In orthodontic treatment, closing spaces, specifically the extraction and scattered spaces of the anterior teeth, requires some auxiliary bias, such as an elastomeric chain. Many factors affect the mechanical properties of elastic chains. In this study, we investigated the relationship of the filament type, the number of loops, and the force degradation of elastomeric chains under thermal cycling conditions.

Methods

The orthogonal design included three filament types (i.e., close, medium, and long). Four, five, and six loops of each elastomeric chain were stretched to have an initial force of 250 g in an artificial saliva environment at 37 °C and thermocycling between 5 and 55 °C three times a day. The remaining force of the elastomeric chains was recorded at different time points (4 h, 24 h, 7 days, 14 days, 21 days, and 28 days), and the percentage of the remaining force was calculated.

Results

The force decreased significantly in the initial 4 h and degraded mostly within the first 24 h. In addition, the percentage of force degradation increased slightly between 1 and 28 days.

Conclusions

Under the same initial force, the longer the connecting body is, the fewer the number of loops and the greater the force degradation of the elastomeric chain are.

Peer Review reports

Background

Elastomeric chains have long been popular in the orthodontic field [1]. They are extensively used for intra-arch teeth movement and force transmission [2]. Orthodontists have used these chains because they are practical, convenient, efficient, affordable, low cost, comfortable for patients, and effective in closing spaces by applying light continuous orthodontic forces, which are preferable to produce dentoskeletal alterations [3,4,5,6,7,8,9]. Elastomeric chains, as an adjunct traction device [10], were first introduced in orthodontics at the end of the 1960s not only for canine distal movement, the closure of the extraction space, and the scattered space of the anterior teeth but also for the correction of reversing teeth [11,12,13].

Elastomeric chains come in three forms: short (open), long (wide), and closed (continuous). They can also be transparent, gray, or a different color depending on their specific properties and intended use [14]. Orthodontic manufacturers may keep their exact composition secret for various reasons [14, 15]. The elastomeric chain is a high-molecular-weight polymer made of synthetic rubber, which is made of polyurethane. Polyurethane is a thermosetting polymer formed by diisocyanate’s reaction and polyols’ polymerisation. This polymer is easy to stretch, and internal stress is generated so that it can rebound to its original length after stretching within limits [16]. In the 1970s, a 28-day study reported that most of the force degradation of elastomeric chains occurs on the first day of stretching and then enters a relatively stable state [17]. Many studies have shown that force occurs immediately after the initial stretching and reduces rapidly during the first 24 h, at which 30–50% of the initial force is lost; the relatively flat stage follows, with the force of 28 days showing an increase of approximately 10% only compared with that on the first day [18,19,20].

It has been established that the force decay of elastomeric chains is influenced by various factors, such as time, manufacturer, prestretching, method of manufacturing, color of elastomeric chains/modules, and the oral environment [21, 22]. Many studies have investigated the effects of temperature on the force degradation of elastomeric chains. To study the force degradation of elastomeric chains, some researchers simulated the oral temperature and compared it with the normal environment [23, 24]. They found that the force degradation increases with temperature at 10 °C, 22 °C, and 37 °C. Some researchers compared the force degradation of elastomeric chains at 60 °C and 37 °C and found that the force degradation in the former is significantly higher than that in the latter [25]. Stevenson et al. [26] tested the tensile properties of elastomeric chains at different temperatures and found that the stress relaxation of elastomeric chains gradually increases as the temperature increases. The mechanical properties of elastic chains are affected by several factors. The intrinsic factors of elastomeric chains, such as the type of filament and the number of loops, also have a certain effect. In addition, the temperature of the oral environment easily changes with food intake. However, the above studies selected only the 37 °C saliva environment, without additional thermal cycling conditions to better simulate the oral environment, and these studies were conducted many years ago. The mechanical properties of elastic chains have improved with the technology advances, requiring the improvement of the experimental conditions to validate the results.

The study's null hypothesis would be that there is no relationship between the filament type, the number of loops, and the force degradation of elastomeric chains under thermal cycling conditions. Therefore, this study investigated the force decay of elastic chains with 250 g of initial force in a 37 °C artificial saliva environment for 28 days at different time points and analyzed the effect of connector types and the number of loops on the force decay of the elastic chain under thermal cycling conditions.

Methods

Samples

The study was approved by the ethics committee of the Stomatology Hospital of China Medical University, Liaoning, China (No: CMUKQ-2021-026). All methods were carried out in accordance with the principles of the declaration of Helsinki.

Three elastomeric chains from one company (G&H Orthodontics, 2165 Earlywood Drive, Franklin, IN 46131 USA) were used in this study. Four, five, and six elastomeric chain loops were prepared. According to the orthogonal design, nine groups were formed, namely, the close, medium, and long elastomeric chain groups (A1B1 to A3B3, A1, A2, A3). B1, B2, and B3 are groups of four, five, and six elastomeric chain loops of different elastomeric chains and loops, each containing 15 samples.

Measurement of force decay

First, five samples from each group were used to measure the stretching length to fabricate a tension-fixing device, where the initial force value is fixed at 250 g, a nail is fixed on a wooden board at point X as a fixed point on one side of the elastomeric chain, and the rubber chain is stretched with a dynamometer until the force value reaches 250 g. Mark the position of the end point of the elastomeric chain at this time as Y, and then stretch the elastomeric chain from point X to point Y each time and measure the force value. The samples from each group were dropped into an artificial saliva environment (0.26% NaH2PO4•H2O, 2.17% NaHPO4•7H2O, 0.9%NaCl) at 37 °C [27] in electrothermal constant temperature water bath (HWS12, Shanghai, China). The artificial saliva was changed every three days. A thermocycler system was set at between 5 and 55 °C for 30 min each time and three times a day with an interval of 4 h. The force of the elastomeric chains was measured at different time points (i.e., 4 h, 24 h, 7 days, 14 days, 21 days, and 28 days) using an Eide Fort digital force gauge (HP-2, Fuji, Japan, accuracy to 0.001kgf).

Statistical analysis

All quantitative data are depicted as the mean ± standard deviation and were analyzed using SPSS 25.0.

The Kolmogorov-Smirnov test was performed to check the normality of the data. The difference in degradation percentage between the groups was analyzed by multiple factor variance analysis. Significant differences were defined as p < 0.05.

Results

Remaining force of the elastomeric chains in each group at different time points

The initial force of the elastomeric chains in each group was 250 g. All the forces in the elastomeric chain were gradually reduced in a time-dependent manner. The force decreased significantly in the initial 4 h, and the remaining force dropped to 177.2 ± 5.0 to 204.6 ± 7.7 g in the next 4 h. At 24 h, the remaining force of Groups A1 and A3 was slightly higher and lower than that of the other groups, respectively. The remaining force of all the elastomeric chains dropped to the lowest after 28 days, and the remaining force of Group A3 is the minimum (Table1).

Table 1 Remaining force of the elastomeric chains in each group at different time points. (unit: g)

Percentage of force degradation on the elastomeric chain in each group was measured at different time points

The percentage of force degradation on the elastomeric chain gradually decreased in a time-dependent manner. The percentage of force degradation at 4 h was approximately 20–35%, while the percentage of force degradation at 24 h was nearly 30–45%. The percentage of force degradation is nearly the same between 24 h and 28 days. At 28 days, the percentage of force degradation was 35–55% (Table 2).

Table 2 Percentage of force degradation (%) in each group at different time points

Curve of force degradation on different loops of elastomeric chain

A1 is a group of close elastomeric chains. As shown in Fig. 1, the curve of the four loops is the lowest at the beginning, and the slope of the curve is relatively large. The five- and six-loop groups are relatively straight. The results show that the force degradation decreased slowly as the number of loops increased (Fig. 1).

Fig. 1
figure 1

Force decay curves of the close elastomeric chains with four, five, and six loops, respectively. A1B1, close elastomeric chains with 4 loops. A1B2, close elastomeric chains with 5 loops. A1B3, close elastomeric chains with 6 loops

A2 is a group of medium elastomeric chains. As shown in Fig. 2, the three curves are relatively close to each other. The trends for B2 and B3 are nearly the same, while B1 exhibits the lowest force. The results indicate that the force degradation increases as the number of loops decreases (Fig. 2).

Fig. 2
figure 2

Force decay curves of the medium elastomeric chains with four, five, and six loops, respectively. A2B1, medium elastomeric chains with 4 loops. A2B2, medium elastomeric chains with 5 loops. A2B3, medium elastomeric chains with 6 loops

A3 is a group of long elastomeric chains. The slope of the three groups is nearly the same. The group of six loops is at the top, while the group of four loops is at the bottom. The figure shows that the force degradation increases as the number of loops decreases (Fig. 3).

Fig. 3
figure 3

Force decay curves of the long elastomeric chains with four, five, and six loops, respectively. A3B1, long elastomeric chains with 4 loops. A3B2, long elastomeric chains with 5 loops. A3B3, long elastomeric chains with 6 loops

Force decay curves of elastomeric chains which had different connectors but the same number of loops

B1 is a group of four-loop elastomeric chains. Figure 4 shows that the slopes of the three curves are nearly the same. The close elastomeric chain is at the top, and the long elastomeric chain is at the bottom. The figure shows that the force degradation of the elastomeric chain gradually increases with the increase in the length of the elastomeric chain (Fig. 4).

Fig. 4
figure 4

Force decay curves of the elastomeric chains have four loops and different connectors. A1B1, close elastomeric chains with 4 loops. A2B1, medium elastomeric chains with 4 loops. A3B1, long elastomeric chains with 4 loops

B2 is a group of five-loop elastomeric chains. Figure 5 shows that the starting points of the three curves are relatively concentrated and begin to disperse with different slopes. The closed elastomeric chain is at the top, and the long elastomeric chain is at the bottom. As the connector length increases, the force degradation of the elastomeric chain gradually increases(Fig. 5).

Fig. 5
figure 5

Force decay curves of the elastomeric chains have five loops and different connectors. A1B2, close elastomeric chains with 5 loops. A2B2, medium elastomeric chains with 5 loops. A3B2, long elastomeric chains with 5 loops

B3 is a group of six-loop elastomeric chains. We can see from Fig. 6 that the three curves are relatively concentrated even though they have some intersections. Overall, the close elastomeric chain is at the top, and the long elastomeric chain is at the bottom (Fig. 6).

Fig. 6
figure 6

Force decay curves of the elastomeric chains have six loops and different connectors. A1B3, close elastomeric chains with 6 loops. A2B3, medium elastomeric chains with 6 loops. A3B3, long elastomeric chains with 6 loops

Variance analysis results in a significant difference in the connector type and the loops number double factors’ effect on the force decay speed of elastomeric chains

The results show significant differences in the type of elastomeric chains and the percentage of force degradation at different time points (p < 0.05). Furthermore, the type of elastomeric chain positively affected the rate of force degradation of the elastomeric chain at different time points, and a significant difference can be observed in the number of loops (p < 0.05), which shows that the number of loops affected the rate of force degradation of the elastomeric chains.

Student–Newman–Keuls (SNK) method analysis results in significant differences in different filament types whose influence on the force decay speed of elastomeric chains

The results show a significant difference between different connector types (p < 0.05) at time points. The remaining percentage of force demonstrated when the same initial force was provided is follows: the percentage of degradation in large filaments was the highest, while the percentage of degradation on the six-loop elastic chains decreased the most.

Discussion

Different brands of elastomeric chains made by diverse manufacturers have dissimilar raw materials and production processes [3]. To eliminate variances in brands, we selected and used one brand of elastomeric chains in this study. In clinical practice, elastomeric chains are mainly used for the distal movement of canine teeth and space closing. The most common position of the elastomeric chain is between the towing hook of the buccal tube of the first molar and the distal canine bracket. Therefore, we could choose the different types and numbers of loops according to individual differences. Four and five loops are most commonly used for the distal movement of canine teeth [28,29,30]. This study chose four, five, and six loops to provide significant clinical guidance. As the orthodontic patients’ re-examination time is four weeks, this study cycle was conducted for 28 days [3].

There are studies on the effect of different elastomeric chains based on their morphology and elongation extent on force degradation and the ability of elastomeric chains to maintain force over time, as well as how they may affect the overall effectiveness of the appliance in moving teeth [3, 31]. The type of filament and number of loops can affect force degradation, with shorter filaments resulting in slower degradation [31]. As the number of loops increases, the force degradation rate decreases. Other variables include the material used to make the loops [3]; and oral environments, such as the presence of saliva or different types of oral bacteria [1, 5, 7, 32] and color stability [4] on the degradation of elastomeric chains.

Ajmi et al. [33] studied the influence of the type of filament, the number of loops, and the initial force on the mechanical properties of elastomeric chains in an artificial saliva environment at 37 °C. We improved the experiment based on previous studies, adding the thermal cycle condition between 5 and 55 °C to simulate the oral environment. At the same number of loops, the remaining force of the close, medium, and long filament types increased, and the percentage of force degradation of the long, medium, and close filament types increased. Therefore, at the same number of loops, the long elastomeric chain exhibited the highest force degradation. This result is consistent with that of Sang Ting [31]. In addition, Eliades et al. [30] found that the stress of close elastomeric chains is concentrated on the loops and the types with filaments whose stress is distributed on the connector. The longer the connector is, the greater the force degradation will be, probably because the connector is weaker than the loops, and the force degradation is greater in the connector than in the loops. The longer the connector is, the greater the percentage of stress distribution and the force decay are.

In addition, the results show that when the connector types were the same, the remaining forces of the four, five, and six loops increased, while the force degradation decreased in a time-dependent manner. According to Eliades' theoretical analysis, increasing the number of loops in an elastomeric chain weakens the structure, which is prone to decay but also reduces the degradation of the chain [30].

The main component of the elastomeric chain is polyurethane. One part of polyurethane is made of polyether, a spiral chain of adjustable length. Therefore, they can be easily stretched. Stretching the spiral chains into a line changes the structure from random to orderly chains. At this point, the tensile stress is produced in the polymer, and it tends to exhibit a disorderly spiral chain. In the initial stage, the polymer only expands its curly molecule and overcomes the hydrogen-propelled and Van der Waals forces between molecules. Subsequently, the polymer is extended into chain-like molecules, which continue to be stretched to produce a relatively large stress to overcome the covalent bonds between the molecules. In these two phases, the tensile deformation is reversible. When stretching is stopped, the elastomeric chain reverts to its initial formation. When the force continues to increase, the covalent bond between the polymers molecules would be destroyed; the elastomeric chain would produce irreversible, permanent deformation. The limit of the polymer covalent bonds between molecules is called the elastic limit of the elastomeric chain [34].

With constant temperature and deformation, the viscous strain component of the polymer increases, and the rebound strain component decreases in a time-dependent manner, leading to a gradual decrease in the restoring force over time. This phenomenon is called stress relaxation, which is the force degradation discussed in this study. In other words, when the temperature is constant, stretching the elastomeric chain persistently at a constant force within the elastic limit also causes permanent deformation over time, resulting in force degradation [35].

Storie et al. [36] tested the size of elastomeric chains in water and artificial saliva. Their results show that (in water and artificial saliva replacement) both led to dimensional changes in the elastomeric chains. The dimensional change of the elastomeric chains in the water and artificial saliva environments was larger than before. Due to the liquid’s absorption in the environment, the size of the elastomeric chain changed, leading to force degradation.

Canine retraction is an important use of elastomeric chains in clinical applications. Knowing whether the remaining force of various elastomeric chains can still meet the need to move the canine teeth after a while is extremely important. Boester et al. [37] pointed out that when the force exerted on the canine teeth is less than 55 g, it can not cause any physiological reaction in the periodontal tissueHowever, alveolar bone resorption are difficult to induce. Therefore, the body of teeth hardly moves.

The initial force was large in this study, the remaining force of the elastomeric chain at 28 days was relatively large, and the remaining force of each elastomeric chain at a different time point could still meet the need to move the canine teeth. However, under thermal cycle conditions, the force degradation rate of the long elastomeric chain was as high as 55.3 ± 9.5%, suggesting that when using an elastomeric chain with a long connector and few loops, attention should be paid to controlling the initial force. If the initial force is extremely low, a rebound force with clinical significance may not be provided at 28 days. In this situation, the re-examination time can be appropriately shortened.

In addition, due to the variability of the oral environment, such as the fluoride medium, some proteins and microbes in oral saliva also affect elastomeric chains. Therefore, the degree of force degradation differs between the in vitro experiment and the physical truth in the mouth [34].

Conclusions

According to the present study findings, the null hypothesis was rejected. We concluded that under the initial force, the longer the connecting body is, the fewer the number of loops and the greater the force degradation of the elastomeric chain.

Availability of data and materials

The datasets generated and analyzed during the current study are not publicly available because of (ownership of data) but are available from the corresponding author upon reasonable request. All data and materials were available at the Stomatology Hospital, China Medical University, Liaoning, China.

References

  1. Mirhashemi A, Saffarshahroudi A, Sodagar A, Atai M. Force-degradation pattern of six different orthodontic elastomeric chains. J Dent (Tehran). 2012;9(4):204–15.

    PubMed  Google Scholar 

  2. Behnaz M, Dalaie K, Hosseinpour S, Namvar F, Kazemi L. The effect of toothpastes with bleaching agents on the force decay of elastomeric orthodontic chains. Eur J Dent. 2017;11:427–31. https://doi.org/10.4103/ejd.ejd_83_17.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Mousavi SM, Mahboubi S, Kalbassi S, Rakhshan V. Force decay of 27 orthodontic elastomeric chains. Iran J Orthod. 2022;17:1–11. https://doi.org/10.22034/IJO.2022.553718.1067.

    Article  Google Scholar 

  4. Moghaddam SF, Kamali MD. Force decay and discoloration of thermoplastic and thermoset orthodontic elastomeric chains. Braz Dent J. 2022. https://doi.org/10.4322/bds.2022.e2612.

    Article  Google Scholar 

  5. Silva Chaves-Filho ACD, Costa AR, Borges LPS, et al. Force degradation of elastomeric chains after storage time and mechanical brushing. Braz Dent J. 2021;32:55–61. https://doi.org/10.1590/0103-6440202104487.

    Article  PubMed  Google Scholar 

  6. Mauro M, Giovana CV, José GN, Vívian FF, Mário VF, Lourenço CS, Ana RC. Evaluation of latex and non-latex intermaxillary elastics strength degradation when submitted to the use of chlorhexidine. Orthod Waves. 2021;80:65–70. https://doi.org/10.1080/13440241.2021.1908749.

    Article  Google Scholar 

  7. Andhare P, Datana S, Agarwal SS, Chopra SS. Comparison of in vivo and in vitro force decay of elastomeric chains/modules: a systematic review and meta analysis. J World Fed Orthod. 2021;10(4):155–62. https://doi.org/10.1016/j.ejwf.2021.07.003.

    Article  PubMed  Google Scholar 

  8. Notaroberto DFC, Martins MME, Goldner MTA, Mendes AM, Quintão CCA. Force decay evaluation of latex and non-latex orthodontic intraoral elastics: in vivo study. Dent Press J Orthod. 2018;23(6):42–7. https://doi.org/10.1590/2177-6709.23.6.042-047.oar.

    Article  Google Scholar 

  9. Ramachandraiah S, Sridharan K, Nishad A, Manjusha KK, Abraham EA, Ramees MM. Force decay characteristics of commonly used elastomeric chains on exposure to various mouth rinses with different alcohol concentration: an in vitro study. J Contemp Dent Pract. 2017;18:813–20.

    PubMed  Google Scholar 

  10. Losito KAB, Lucato AS, Tubel CAM, Correa CA, dos Santos JCB. Force decay in orthodontic elastomeric chains after immersion in disinfection solutions. Braz J Oral Sci. 2014;13:266–9. https://doi.org/10.20396/bjos.v13i4.8640856.

    Article  Google Scholar 

  11. Andreasen GF, Bishara S. Comparison of alastik chains with elastics involved with intra-arch molar to molar forces. Angle Orthod. 1970;40:151–8. https://doi.org/10.1043/0003-3219(1970)040%3c0151:COACWE%3e2.0.CO;2.

    Article  PubMed  Google Scholar 

  12. Kanuru RK, Azaneen M, Narayana V, Kolasani B, Indukuri RR, Babu PF. Comparison of canine retraction by in vivo method using four brands of elastomeric power chain. J Int Soc Prev Commun Dent. 2014;4:S32–7. https://doi.org/10.4103/2231-0762.144586.

    Article  Google Scholar 

  13. Kroczek C, Kula KS, Stewart KT, Baldwin JJ, Fu T, Chen J. Comparison of the orthodontic load systems created with elastomeric power chain to close extraction spaces on different rectangular archwires. Am J Orthod Dentofac Orthop. 2012;141:262–8. https://doi.org/10.1016/j.ajodo.2011.09.007.

    Article  Google Scholar 

  14. Halimi A, Benyahia H, Doukkali A, Azeroual MF, Zaoui F. A systematic review of force decay in orthodontic elastomeric power chains. Int Orthod. 2012;10:223–40. https://doi.org/10.1016/j.ortho.2012.06.013.

    Article  PubMed  Google Scholar 

  15. Barrie WJ, Spence JA. Elastics–their properties and clinical applications in orthodontic fixed appliance therapy. Br J Orthod. 1974;1:167–71. https://doi.org/10.1179/bjo.1.4.167.

    Article  PubMed  Google Scholar 

  16. Yang L, Lv C, Yan F, Feng J. Force degradation of orthodontic latex elastics analyzed in vivo and in vitro. Am J Orthod Dentofac Orthop. 2020;157:313–9. https://doi.org/10.1016/j.ajodo.2019.03.028.

    Article  Google Scholar 

  17. Bishara SE, Andreasen GF. A comparison of time related forces between plastic alastiks and latex elastics. Angle Orthod. 1970;40:319–28. https://doi.org/10.1043/0003-3219(1970)040%3c0319:ACOTRF%3e2.0.CO;2.

    Article  PubMed  Google Scholar 

  18. Behnaz M, Namvar F, Sohrabi S, Parishanian M. Effect of bleaching mouthwash on force decay of orthodontic elastomeric chains. J Contemp Dent Pract. 2018;19:221–5. https://doi.org/10.5005/jp-journals-10024-2240.

    Article  PubMed  Google Scholar 

  19. Brignardello-Petersen R. Nickel-titanium springs probably result in a small increase in orthodontic space closure rate compared with elastomeric power chains, but insufficient or no evidence about relative effects on other outcomes. J Am Dent Assoc. 2018;149:e105. https://doi.org/10.1016/j.adaj.2018.01.039.

    Article  PubMed  Google Scholar 

  20. Lo Giudice A, Portelli M, Militi A, et al. Is static friction affected by aging and amount of elastomeric ligatures in orthodontic sliding mechanics? an in-vitro investigation. J Biol Regul Homeost Agents. 2018;32:67–73.

    PubMed  Google Scholar 

  21. Andhare P, Datana S, Agarwal SS, Chopra SS. Comparison of in vivo and in vitro force decay of elastomeric chains/modules: a systematic review and meta analysis. J World Fed Orthod. 2021;10:155–62. https://doi.org/10.1016/j.ejwf.2021.07.003.

    Article  PubMed  Google Scholar 

  22. Braga E, Souza G, Barretto P, Ferraz C, Pithon M. Experimental evaluation of strength degradation of orthodontic chain elastics immersed in hot beverages. J Indian Orthod Soc. 2019;53(4):244–8. https://doi.org/10.1177/0301574219867540.

    Article  Google Scholar 

  23. Richardson P, Martín-Fabiani I. Competition between crystallization and coalescence during the film formation of poly(chloroprene) latex and effects on mechanical properties. Ind Eng Chem Res. 2019;58:21031–43. https://doi.org/10.1021/acs.iecr.9b02279.

    Article  Google Scholar 

  24. Balhoff DA, Shuldberg M, Hagan JL, Ballard RW, Armbruster PC. Force decay of elastomeric chains—a mechanical design and product comparison study. J Orthod. 2011;38:40–7. https://doi.org/10.1179/14653121141227.

    Article  PubMed  Google Scholar 

  25. Malanon S, Dechkunakorn S, Anuwongnukroh N, Wichai W. Comparison of three commercial latex and non-latex orthodontic elastic bands. Key Eng Mater. 2019;814:354–9. https://doi.org/10.4028/www.scientific.net/KEM.814.354.

    Article  Google Scholar 

  26. Stevenson JS, Kusy RP. Force application and decay characteristics of untreated and treated polyurethane elastomeric chains. Angle Orthod. 1994;64:455–67. https://doi.org/10.1043/0003-3219(1994)064%3c0455:FAADCO%3e2.0.CO;2.

    Article  PubMed  Google Scholar 

  27. Kardach H, Biedziak B, Olszewska A, Golusińska-Kardach E, Sokalski J. The mechanical strength of orthodontic elastomeric memory chains and plastic chains: an in vitro study. Adv Clin Exp Med. 2017;26(3):373–8. https://doi.org/10.17219/acem/61957.

    Article  PubMed  Google Scholar 

  28. Meyns J, Brasil DM, Mazzi-Chaves JF, Politis C, Jacobs R. The clinical outcome of skeletal anchorage in interceptive treatment (in growing patients) for class III malocclusion. Int J Oral Maxillofac Surg. 2018;47:1003–10. https://doi.org/10.1016/j.ijom.2018.04.011.

    Article  PubMed  Google Scholar 

  29. Arumughan S, Somaiah S, Muddaiah S, Shetty B, Reddy G, Roopa S. A comparison of the rate of retraction with low-level laser therapy and conventional retraction technique. Contemp Clin Dent. 2018;9:260–6. https://doi.org/10.4103/ccd.ccd_857_17.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Eliades E, Eliades G, Watts DC. Structural conformation of in vitro and in vivo aged orthodontic elastomeric modules. Eur J Orthod. 1999;21:649–58. https://doi.org/10.1093/ejo/21.6.649.

    Article  PubMed  Google Scholar 

  31. Sang T, Wu J. Factors on force degradation of elastomeric chains in vitro. Shanghai Kou Qiang Yi Xue. 2008;17:638–42.

  32. Menon VV, Madhavan S, Chacko T, Gopalakrishnan S, Jacob J, Parayancode A. Comparative assessment of force decay of the elastomeric chain with the use of various mouth rinses in simulated oral environment: an in vitro study. J Pharm Bioallied Sci. 2019;11:S269–73. https://doi.org/10.4103/JPBS.JPBS_9_19.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Ajami S, Farjood A, Zare M. Synergic effect of salivary pH baselines and low pH intakes on the force relaxation of orthodontic latex elastics. Dent Res J (Isfahan). 2017;14:68–72. https://doi.org/10.4103/1735-3327.201129.

    Article  PubMed  Google Scholar 

  34. Mude N, Chokhani D, Kharche A, Khudare P, Joteppa V, Mohani S. Changes in the force degradation of orthodontic elastomeric chains subjected to different formulations of chlorhexidine mouth rinse: an in vitro study. Int J Health Sci. 2022;6(S1):12572–85. https://doi.org/10.53730/ijhs.v6nS1.8340.

    Article  Google Scholar 

  35. Gioka C, Zinelis S, Eliades T, Eliades G. Orthodontic latex elastics: a force relaxation study. Angle Orthod. 2006;76(3):475–9. https://doi.org/10.1043/0003-3219(2006)076[0475:OLE]2.0.CO;2.

    Article  PubMed  Google Scholar 

  36. Storie DJ, Regennitter F, von Fraunhofer JA. Characteristics of a fluoride-releasing elastomeric chain. Angle Orthod. 1994;64(3):199–210. https://doi.org/10.1043/0003-3219(1994)064%3c0199:COAFEC%3e2.0.CO;2.

    Article  PubMed  Google Scholar 

  37. Boester CH, Johnston LE. A clinical investigation of the concepts of differential and optimal force in canine retraction. Angle Orthod. 1974;44:113–9. https://doi.org/10.1043/0003-3219(1974)044%3c0113:ACIOTC%3e2.0.CO;2.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank teachers and students in the Stomatology of china medical university for their continuous help and support.

Funding

This research was funded by This study was supported by a grant from Youth backbone support program of China medical university (QGZ2018070), and Liaoning province key research and development plan project (2020JH2/10300038).

Author information

Authors and Affiliations

Authors

Contributions

BZ and MAAA contributed to conception and design of study, original draft preparation.; allocation, and editing. BZ, MAAA, YZ and ZL contributed to analysis and/or interpretation of data. YZ and ZL contributed to drafting and revising the manuscript critically for important intellectual content. YL contributed to the critical revision, supervision, and data curation. All authors read and approved the final manuscript to be published.

Corresponding author

Correspondence to Yi Liu.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the ethics committee of the Stomatology Hospital of China Medical University, Liaoning, China (No: CMUKQ-2021-026). All methods were carried out in accordance with the principles of the declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zheng, B., Al-Somairi, M.A.A., Li, Z. et al. Effect of filament types and loops number on the force degradation of elastomeric chains used for orthodontic treatment: an in-vitro study. BMC Oral Health 23, 113 (2023). https://doi.org/10.1186/s12903-023-02812-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-023-02812-7

Keywords