Comparative analysis of anchorage strength and histomorphometric changes after implantation of miniscrews in adults and adolescents: an experimental study in Beagles
BMC Oral Health volume 23, Article number: 639 (2023)
Abstract
Objectives
This study aimed to explore the differences in anchorage strength and histomorphometric changes in orthodontic miniscrews between adult and adolescent beagles.
Material and method
Six adult beagles and six young beagles were used as experimental subjects, and eight miniscrews were symmetrically placed in the posterior mandible of each dog. Measurement of the displacement (mm) of two adjacent miniscrews after load application was performed to compare the anchorage strength between the adult and adolescent groups. Three intravital bone fluorochromes (oxytetracycline, calcein green, xylenol orange) were administered postoperatively to mark the active bone-forming surface. Subsequently, the mineral apposition rate and bone-implant contact ratio were measured for dynamic and static histomorphometry. Finally, the expression levels of the RANKL/OPG ratio were evaluated by immunohistochemistry.
Results
The average displacement of miniscrews in the adult group was significantly less than that in the adolescent group after load application. For histomorphometry analysis, the mineral exposure rate in the adolescent group was higher than that in the adult group with or without force application. In addition, more fractures and new bone formation but deceased bone-implant contact ratios were observed in the adolescent group than in the adult group. The ratio of RANKL/OPG expression increased more in the adolescent group than in the adult group.
Conclusion
Miniscrews do not remain in the same position as skeletal anchors, and the amount of displacement was higher in adolescent group than that in adult group, reflecting the weaker anchorage strength of miniscrews in adolescents due to the higher bone turnover rate and active bone remodelling. Therefore, it is feasible to apply orthodontic loading to the miniscrews in adult patients earlier, even immediately, but it is recommended to wait a period for the adolescents.
Introduction
Anchorage, defined as resistance to undesired tooth movement, is the key management strategy during orthodontic treatment. To enhance anchorage, many mechanical apparatuses have been designed and applied in orthodontics. Among these, orthodontic miniscrews are preferred because of their small size, location flexibility, low cost and convenient operation. Above all, miniscrews provide the stronger skeletal anchorage without the cooperation of patients compared with conventional anchorage reinforcement [1]. Therefore, miniscrews are widely used in various malocclusions, such as maxillary protrusion, mandibular retrusion, anterior deep overjet, and excessive tooth display [2,3,4]. During treatment, miniscrews must keep stable in bone; thus, many researchers have worked to improve osseointegration by imparting biocompatibility to the surface, adequate statistical analysis of anatomical structures, optimization of mechanical properties, enhancement of corrosion resistance and so on [5, 6].
Although the efficacy and safety of miniscrews have been systematically demonstrated, the failure and looseness of miniscrews are still frequently encountered [7,8,9]. Most previous studies reported that the success rate of miniscrews varies between approximately 70% and 90%, which could be influenced by various host and objective factors including bone quality, miniscrew design, placement technique and loading conditions (period, magnitude, direction, etc.) [10,11,12]. One of the important host factors was the age of patients. A recent study that evaluated 889 miniscrews in 347 patients showed that younger people with miniscrews inserted in the retromaxillary or retromandibular regions had a higher progressive susceptibility to failure [13]. Several studies also found that the success rate of miniscrews is lower among patients younger than 20 years [14,15,16]. However, a recent systematic literature review and meta-analysis identified some controversies surrounding the possible association between miniscrew failure and age [17]. Several studies showed no association between age and failure [18, 19]. Moreover, it has been revealed there was a 5% increase in failure risk for every one-year increase in age among participants older than 30 years by a Cox proportional-hazards model [20].
Although many potential risk factors associated with miniscrews are known, loosening of miniscrews still occurs frequently in clinical orthodontic treatment due to the increasing enormous number of patients [21], and the cause of the different failure rates in adolescents compared with adults is still unclear. Many orthodontists proposed that this difference between adults and growing patients in stability may be related to their difference in bone density and cortical bone thickness [22,23,24]. However, most studies provide evidence mainly based on clinical indicators or radiographic image analysis [25, 26]. In this study, we attempted to investigate the difference in anchorage strength and histomorphometric changes between adolescents and adults by an experimental in vivo model to provide a theoretical basis for better clinical application of miniscrews.
Materials and methods
Animal preparation
A total of twelve male beagles were used in this randomized controlled trial study. Six of them were aged 19–20 months (with eruption of the mandibular third molar, weight 13–14 kg) and were defined as the adult group, and the other six that were aged 6.2–6.8 months (mandibular second and third molars without eruption, weight 8-8.5 kg) were defined as the adolescent group. All animal experiments were approved by the Animal Ethics Committee of Provincial Hospital Affiliated with Shandong University (NO.2,018,091), and all experiments were performed in accordance with the Animal Ethics Procedures and Guidelines of the People’s Republic of China. The animals were observed for 2 weeks to confirm that they were healthy and adapted to the new feeding location before the study. During the entire process, the living environment, food and other external factors of the two groups of animals were ensured to be consistent and stable. The experiments were performed according to a timeline (Fig. 1A).
Implantation of miniscrews
All procedures were performed under systemic (30 mg/kg 3% sodium pentobarbital) and local infiltration (2% lidocaine with 0.01% epinephrine) anaesthesia. Eight commercially pure titanium miniscrews (diameter 1.6 mm, screw thread length 6.0 mm, Medicon, Germany) were implanted in each animal through the drilling method, four on each side. The miniscrews were scheduled to be placed in the area of premolars and molars in the mandible based on the anatomic features of beagles. Side (left or right) and site (anterior, middle, or posterior) were alternatively assigned for each dog, and the implant sites were the interroots of P2 and P3, P3 and P4, P4 and M1, and proximal and distal M1 (P = premolars; M = molars), which were defined as C, D, E, and F, respectively (Fig. 1B-C). We implanted the miniscrews at suitable sites with an insertion angle ranging from 60° to 70°, and the height to the alveolar bone crest was as consistent as possible. After surgery, 800,000 units of penicillin were injected intramuscularly every day for 7 consecutive days. Dental X-rays were taken after one week to check the placement (Fig. 1D), and all surgical procedures were performed by the same operator. During the subsequent evaluation, the researchers who were in charge of the data analysis process were not aware of the group allocation to ensure the accuracy of the results.
Evaluation of the anchorage strength of miniscrews
To evaluate the anchorage strength of miniscrews, the distance between miniscrews was measured before and after accepting stretching force. Three weeks after implantation, the forces were applied to the interradicular miniscrews C and D ipsilaterally by a nickel titanium tension spring connecting them (Fig. 1E). All groups were assigned to have constant stretch tension of 150 g bilaterally, and the nickel titanium tension spring was adjusted after reloading to maintain the force at the same level. The distance between them was measured with a Vernier calliper at three time points: 0 w, 3 w, and 10 w. The measuring point was the central point on the surface of the head of the miniscrews with two forces, and the average value was taken 3 times each time by the same operators (Fig. 1F). All loading and measurements were performed by the same operator, and the above procedures and the injection of fluorescent markers were performed under general anaesthesia.
Intravital bone fluorochrome staining
Four spots after implantation were intravenously injected four times within 11 weeks to mark active bone-forming surfaces: oxytetracycline (20 mg/kg) (Terramicin Long-acting, Zoetis, Italy) was injected 1 week after implantation; calcein green (8 mg/kg) was injected 3 weeks after the operation (Sigma C0875); and xylenol orange (120 mg/kg) (Fluka 33,825) was injected 6 weeks after the operation. At 11 weeks after surgery, oxytetracycline (20 mg/kg) (Terramicin Long-acting, Zoetis, Italy) was injected. The in vivo experimental design is shown in Fig. 1.
Specimen preparation
Thirteen weeks after implantation, all animals were sedated and anaesthetized as previously described. The carotid artery was perfused with 4% neutral paraformaldehyde, and the femoral artery was exsanguinated. The mandible containing miniscrews was separated and removed, and then the lower margin of the mandible was trimmed to guarantee that the longitudinal axis of the bone was parallel to the direction of the force. The left specimen was fixed using 4% paraformaldehyde for decalcification, and the right specimen was immersed in 75% alcohol for undecalcified bone processing. After dehydration and infiltration, the mandibular specimen was embedded in poly(methyl methacrylate) (Merck, Schuchardt, Hohenbrunn, Germany). Then, each block was sectioned by using a saw microtome (SP 1600; Leica Instruments, Nussloch, Germany), obtaining a series of Sect. 60 μm in thickness. One unstained section of each specimen was used for vital bone fluorescence labelling observation, and three consecutive sections of the central portion of the miniscrew site were also used for static histomorphometric evaluation after being stained with 1% toluidine blue.
Histologic observation and histomorphometric evaluation
One unstained section for each specimen was placed under a fluorescence microscope (OLYMPUS IX71). Fluorescence images of the tail, middle and neck of the miniscrews were collected, and the mineral apposition rate (MAR, µm/day) was calculated for dynamic histomorphometry with Image Pro-plus software to evaluate the rate of new bone formation. The MAR is the rate of progression of the mineralization front between two consecutive fluorochromes as an index of osteoblast activity. In addition, static histomorphometric evaluations were carried out with Leitz Microvid equipment connected to an IBM XT computer with an image analysis system (Nikon Em120; ACT-1). Image-Pro Plus software was used to perform static bone morphometry at the implant-bone interface. The bone-implant contact ratio (BIC) was calculated by the following formula: BIC (%) = implant surface length in contact with osseous tissue ÷ total surface length of implant × 100%.
Immunohistochemistry staining
Paraffin-embedded sagittal sections of decalcified mandible tissue were sliced 5 μm thick, deparaffinized in xylene two times and dehydrated with gradient concentrations of ethanol. To block endogenous peroxidase activity, the slides were incubated with 3% hydrogen peroxide for 15 min and blocked with 5% BSA for 30 min at room temperature to prevent nonspecific binding. Then, the slides were incubated with the primary antibodies anti-OPG (1:50, ab73400; Abcam) and anti-RANKL (1:200, ab9957; Abcam) at 4 °C overnight. After being immersed in a secondary antibody for 1 h at room temperature, the sections were washed with PBS (pH 7.2–7.6) three times for 5 min each. Then, 3,3’- diaminobenzidine (DAB) (ZLI-9018; Zhongshan Jinqiao Biotechnology Co.) was used for the colour reaction, which was stopped with distilled water, followed by counterstaining with haematoxylin. Finally, dehydration and sealing were performed with neutral gum. Slides were photographed using a light microscope (Olympus, Tokyo, Japan), and the fractional (%) stained area was calculated using Image-Pro Plus software.
Statistical analysis
All in vitro experiments were performed at least three times, and the measurements were conducted twice by the same investigator at an interval of 2 weeks. Reliability tests were applied to ensure the stability, internal consistency and equivalence of the main measurement properties. Independent-sample t tests were used to compare the differences between the adult group and the adolescent group in the displacement of miniscrews, MAR and BIC. Statistical analysis was carried out with the SPSS 10.0 software package. Differences with probabilities less than 5% (P < 0.05) were considered statistically significant.
Results
Evaluation of the anchorage strength of miniscrews
All miniscrews were successfully implanted and no loose or failed screws were found in the healing stage. To evaluate the anchorage strength of the miniscrews, the distance between the miniscrews was measured and recorded after accepting mechanical force (Table S1). Then the displacement of the miniscrews was calculated and is shown in Table 1. The observation variables were independent, normally distributed and the variance between the two groups of observation variables was equal. The observations revealed that the average displacement of the adult group (n = 12) was less than that of the adolescent group (n = 12) at the three loading time points, and the difference was statistically significant (P < 0.01). Furthermore, the displacement of the miniscrews in the first 3 weeks was greater than that in the last 7 weeks, which indicated that the displacement of miniscrews mainly occurred in the first 3 weeks.
Evaluation of bone fluorochrome staining
Fluorescent labelling demonstrated the development of newly formed bone. There were three fluorescent labels in the middle part of the miniscrew (Fig. 2A). In unloading screws, we could see three fluorescent bands: green near the margin, red and yellow outwards. In loading screws, we could see fluorescent expression around the implant, but there were no obvious bands. Moreover, there was more obvious bone deposition on side B than on side A (A was the side with the same direction of force applied to the miniscrew, and side B was the other side). The bone mineral appositional rate (MAR) was determined by measuring the distance between the centres of the two fluorescent bands divided by the time of labelling, which reflected how fast the mineralization front proceeded during bone remodelling. The results showed that the MAR in the adolescent group was higher than that in the adult group with or without spring application (Fig. 2B), indicating that the speed of bone reconstruction around unstrained miniscrews was faster in the adolescent group than in the adult group.
Observation of the miniscrew-bone interface
Of the 96 miniscrews enrolled, the BIC percentage ranged from 32.37 to 66.85%. The means and SDs of the BIC values are presented in Table S2. The staining of undecalcified sections (Fig. 3A) revealed that the miniscrews bound well to the bone in the unloading groups, and fractures and new bone formation were observed in the adolescent group when compared with the adult group. Additionally, more osteoids were formed in the space between the screws and bone in the loading groups of adolescents. For quantitative analysis, the results showed that the load did not affect the respective BIC in the adults (p = 0.177), but a decreased BIC was observed in the adolescents. Both the loaded and unloaded miniscrews showed significant differences between the adult group and the adolescent group in their respective BIC values (Fig. 3B).
Haematoxylin-eosin staining and immunohistochemistry
The results of HE staining (Fig. 4A) revealed osteoblasts and osteoclasts in the peripheral bone of the miniscrews. Specifically, the osteoclasts of the juvenile beagles were stronger than those of the adult beagles in both the unloading and loading groups, which means that osteogenesis in juvenile beagles was better than that in adult beagles. The purpose of immunohistochemistry staining was to detect the expression of OPG and RANKL protein in the peri-screw bone area (Fig. 4B). The expression of OPG in bone around the miniscrews decreased after 10 weeks of force application in both groups compared with the unloading condition, and the reduction in side A of the applied miniscrews in the adolescent group decreased the most. The expression of RANKL was increased more in the adolescent group than in the adult group, and the increase was the largest on the A side of the miniscrews. The mean integrated optical density values are shown in Table S3. The ratio of RANKL/OPG expression increased in both groups; the increase on side A was the largest in the adolescent group, and the increase on side B was the smallest in the adolescent group. The results showed that the bone resorption on the A side of the adolescent miniscrews was the highest, and the bone resorption on the B side of the adolescent miniscrews was the lowest.
Discussion
Considerable interest in miniscrews has grown because of the strong anchorage control they provide for executing hitherto impossible tooth movement[27]. However, loosening of miniscrews still occurs frequently in daily clinical practice due to the enormous number of patients, especially younger patients[13]. The present study was designed innovatively to further investigate the differences in anchorage stability of miniscrews influenced by age-related factors from the perspective of histology.
Orthodontic treatment in adolescents and adults displays multiple differences, such as duration, tooth movement parameters, accidental periodontal adverse effects, biological response and growth potential, which are of great concern for orthodontists[28,29,30]. There is a reduced rate of tooth movement in adults along with more pain and discomfort than in adolescents, and the discrepancy might be associated with the fact that the bone characteristics of adults differ from those of adolescents[31]. At the same time, the primary stability of miniscrews depends on bone characteristics and other factors such as implant geometric design (length, diameter, thread type, thread shape, thread pitch, thread design), and surgical technique-related factors (cortical predrilling insertion angle) [32,33,34]. Therefore, bone characteristics tend to be a prerequisite for differences in the outcomes of the two groups.
Some previous studies considered miniscrews to be an absolute anchorage, that could remain nearly immobile, or at least there was no clinically significant displacement compared with tooth movement [35, 36]. A growing number of studies have begun to focus on the displacement of miniscrews, and the displacement of the miniscrews were used to reflect the mechanical retention force [37]. Orthodontic loading-induced displacement distinguishes between primary and secondary displacement, it has been proposed that the sum of displacements should not exceed 1 mm[38]. In another study, the total displacement (mean value ≤ 0.78 mm) of forty-one miniscrews were measured submitted to force after a 5 months period[39]. These results are basically consistent with our study, the mean maximum displacement during 10 weeks was 0.625 mm observed in adolescent group.
Primary displacement refers to the immediate displacement subjected to functional load due to the elastic and plastic properties of the bones, and secondary displacement is defined as the long-term displacement of miniscrews due to bone remodelling processes[36]. There is a lower modulus of elasticity, bending strength, and bone density in bone of the younger, all of which contribute to the bone bending tendency under mechanical loading[26, 40, 41]. The initial retention of miniscrews also rely on mechanical locking provided by the adequate cortical bone thickness and bone mineral density [42, 43]. Data observed from a clinical point of view has suggested that the greater the cortical thickness, the lower will be the miniscrew displacement[38]. Studies have shown that adolescents presented with significantly thinner alveolar cortical bone thickness and lower cancellous bone density at different levels from the alveolar crest than adults[44, 45], implying that mechanical latches from the alveolar bone tend to be a weaker support in a growing population. This is one of the main reasons why the average displacement of miniscrews in the adult group was obviously shorter than that in the adolescent group, indicating the stability of miniscrew anchorage was compromised in adolescents with thinner cortical layers. After all, greater physiological anchorage loss also occurs in adolescents than in adults[46].
It has been demonstrated that a 1 mm zone of devitalized bone develops quickly around the miniscrews, particularly in the cortical region, due to surgical trauma and stress necrosis [47]. The devitalized interface cortical healing relies on continuous dynamic bone remodelling in cancellous bone to ensure lasting direct bone-to-miniscrew anchorage, which is associated with secondary displacement tightly. Accumulated researches have showed the influence of age on bone turnover marker and bone remodelling. Higher concentrations of bone turnover and metabolic indicators were observed in younger individuals[25], indicating that there would also be more secondary displacement in adolescent group. Besides, we implanted the miniscrews with an insertion angle ranging from 60° to 70° based on the previous suggestions[48], while a finite element analysis demonstrated that insertion of miniscrews at angles less than or greater than 90 degrees to the alveolar process bone might decrease the anchorage stability of the miniscrew[49]. The immediate compression force increased the stress concentration at cortical bone, the angular acceleration resulted in a composite motion combining more tipping movement at the initial stage. It is even more of a challenge for bone quality put forwards by the increased torque. Crucially, the stability is at the lowest level during the third week, just at the end of the resorptive phase, especially in sites with thin cortical layers[50, 51]. All the above mentioned is aligned with our results that the displacement of both groups tended to plateau in the first 3 weeks and decrease thereafter in the following weeks.
The bone dynamics adjacent to miniscrews are thought to be active to maintain a vital bone-implant interface, which is improved by both direct bone deposition and mineral tissue integration[52]. Histomorphometry is a useful technique for evaluating the MAR and cellular events. We employed the BIC and MAR as parameters for static and dynamic histomorphometric analysis, respectively. After implantation, not only mineralized bone tissue contact but also osteoblasts firmly attached to the miniature screws in the direct-contact area, with gaps of hundreds of microns in other areas where the recruiting osteogenic cells subsequently become stationary osteoblasts to secrete the osteoid matrix, followed by mineralization to form irregular woven bone that encroaches on the miniscrew surface[53, 54]. The BIC represents the amount of bone directly in contact with the miniscrew surface, and the MAR represents the rate of progression of the mineralization front between two consecutive fluorochromes as an index of osteoblast activity[55, 56]. An increase in the BIC and MAR would be desirable as the bone undergoes modelling and remodelling. In our study, the BIC of all miniscrews in adults and adolescents ranged from 32.37 to 66.85%, although both groups showed good osseointegration. It is still unclear whether a 60% BIC is sufficient and better for miniscrew stability and service than a 32% BIC; however, a higher BIC in adults better indicates the state of adaptation at the interface. Nevertheless, the MAR in adolescents was faster than that in adults. It is evident from the results that bone formative responses are higher in adolescents than in adults, while higher rates of osteoclast differentiation were also coupled with significantly increased bone turnover activity in the younger population[28]. That is why a faster MAR but lower BIC presents in adolescents, and as time goes on, their BIC would be similar owing to the faster MAR in adolescents than in adults. The higher bone turnover rate in the adolescent group was a double-edged sword, which complicated the achievement of optimal mechanical stability, leading to faster bone modelling and remodelling but weaker anchorage strength of miniscrews within a short time, and the stability of the miniscrews in adults seemed more reasonable.
In addition, the immune-expression patterns of RANKL and OPG also emphasized the difference between adolescents and adults. OPG is expressed by osteoblasts, which might prevent osteoclasis by binding to the key osteoclastogenic cytokine RANKL and is pro-osteoblastic in bone mineralization and calcium ion homeostasis[57]. The ratio of RANKL/OPG is a determinant in regulating the activation and function of osteoclasts[58]. Slower bone turnover progress as well as maintenance of the RANKL/OPG ratio were observed in adults, which further confirmed the anchorage strength results.
The major limitation of this research was the single design. We adopted one type of miniscrew, the same insertion method, force application, and placement sites to reduce other interference factors. Thus, it is difficult to illustrate and summarize the problems in all situations. As far as the present study is concerned, it is a pilot and exploratory study, we preliminarily excavated the anchorage strength and histomorphometric changes distinguishing the susceptibility to failure of orthodontic miniscrews between adolescent and adult groups.
Conclusion
Miniscrews do not remain in the same position as skeletal anchors, and the amount of displacement was higher in adolescent group than that in adult group, reflecting the weaker anchorage strength of miniscrews in adolescents due to the higher bone turnover rate and active bone remodelling. Therefore, it is possible to apply orthodontic loading to the miniscrews in adult patients earlier, even immediately, but it is recommended to wait a period of time for the adolescents.
Data Availability
The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.
References
Liu Y, Yang ZJ, Zhou J, Xiong P, Wang Q, Yang Y, Hu Y, Hu JT. Comparison of Anchorage Efficiency of Orthodontic Mini-implant and Conventional Anchorage reinforcement in patients requiring Maximum Orthodontic Anchorage: a systematic review and Meta-analysis. J Evid Based Dent Pract. 2020;20:101401.
Babanouri N, Ajami S, Salehi P. Effect of miniscrew-facilitated micro-osteoperforation on the rate of orthodontic tooth movement: a single-center, split-mouth, randomized, controlled trial. Prog Orthod. 2020;21:7.
Solano Mendoza P, Aceytuno Poch P, Solano Reina E. B. Solano Mendoza, Skeletal, Dentoalveolar and Dental Changes after “Miniscrew Assisted Rapid Palatal Expansion” Evaluated with Cone Beam Computed Tomography, J Clin Med 11 (2022).
Nienkemper M, Wilmes B, Pauls A, Drescher D. Multipurpose use of orthodontic mini-implants to achieve different treatment goals. J Orofac Orthop. 2012;73:467–76.
Im C, Park JH, Jeon YM, Kim JG, Jang YS, Lee MH, Jeon WY, Kim JM, Bae TS. Improvement of osseointegration of Ti-6Al-4V ELI alloy orthodontic miniscrews through anodization, cyclic pre-calcification, and heat treatments. Prog Orthod. 2022;23:11.
Kang HK, Chu TM, Dechow P, Stewart K, Kyung HM, Liu SS. Laser-treated stainless steel miniscrew implants: 3D surface roughness, bone-implant contact, and fracture resistance analysis. Eur J Orthod. 2016;38:154–62.
Alharbi F, Almuzian M, Bearn D. Miniscrews failure rate in orthodontics: systematic review and meta-analysis. Eur J Orthod. 2018;40:519–30.
Wang K, Fan H, Yang H, Li J, Xie W. Efficacy and safety of micro-implant anchorage in Angle class II malocclusion orthodontic treatment: a protocol for systematic review and meta-analysis. Med (Baltim). 2020;99:e23221.
Gurdan Z, Szalma J. Evaluation of the success and complication rates of self-drilling orthodontic mini-implants. Niger J Clin Pract. 2018;21:546–52.
Kaaouara Y, Sara EA, Rerhrhaye W. Perception of miniscrew anchorage devices by patients. Int Orthod. 2018;16:676–83.
Reynders R, Ronchi L, Bipat S. Mini-implants in orthodontics: a systematic review of the literature, Am J Orthod Dentofacial Orthop 135 (2009) 564 e561-519; discussion 564–565.
Arqub SA, Gandhi V, Mehta S, Palo L, Upadhyay M, Yadav S. Survival estimates and risk factors for failure of palatal and buccal mini-implants. Angle Orthod. 2021;91:756–63.
Xin Y, Wu Y, Chen C, Wang C, Zhao L. Miniscrews for orthodontic anchorage: analysis of risk factors correlated with the progressive susceptibility to failure. Am J Orthod Dentofacial Orthop. 2022;162:e192–e202.
Dalessandri D, Salgarello S, Dalessandri M, Lazzaroni E, Piancino M, Paganelli C, Maiorana C, Santoro F. Determinants for success rates of temporary anchorage devices in orthodontics: a meta-analysis (n > 50). Eur J Orthod. 2014;36:303–13.
Tsui WK, Chua HD, Cheung LK. Bone anchor systems for orthodontic application: a systematic review. Int J Oral Maxillofac Surg. 2012;41:1427–38.
Lee SJ, Ahn SJ, Lee JW, Kim SH, Kim TW. Survival analysis of orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2010;137:194–9.
Casana-Ruiz MD, Bellot-Arcis C, Paredes-Gallardo V, Garcia-Sanz V, Almerich-Silla JM, Montiel-Company JM. Risk factors for orthodontic mini-implants in skeletal anchorage biological stability: a systematic literature review and meta-analysis. Sci Rep. 2020;10:5848.
Moon CH, Park HK, Nam JS, Im JS, Baek SH. Relationship between vertical skeletal pattern and success rate of orthodontic mini-implants. Am J Orthod Dentofacial Orthop. 2010;138:51–7.
Chen YJ, Chang HH, Lin HY, Lai EH, Hung HC, Yao CC. Stability of miniplates and miniscrews used for orthodontic anchorage: experience with 492 temporary anchorage devices. Clin Oral Implants Res. 2008;19:1188–96.
Tsai CC, Chang HP, Pan CY, Chou ST, Tseng YC. A prospective study of factors associated with orthodontic mini-implant survival. J Oral Sci. 2016;58:515–21.
Ramirez-Ossa DM, Escobar-Correa N, Ramirez-Bustamante MA, Agudelo-Suarez AA. An Umbrella Review of the effectiveness of Temporary Anchorage Devices and the factors that contribute to their success or failure. J Evid Based Dent Pract. 2020;20:101402.
Lee Y, Choi SH, Yu HS, Erenebat T, Liu J, Cha JY. Stability and success rate of dual-thread miniscrews. Angle Orthod. 2021;91:509–14.
Wehrbein H. Bone quality in the midpalate for temporary anchorage devices. Clin Oral Implants Res. 2009;20:45–9.
Jing Z, Wu Y, Jiang W, Zhao L, Jing D, Zhang N, Cao X, Xu Z, Zhao Z. Factors affecting the clinical success rate of Miniscrew Implants for Orthodontic Treatment. Int J Oral Maxillofac Implants. 2016;31:835–41.
Shao J, Zhou SS, Qu Y, Liang BB, Yu QH, Wu J. Correlation between bone turnover and metabolic markers with age and gender: a cross-sectional study of hospital information system data. BMC Musculoskelet Disord. 2020;21:603.
Zheng Y, Zhu C, Zhu M, Lei L. Difference in the alveolar bone remodelling between the adolescents and adults during upper incisor retraction: a retrospective study. Sci Rep. 2022;12:9161.
Mallick S, Murali PS, Kuttappa MN, Shetty P, Nair A. Optimal sites for mini-implant insertion in the lingual or palatal alveolar cortical bone as assessed by cone beam computed tomography in South indian population. Orthod Craniofac Res. 2021;24:121–9.
Abbing A, Koretsi V, Eliades T, Papageorgiou SN. Duration of orthodontic treatment with fixed appliances in adolescents and adults: a systematic review with meta-analysis. Prog Orthod. 2020;21:37.
Seidel CL, Gerlach RG, Weider M, Wolfel T, Schwarz V, Strobel A, Schmetzer H, Bogdan C, Golz L. Influence of probiotics on the periodontium, the oral microbiota and the immune response during orthodontic treatment in adolescent and adult patients (ProMB Trial): study protocol for a prospective, double-blind, controlled, randomized clinical trial. BMC Oral Health. 2022;22:148.
George P, George JK, Krishnan V, Vijayaraghavan N, Rajendran SR, Chandran BM, Thulasidharan UM. Periodontal ligament cells in adolescents and adults: genetic level responses to orthodontic forces. Am J Orthod Dentofacial Orthop. 2020;158:816–23.
Chang CJ, Lin WC, Chen MY, Chang HC. Evaluation of total bone and cortical bone thickness of the palate for temporary anchorage device insertion. J Dent Sci. 2021;16:636–42.
Sreenivasagan S, Subramanian AK, Nivethigaa B. Assessment of insertion torque of mini-implant and its correlation with Primary Stability and Pain levels in Orthodontic Patients. J Contemp Dent Pract. 2021;22:84–8.
Singh J, Singh SK, Gupta AR, Nayak SC, Vatsa R, Priyadarshni P. Comparative evaluation of Primary Stability of two different types of Orthodontic Mini-Implants. J Pharm Bioallied Sci. 2021;13:128–S131.
Lioubavina-Hack N, Lang NP, Karring T. Significance of primary stability for osseointegration of dental implants. Clin Oral Implants Res. 2006;17:244–50.
Chen G, Chen S, Zhang XY, Jiang RP, Liu Y, Shi FH, Xu TM. A new method to evaluate the positional stability of a self-drilling miniscrew. Orthod Craniofac Res. 2015;18:125–33.
Nienkemper M, Handschel J, Drescher D. Systematic review of mini-implant displacement under orthodontic loading. Int J Oral Sci. 2014;6:1–6.
Miyawaki S, Tomonari H, Yagi T, Kuninori T, Oga Y, Kikuchi M. Development of a novel spike-like auxiliary skeletal anchorage device to enhance miniscrew stability. Am J Orthod Dentofacial Orthop. 2015;148:338–44.
Migliorati M, De Mari A, Annarumma F, Aghazada H, Battista G, Campobasso A, Menini M, Lo Giudice A, Cevidanes LHS, Drago S. Three-dimensional analysis of miniscrew position changes during bone-borne expansion in young and late adolescent patients. Prog Orthod. 2023;24:20.
Alves M Jr, Baratieri C, Nojima LI. Assessment of mini-implant displacement using cone beam computed tomography. Clin Oral Implants Res. 2011;22:1151–6.
Ohiomoba H, Sonis A, Yansane A, Friedland B. Quantitative evaluation of maxillary alveolar cortical bone thickness and density using computed tomography imaging. Am J Orthod Dentofacial Orthop. 2017;151:82–91.
Grimm FM. Bone bending, a feature of orthodontic tooth movement. Am J Orthod. 1972;62:384–93.
Feng Y, Kong WD, Cen WJ, Zhou XZ, Zhang W, Li QT, Guo HY, Yu JW. Finite element analysis of the effect of power arm locations on tooth movement in extraction space closure with miniscrew anchorage in customized lingual orthodontic treatment. Am J Orthod Dentofacial Orthop. 2019;156:210–9.
Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003;124:373–8.
Han S, Bayome M, Lee J, Lee YJ, Song HH, Kook YA. Evaluation of palatal bone density in adults and adolescents for application of skeletal anchorage devices. Angle Orthod. 2012;82:625–31.
Cassetta M, Sofan AA, Altieri F, Barbato E. Evaluation of alveolar cortical bone thickness and density for orthodontic mini-implant placement. J Clin Exp Dent. 2013;5:e245–252.
Ruan MJ, Chen G, Xu TM. Comparison of orthodontic tooth movement between adolescents and adults based on implant superimposition. PLoS ONE. 2018;13:e0197281.
Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52:155–70.
Zhao L, Xu Z, Wei X, Zhao Z, Yang Z, Zhang L, Li J, Tang T. Effect of placement angle on the stability of loaded titanium microscrews: a microcomputed tomographic and biomechanical analysis. Am J Orthod Dentofacial Orthop. 2011;139:628–35.
Perillo L, Jamilian A, Shafieyoon A, Karimi H, Cozzani M. Finite element analysis of miniscrew placement in mandibular alveolar bone with varied angulations. Eur J Orthod. 2015;37:56–9.
Ure DS, Oliver DR, Kim KB, Melo AC, Buschang PH. Stability changes of miniscrew implants over time. Angle Orthod. 2011;81:994–1000.
Wei X, Zhao L, Xu Z, Tang T, Zhao Z. Effects of cortical bone thickness at different healing times on microscrew stability. Angle Orthod. 2011;81:760–6.
Garetto LP, Chen J, Parr JA, Roberts WE. Remodelling dynamics of bone supporting rigidly fixed titanium implants: a histomorphometric comparison in four species including humans. Implant Dent. 1995;4:235–43.
Davies JE. Mechanisms of endosseous integration. Int J Prosthodont. 1998;11:391–401.
Insua A, Monje A, Wang HL, Miron RJ. Basis of bone metabolism around dental implants during osseointegration and peri-implant bone loss. J Biomed Mater Res A. 2017;105:2075–89.
Heriveaux Y, Le Cann S, Fraulob M, Vennat E, Nguyen VH, Haiat G. Mechanical micromodeling of stress-shielding at the bone-implant interphase under shear loading, Med Biol Eng Comput (2022).
Rutten S, Nolte PA, Korstjens CM, van Duin MA, Klein-Nulend J. Low-intensity pulsed ultrasound increases bone volume, osteoid thickness and mineral apposition rate in the area of fracture healing in patients with a delayed union of the osteotomized fibula. Bone. 2008;43:348–54.
Theodoridis C, Doulkeridou C, Menexes G, Vouros I. Comparison of RANKL and OPG levels in peri-implant crevicular fluid between healthy and diseased peri-implant tissues. A systematic review and meta-analysis. Clin Oral Investig. 2022;26:823–36.
Marcadet L, Bouredji Z, Argaw A, Frenette J. The roles of RANK/RANKL/OPG in Cardiac, skeletal, and smooth muscles in Health and Disease. Front Cell Dev Biol. 2022;10:903657.
Acknowledgements
Not applicable.
Funding
Not applicable.
Author information
Authors and Affiliations
Contributions
Zhiqiang Wang performed the examination and analyzed the results, Tingting Jia contributed to data curation, and Yi Zhao was a major contributor in original draft preparation, manuscript review and editing. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Ethics approval and consent to participate
All animal experiments were approved by the Animal Ethics Committee of Provincial Hospital Affiliated to Shandong University, and all experiments were performed in accordance with the ARRIVE guidelines (https://arriveguidelines.org) and Animal Ethics Procedures and Guidelines of the People’s Republic of China.
Consent for publication
Not applicable.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Zhao, Y., Jia, T. & Wang, Z. Comparative analysis of anchorage strength and histomorphometric changes after implantation of miniscrews in adults and adolescents: an experimental study in Beagles. BMC Oral Health 23, 639 (2023). https://doi.org/10.1186/s12903-023-03318-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12903-023-03318-y