Do metal artifact reduction algorithms influence the detection of implant-related injuries to the inferior alveolar canal in CBCT images?

Abstract

Background

The routine application of dental implants for replacing missing teeth has revolutionized restorative and prosthetic dentistry. However, cone beam computed tomography (CBCT) evaluations of structures adjacent to the implants are limited by metal artifacts. There are several methods for reducing metal artifacts, but this remains a challenging task. This study aimed to examine the effectiveness of metal artifact reduction (MAR) algorithms in identifying injuries of implants to the inferior alveolar canal in CBCT images.

Method

In this in vitro study, mono-cortical bone windows were created and the inferior alveolar canal was revealed. Using 36 implants, pilot drill and penetration damage of the implant tip into the canal was simulated and compared to the control implants with distance from the canal. CBCT images were evaluated by four experienced observers with and without the MAR algorithm and compared to direct vision as the gold standard. The values of accuracy, sensitivity, and specificity were obtained and compared by receiver operating characteristic (ROC) curve (α = 0.05).

Result

The area under the ROC curve values for detection of pilot drill injuries varied between 0.840–0.917 and 0.639–0.854 in the active and inactive MAR conditions, respectively. The increase in ROC area was only significant for one of the observers (P = 0.010). For diagnosing penetrative injuries, the area under the ROC curve values was between 0.990–1.000 and 0.722–1.000 in the active and inactive MAR conditions, respectively. The improvement of ROC curve values in active MAR mode was only significant for one of the observers (P = 0.006).

Conclusion

Activation of MAR improved the diagnostic values of CBCT images in detecting both types of implant-related injuries to the inferior alveolar canal. However, for most observers, this increase was not statistically significant.

Peer Review reports

Introduction

The routine application of dental implants for replacing missing teeth has revolutionized restorative and prosthetic dentistry. In recent years, there has been an upsurge in demand for dental implants used for prosthetic and aesthetic purposes [1, 2]. Radiology is a frequently employed diagnostic modality in diverse phases of implant therapy encompassing pre-treatment, treatment strategy, implant insertion, and postoperative monitoring [3]. While three-dimensional imaging using cone beam computed tomography (CBCT) is among the most commonly used imaging modalities for the presurgical evaluation and monitoring of symptomatic implants, it is limited by its ability to detect details in areas adjacent to metallic objects or objects with a high atomic number as a result of metal artifacts [4,5,6].

Although several methods and models have been proposed to reduce these artifacts, the successful elimination of metal-induced artifacts continues to pose a considerable challenge, and the efficacy of the proposed techniques remains a topic of debate [7, 8]. Metal artifact reduction (MAR) algorithms have been developed to repair the gray values altered as a result of artifacts. The effectiveness of MAR algorithms has been studied concerning different metallic or high atomic number materials, from gutta-percha and stainless steel to zirconia, and for different diagnostic tasks, including detection of caries and dental fractures, among other applications [9,10,11]. The effectiveness of these MAR algorithms has proven to be task-specific [12]. To our knowledge, only one previous research has been found regarding the artifact reduction algorithm’s efficacy in identifying injuries to the inferior alveolar canal, which showed a negative effect of the algorithm [13]. Given the importance of detecting these types of injuries to the inferior alveolar canal and the result being task-specific, this study aimed to examine the effectiveness of the native metal artifact reduction algorithm of Sirona CBCT scanner in identifying injuries of implants to the inferior alveolar canal in CBCT images.

Methods and materials

The present study was carried out in Department of Oral and Maxillofacial Radiology, Isfahan University of Medical Sciences, Iran. The Research Ethics Committee at Isfahan University of Medical Sciences has approved this study (#IR.MUI.REC.1400.072, approval date: 03/08/2022). Based on the correlation sample size formula the sample size for each group was 12 implants for CI = 95%, α = 0.05 to determine a difference of 0.25 between the control and experimental groups.

Preparation of specimens

Twelve fresh hemimandibles of sheep (slaughtered on the previous day) were carefully selected, and the soft tissue was completely removed. In order to determine the condition and position of the inferior alveolar canal, an initial radiograph of these hemimandibles was taken with a phosphor plate size 4 (Durr Dental, Bietigheim-Bissingen, Germany). These initial radiographs allowed for screening of the condition of the inferior alveolar canal prior to surgical procedure.

Surgical procedures and implant insertion

Monocortical bone windows were created from the lingual side, which allowed for the inferior alveolar canal to be revealed and directly visualized. 36 titanium dental implants (Bionic, Nik Kasht Asia, Tehran, Iran) with sizes 4 × 10 mm and 4 × 12 mm were placed in the mandibles according to the type of injuries. Two types of inferior alveolar nerve injuries were simulated in this study: pilot drill damage and implant tip penetration into the canal. Twelve implants were simulated for each group. In the pilot drill injury, the upper border of the inferior alveolar canal was penetrated by the pilot drill, but the implant tip was placed about 1 mm above the injury site. With implant penetration damage, drilling was done in such a way that the final position of the implant tip was about 1 mm inside the inferior alveolar canal. Twelve implants were placed as a control group 1 mm above the upper border of the canal (Fig. 1). After that, the window was closed again and the hemimandibles were prepared for the imaging stage.

Fig. 1
figure 1

Illustration depicting preparation of sheep mandibles and insertion of dental implants

Preparing the imaging phantom

The bone segment containing the implants was cut from the hemimandibles using a handpiece. A cranium model (Anatokala, Tehran, Iran) with mandible and maxilla in occlusion was used as an imaging phantom. Bone blocks containing the implants were fixed to the imaging phantom with glue. Soft tissue was replicated by 10 mm of base plate wax (Polywax, Izmir, Turkey) (Fig. 2).

Fig. 2
figure 2

– Illustration depicting preparation of the imaging phantom using cut sheep mandible bone blocks containing implants

Preparation of CBCT images

CBCT images were taken by placing the prepared imaging phantom in Galileos scanner (Sirona, Bensheim, Germany) with exposure conditions of 85 kVp, 21 mAs, voxel size 280 micrometers and field of view 15 × 15 cm. For each sample, two scans were performed: with and without activation of MAR algorithm. CBCT images were viewed and evaluated in Sidexis 4 software (Sirona, Bensheim, Germany).

Observers and reading sessions

Evaluation of the images was done by 4 observers: one oral and maxillofacial radiologist, two oral and maxillofacial surgeon, and one dentist all with more than 5 years of experience in analysis of CBCT images. A calibration session was held for the observers discussing about three cases unrelated to the study. Evaluation of the images was done in a semi-dark and quiet room. The images in each set (off or on MAR) were provided to each observer on two separate reading sessions with a 2-week interval. The observers were unaware of the presence or absence of damage in the images as well as if the image set they are viewing is with or without the activation of MAR. They also had no contact with each other in the reading sessions. The observers were free to choose the desired views and visual settings for diagnosis (Fig. 3). The observers’ responses were recorded by a 4-point Likert scale: 1 control, 2 pilot drill, 3 implant penetration, 4 uncertain. Uncertain cases were excluded from the analysis.

Fig. 3
figure 3

CBCT images of (a) control implants (b) implants with penetration injury (left on the image) and pilot drill injury (right on the image)

Statistical analysis

Statistical analysis was performed using Statistical Package for the Social Sciences (SPSS, version 25, Armonk, NY, USA). Intra- and inter-observer agreements were calculated using Cohen’s kappa. Sensitivity, specificity, and area under the receiver operating characteristic (ROC) curve were calculated for each observer. P < 0.05 was considered statistically significant.

Results

Intraobserver and intraobserver agreements ranged from 0.78 to 0.94 and 0.88 to 0.94, respectively indicating excellent agreements.

Figures 4 and 5 depict the ROC curves of CBCT images with and without activation of MAR for detection of pilot drill and penetrative injuries of the inferior alveolar canal, respectively. The area under ROC curve values for detection of pilot drill injuries ranged between 0.840 and 0.917 in the active MAR and between 0.639 and 0.854 in the inactive MAR conditions. For all observers, activation of MAR led to an increase in the area under the ROC curve when detecting pilot drill injuries to the inferior alveolar canal. However, this difference was not statistically significant (P > 0.05), except for one of the observers (P = 0.010) (Table 1). For diagnosing penetrative injuries, the area under ROC curve values varied between 0.990 and 1.000 in the active MAR and between 0.722 and 1.000 in the inactive MAR conditions. A similar trend was observed for the detection of penetrative injuries to the inferior alveolar canal, where the area under the ROC curve values was only significantly increased by activation of MAR for one observer (P = 0.006) and not significantly different for the other three observers (P > 0.05) (Table 2).

Fig. 4
figure 4

– The receiver operating characteristic (ROC) curves for detection of pilot drill injuries to the inferior alveolar canal with activation of MAR (a) and without activation of MAR (b)

Fig. 5
figure 5

The receiver operating characteristic (ROC) curves for detection of penetrative injuries to the inferior alveolar canal with activation of MAR (a) and without activation of MAR (b)

Table 1 Diagnostic values of CBCT images with/without activation of MAR for detecting pilot drill injuries to the inferior alveolar canal
Table 2 Diagnostic values of CBCT images with/without activation of MAR for detecting penetrative injuries to the inferior alveolar canal

Discussion

For detecting implant-related injuries to the inferior alveolar canal, although activation of MAR improved the diagnostic values of CBCT images, this increase was not statistically significant for most observers.

Studies have shown that as high as 13% of implants placed in the posterior mandible can cause injury to the inferior alveolar nerve [14,15,16]. Mechanical, chemical, and thermal elements can cause injury to the inferior alveolar nerve during or after implant insertion. Direct mechanical trauma from implant drills, implant tip, and bone debris are frequent and can result in pressure on the nerve, its entrapment, or even its rupture [17]. The occurrence of such structural injuries has the potential to give rise to clinical complaints such as anesthesia, hypoesthesia, dysesthesia, or pain. If these injuries are permanent, they affect all aspects of life, reduce the quality of life, and can cause legal problems for dental practitioners [18]. Hence, it is imperative to ensure precise detection of injuries to the inferior alveolar nerve and identify the associated causes and iatrogenic factors.

In this context, CBCT images are helpful in diagnosing complications arising from symptomatic implants. However, as metallic objects with high atomic numbers, these implants produce artifacts including photon starvation, cupping, and beam hardening streaks [19]. The presence of these artifacts can alter CBCT gray values, particularly in the areas immediately adjacent to dental implants, and make the diagnosis of implant complications, such as fenestration, perforation of bony borders, and bone loss challenging. MAR algorithms have been developed to repair these altered gray values. They are usually categorized into five different approaches: the physical modeling-based approach, the projection completion-based approach, the dual energy-based approach, the iterative reconstruction-based approach, and the deep learning-based approach [20].

In the study of de Freitas et al. the influence of metal artifact reduction on the diagnosis of contact between implant and mandibular canal was investigated. They showed that MAR algorithm has a negative effect on the diagnosis. This is in contrast to our findings and may be attributed to the difference in algorithm, the CBCT device, or the difference in implant type [13]. Salemi et al. have investigated the effects of MAR algorithms on the detection of fenestration and dehiscence adjacent to titanium dental implants. Their findings suggested that MAR algorithms did not improve the diagnosis of these bony defects. In fact, accuracy, sensitivity, and specificity decreased in MAR-activated images [21]. The same trend was also observed in the study of Sheikhi et al. in which sensitivity and accuracy were higher in off MAR condition for both fenestration and dehiscence [22]. In the present study, however, the values of area under curve, sensitivity, and specificity generally improved with activation of MAR, although this improvement was not statistically significant for most observers. This inconsistency may be attributed to different diagnostic tasks, as well as different algorithms and CBCT scanners. Salemi et al. used Planmeca Promax 3D and Soredex Cranex 3D CBCT scanners. In the study of Sheikhi et al., similar to our study, Sirona Galileos CBCT unit was used. Moreover, our study applied more observers. In another study, Bagis et al. reported that MAR algorithm improved the accuracy of detection of peri-implant fenestrations. Additionally, MAR algorithm was more effective in enhancing the diagnosis for titanium implants compared to zirconium ones [23]. The implants used in the present study were titanium grade 5 alloys with 6 wt% aluminum and 4% vanadium. The effect of MAR on reducing artifacts arising from different implant materials and alloys is another topic worth further exploring.

Previous studies have reported observer variability for diagnosing implant complications with or without MAR. Salemi et al. noted poor to moderate and good to excellent agreement of two experienced radiologists in on and off MAR modes, respectively [21]. Bagis et al. found moderate agreement without MAR and denoising options and very good agreement with MAR option for their three observers [23]. Therefore, in this study we tried to account for the observer variability by using four observers. Our findings indicated excellent agreement of the observers. Observer variability can be attributed to factors including training and experience [24]. This can also explain why enhancements and modifications in radiologic images can affect the diagnosis of one observer, while having no effect on that of another one.

A limitation of the present study was using sheep hemimandibles in which the bone density as well as the cortication of the inferior alveolar canal borders may be different compared to human mandibles. Sheep mandibles have been used in other studies for placement of dental implants [16, 23]. In the present study, the diagnostic potential of activation of MAR algorithm was tested for detection of two types of simulated injuries to the inferior alveolar canal. These defects were simulated in about 1 mm limit of the canal, as in clinical scenarios, these injuries usually originate from minor miscalculations of available bone height in pre-operative radiographic images. Detection of pilot drill injuries to the inferior alveolar canal can be a more challenging task, as the canal borders are not visible in all cases, especially in the presence of detrimental metallic artifacts of dental implants. The findings of this study, when combined with those of other studies, can be useful for clinicians and radiologists in selecting the most appropriate tools in diagnosis of complications of dental implants. Although in most cases, purchasing the metal artifact reduction algorithm requires additional expenses, even a slight diagnostic advantage can justify its application.

Conclusion

Activation of MAR improved the diagnostic values of CBCT images in detecting implant-related injuries to the inferior alveolar canal. However, this increase was not statistically significant for most observers.

Data availability

The datasets used and/or analyzed during the current study available from the corresponding author on reasonable request.

References

  1. Probst LF, Vanni T, Cavalcante DFB, Silva ETD, Cavalcanti YW, Passeri LA et al. Cost-effectiveness of implant-supported dental prosthesis compared to conventional dental prosthesis. Rev Saude Publica. 2019;53.

  2. Eklund SA. Trends in dental treatment, 1992 to 2007. J Am Dent Assoc. 2010;141(4):391–9.

    Article  PubMed  Google Scholar 

  3. Mehdizadeh M, Booshehri SG, Kazemzadeh F, Soltani P, Motamedi MRK. Level of knowledge of dental practitioners in Isfahan, Iran about cone-beam computed tomography and digital radiography. Imaging Sci Dentistry. 2015;45(2):133–5.

    Article  Google Scholar 

  4. Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40(5):265–73.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Abdinian M, Moshkforoush S, Hemati H, Soltani P, Moshkforoushan M, Spagnuolo G. Comparison of cone beam computed tomography and digital radiography in detecting separated endodontic files and strip perforation. Appl Sci. 2020;10(23):8726.

    Article  CAS  Google Scholar 

  6. Alemam S, Abuelsadat S, Saber S, Elsewify T. Accuracy, sensitivity and specificity of three imaging modalities in detection of separated intracanal instruments. Giornale Italiano Di Endodonzia. 2020;34(1).

  7. Queiroz PM, Oliveira ML, Groppo FC, Haiter-Neto F, Freitas DQ. Evaluation of metal artefact reduction in cone-beam computed tomography images of different dental materials. Clin Oral Investig. 2018;22(1):419–23.

    Article  PubMed  Google Scholar 

  8. Bezerra IS, Neves FS, Vasconcelos TV, Ambrosano GM, Freitas DQ. Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts. Dentomaxillofac Radiol. 2015;44(6):20140428.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Oliveira MR, Sousa TO, Caetano AF, de Paiva RR, Valladares-Neto J, Yamamoto-Silva FP, et al. Influence of CBCT metal artifact reduction on vertical radicular fracture detection. Imaging Sci Dentistry. 2021;51(1):55.

    Article  Google Scholar 

  10. Nascimento EHL, Gaêta-Araujo H, Fontenele RC, Oliveira-Santos N, Oliveira-Santos C, Freitas DQ. Do the number of basis images and metal artifact reduction affect the production of artifacts near and far from zirconium dental implants in CBCT? Clin Oral Invest. 2021;25:5281–91.

    Article  Google Scholar 

  11. Cebe F, Aktan AM, Ozsevik AS, Ciftci ME, Surmelioglu HD. The effects of different restorative materials on the detection of approximal caries in cone-beam computed tomography scans with and without metal artifact reduction mode. Oral surgery, oral medicine, oral pathology and oral radiology. 2017;123(3):392–400.

  12. Nomier AS, Gaweesh YSE-D, Taalab MR, El Sadat SA. Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study. BMC Oral Health. 2022;22(1):1–11.

    Article  Google Scholar 

  13. de Freitas BN, da Motta RJG, Pauwels R, Oliveira-Santos C, Tirapelli C. Influence of metal artefact reduction on the diagnosis of contact between implant and mandibular canal in cone beam computed tomography: an ex-vivo study. Clin Oral Implants Res. 2023;34(7):741–50.

    Article  PubMed  Google Scholar 

  14. Juodzbalys G, Wang HL, Sabalys G. Injury of the Inferior Alveolar nerve during Implant Placement: a literature review. J Oral Maxillofac Res. 2011;2(1):e1.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Steinberg MJ, Kelly PD. Implant-related nerve injuries. Dent Clin North Am. 2015;59(2):357–73.

    Article  PubMed  Google Scholar 

  16. Sirin Y, Yildirimturk S, Horasan S, Guven K. Diagnostic potential of panoramic radiography and CBCT in detecting Implant-related Ex vivo injuries of the Inferior Alveolar Canal Border. J Oral Implantol. 2020;46(3):206–13.

    Article  PubMed  Google Scholar 

  17. Bagheri SC, Meyer RA. Management of mandibular nerve injuries from dental implants. Atlas Oral Maxillofac Surg Clin North Am. 2011;19(1):47–61.

    Article  PubMed  Google Scholar 

  18. Pogrel MA, Jergensen R, Burgon E, Hulme D. Long-term outcome of trigeminal nerve injuries related to dental treatment. J Oral Maxillofac Surg. 2011;69(9):2284–8.

    Article  PubMed  Google Scholar 

  19. Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012;4(2):229–40.

    Article  Google Scholar 

  20. Hegazy MA, Cho MH, Cho MH, Lee SY. Metal artifact reduction in Dental CBCT images using direct Sinogram correction combined with metal path-length weighting. Sensors. 2023;23(3):1288.

    Article  ADS  PubMed  PubMed Central  Google Scholar 

  21. Salemi F, Jamalpour MR, Eskandarloo A, Tapak L, Rahimi N. Efficacy of metal artifact reduction algorithm of cone-beam computed tomography for detection of fenestration and dehiscence around dental implants. J Biomedical Phys Eng. 2021;11(3):305.

    Article  Google Scholar 

  22. Sheikhi M, Behfarnia P, Mostajabi M, Nasri N. The efficacy of metal artifact reduction (MAR) algorithm in cone-beam computed tomography on the diagnostic accuracy of fenestration and dehiscence around dental implants. J Periodontol. 2020;91(2):209–14.

    Article  PubMed  Google Scholar 

  23. Bagis N, Kurt MH, Evli C, Camgoz M, Atakan C, Peker Ozturk H, et al. Evaluation of a metal artifact reduction algorithm and an adaptive image noise optimization filter in the estimation of peri-implant fenestration defects using cone beam computed tomography: an in-vitro study. Oral Radiol. 2022;38(3):325–35.

    Article  PubMed  Google Scholar 

  24. Parker JM, Mol A, Rivera EM, Tawil PZ. Cone-beam computed tomography uses in clinical endodontics: observer variability in detecting periapical lesions. J Endod. 2017;43(2):184–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

There are no acknowledgments.

Funding

This study was financially supported by Isfahan University of Medical Sciences (#3400940).

Author information

Authors and Affiliations

Authors

Contributions

P.S and H.D designed the study, M.A prepared the specimen, and G.S analyzed the data. K.B performed the intervention. C.R contributed through revision process and study design. all authors contributed to writing and reviewed the article.

Corresponding author

Correspondence to Kimia Baghaei.

Ethics declarations

Ethics approval and consent to participate

The Research Ethics Committee at Isfahan University of Medical Sciences has approved this study (#IR.MUI.REC.1400.072, approval date: 03/08/2022). The sheep mandible was purchased from a local butcher. There were no human subjects involved in this study.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions