Phantom study of a fully automatic radioactive seed placement robot for the treatment of skull base tumours

Abstract

Background

Interstitial brachytherapy is a form of intensive local irradiation that facilitates the effective protection of surrounding structures and the preservation of organ functions, resulting in a favourable therapeutic response. As surgical robots can perform needle placement with a high level of accuracy, our team developed a fully automatic radioactive seed placement robot, and this study aimed to evaluate the accuracy and feasibility of fully automatic radioactive seed placement for the treatment of tumours in the skull base.

Methods

A fully automatic radioactive seed placement robot was established, and 4 phantoms of skull base tumours were built for experimental validation. All the phantoms were subjected to computed tomography (CT) scans. Then, the CT data were imported into the Remebot software to design the preoperative seed placement plan. After the phantoms were fixed in place, navigation registration of the Remebot was carried out, and the automatic seed placement device was controlled to complete the needle insertion and particle placement operations. After all of the seeds were implanted in the 4 phantoms, postoperative image scanning was performed, and the results were verified via image fusion.

Results

A total of 120 seeds were implanted in 4 phantoms. The average error of seed placement was (2.51 ± 1.44) mm.

Conclusion

This study presents an innovative, fully automated radioactive particle implantation system utilizing the Remebot device, which can successfully complete automated localization, needle insertion, and radioactive particle implantation procedures for skull base tumours. The phantom experiments showed the robotic system to be reliable, stable, efficient and safe. However, further research on the needle-soft tissue interaction and deformation mechanism of needle puncture is still needed.

Peer Review reports

Introduction

Brachytherapy is defined as the short-distance treatment of malignant tumours with a radioactive isotope placed on, in, or near the lesion to continuously emit γ-rays [1], perpetually destroying tumour cells [2]. Because the dose decreases exponentially with increasing distance from the radioactive source, tumours can be effectively damaged while minimizing damage to the surrounding normal tissue [3]. Brachytherapy has been widely used in treating several types of cancer, including head and neck [4], prostate [5] and lung [6] cancer. Given the complexity of the anatomical structures in the skull base, which include the eyeball, numerous blood vessels (such as the internal carotid and jugular veins), cranial nerves IX-XII, and spinal cord tissues, all of which are within a confined space [7], resection of head and neck malignancies usually causes facial defects and aesthetic deformities. Additionally, free or pedicled flaps are even sometimes needed for postoperative reconstruction [8, 9]. Brachytherapy has a unique advantage for treating malignant tumours located in the skull base, especially for patients with advanced or recurrent cancer or as a supplementary therapy after surgical resection. However, there is an inevitable issue that persists: staff exposure. Although a single radioactive particle emits minimal radiation doses and has a limited range, the cumulative dose in the target area is relatively high because of the accumulation of radioactive particles in the patient’s body, occasionally reaching levels as high as 120 Gy. Even with the use of protective gear, including lead aprons, collars, and caps, the crystalline lens of the eye remains exposed to radiation during the procedure and is particularly vulnerable to radiation damage.

Currently, the use of robots and advanced technology in brachytherapy is increasing, and there is a substantial body of literature on robot-assisted brachytherapy for lung [10], bladder [11, 12], and prostate [13] cancer. Robot-assisted needle insertion technology will help surgeons perform high-accuracy seed implantation, improve the calculation of optimal seed placement and reduce radiation exposure to medical staff [14]. Moreover, similar to remote after-loading technology, physicians only need to stay in a dedicated shielded room with audio/visual surveillance. Considering the issues described above, the development of an automated radioactive particle implantation device could facilitate the replacement of manual implantation by surgeons via robotic procedures.

Our preliminary experiments suggest that the accuracy of Remebot puncture at the skull base meets clinical requirements [15]. The robotic arm moves axially with precision, as directed by the surgeon, enabling precise linear movement towards or away from the target point, potentially allowing robot to replace surgeons in conducting puncture and injection procedures. This study aimed to assess the accuracy and feasibility of a fully automated radioactive seed implantation robot for treating skull base tumours.

Materials and methods

Simulation skull models (1:1, PNATOMY, Zhangjiagang Kexin Scientific Instrument Co., Ltd., Suzhou, China), ultralight clay (Ninitegong, Shandong Zhiqi Hangyi Puzzle Toy Co., Ltd., Linyi, China), a large amount of Kirschner wire (diameter 6 mm, length 200 mm), an appropriate amount of fresh pork and pigskin, MK06A markers (Beijing Baihuiweikang Technology Co., Ltd., Beijing, China), a Remebot robot, and 3 − 0 sutures were used.

Fully automatic radioactive seed placement robot

The end-effector, a fully automatic radioactive seed placement device consisting mainly of a quick-release connector, linear module, calibration plate and radioactive seed cartridge, was rigidly mounted on the Remebot flange. The linear module consisted of three parts: a stepping motor, an actuator and a ball screw. The ball screw mechanism could convert the rotating motion of the motor into linear motion by controlling the forwards and reverse motion of the stepper motor to realize the insertion and withdrawal of the needle. The rotational speed of the stepper motor could be controlled by sending different instructions, and the speed of needle insertion and withdrawal could be adjusted to meet various puncture needs. The calibration plate was used to install the calibration block with a specific size and pattern, and the camera of the Remebot could realize the calibration and registration of the tool centre point at the end of the puncture through identification of the calibration block.

Model and radioactive seeds

A fresh pork specimen measuring 7 cm × 4 cm × 3 cm was uniformly encased in ultralight clay. Cavities, including the pterygopalatine fossa and posterior nasal aperture, were initially filled with sufficient ultralight clay. This approach ensured the stability of the tumour model, ensuring that it was securely established. The ultralight clay encasing the fresh pork was situated between the skull base and the bilateral mandibular ramus to simulate a skull base tumour. Holes were drilled using a dental handpiece from the left to the right mandibular angle along the inferior mandibular margin on the 1:1 skull model, with an interval of approximately 2 cm separating each pair of holes. Two holes were drilled bilaterally behind the hypoglossal canal and adjacent to the foramen magnum on the occipital bone of the skull model. An arch-shaped specimen of fresh pigskin measuring 15 cm × 10 cm was trimmed, and subcutaneous fat was removed to leave only the epidermis and dermis. The prepared fresh pigskin was affixed beneath the mandible and near the foramen magnum of the occipital bone using 3–0 sutures. The Kirschner wire was cut into 5-mm-long pieces for use instead of radioactive seeds to avoid radiation exposure to the researchers.

Preoperative design

MK06A markers were affixed to the forehead and bilateral temporal regions of each skull model. The skull models were placed in a supine position for computed tomography (CT) imaging. The CT data were imported into Remebot 4.0 software for reconstruction and design of a radioactive particle implantation plan. The implantation needle trajectories and seed positions were planned for even distribution within the tumour model. A horizontal particle interval of 10 mm was set on the same axial plane of the CT images. Either two or three particles were implanted in each needle track. The deepest particle in each needle track was designated the target point. After retracting 10 mm from the target point along the needle track towards the surface, the position of the second particle was designed. In each needle track, 2 or 3 particles were implanted, for a total of 30 particles per skull model.

Procedure

Following stabilization of the skull model’s spatial position using a head frame, robot registration was performed. (Fig. 1) To enable the automatic particle implantation device to perform the puncture action, the distance between the tip of the needle on the robotic arm’s puncture sheath and the target point was set to 70 mm. Once the robotic arm was in position, it was controlled via Remebot 4.0 software to advance axially 70 mm towards the target point, after which the skin was penetrated to reach the target site. Commands are input to the automatic implantation device, which advanced or retracted the inner needle to execute the particle implantation process. The robotic arm was manoeuvred to axially retract by 10 mm for the implantation of a subsequent particle, followed by an axial retraction of 60 mm to remove the sheath needle from the skin. Once the skin had resettled into its initial position, the spatial location of the actual skin puncture was marked using a navigation probe, and the corresponding spatial coordinates were noted within the software. The same technique was then applied for puncturing the following target point.

Postoperative evaluation

After all the particles were implanted, a CT scan was performed again. The number of particles on the postoperative CT images was calculated to confirm whether the preoperative design was met for 30 particles. The data were imported into Remebot 4.0 software for reconstruction. The postoperative images were merged with the preoperative surgical plan. The pre- and post-surgical coordinates are denoted as (X1, Y1, Z1) and (X2, Y2, Z2), respectively.

$$d=\sqrt{\varDelta {X}^{2}+\varDelta {\text{Y}}^{2}+\varDelta {Z}^{2}}$$

The difference between the pre- and post-operative spatial coordinates of the radioactive particles was calculated to evaluate the accuracy of the fully automatic radioactive seed placement robot.

Fig. 1
figure 1

Following the stabilization of the skull model’s spatial position using a head frame, robot registration is performed

Statistical analysis

Analysis was performed using IBM SPSS software version 20 (IBM Corp., Armonk, NY, USA). The placement error was calculated by comparing the preoperatively planned and actual postoperative placement points according to the Euclidean distance between the centres of the planned and actual placement points. One-way ANOVA was used for pairwise comparison of the placement error of the 30 particles implanted in the 4 skull phantom experiments, and a p value less than 0.05 was considered to indicate statistical significance.

Results

A total of 120 seeds were implanted in 4 phantoms. The average error of seed placement positions was (2.51 ± 1.44) mm (Table 1; Fig. 2), Fig. 3 shows the result of superimposition of pre-(yellow) and post-(blue) needle pathway.

Table 1 One-way analysis of variance of phantom
Fig. 2
figure 2

Postoperative CT images of 4 phantoms

Fig. 3
figure 3

The superimposition of pre-(yellow) and post-(blue) needle pathway shows the deviation of puncture points and angle

Discussion

Compared with external beam radiotherapy (EBRT), brachytherapy offers several advantages: (1) A high cumulative dose can be delivered to the target region, and the implantation of iodine-125 particles can maximally increase the dose delivered to the targeted area while significantly reducing the dose delivered to the surrounding normal tissues. (2) Compared with the fractionated, high-dose exposures involved in EBRT, brachytherapy consists of sustained, low-dose irradiation within the tumour, continuously targeting actively proliferating cells. (3) Radioactive seed implantation is highly conformal, minimizing trauma to normal tissues. Iodine-125 particle sources are characterized by a rapid decrease in γ-ray energy with increasing distance from the radiation source. (4) Following tumour reduction through treatment, particles tend to accumulate at the tumour’s centre, a phenomenon that maintains the radiation dose delivered to the target area to a certain degree as the radiation source decays [4]. Theoretically, the curative effect of brachytherapy highly depends on the accuracy of seed implantation. Poor accuracy may seriously affect the therapeutic efficacy and result in damage to the surrounding normal tissue. At present, surgeons can achieve accurate radioactive seed implantation with the assistance of 3D-printed templates [16, 17]. Nevertheless, surgeons still have to perform the procedure at the patient’s bedside. Thus, there still exists another inevitable issue, which is staff exposure, especially for surgeons with prolonged exposure to brachytherapy. If a robot could perform seed implantation in place of surgeons, the surgeons could stay at a safe distance from the radioactive seeds.

The last decade has witnessed major progress in the field of minimally invasive and robot-assisted surgeries [18], for instance, in skull base biopsy [19] and kidney [20] and breast [21] surgery. In prostate brachytherapy [22], a placement accuracy of millimetres is needed, and a needle placement accuracy of < 2.7 mm is generally considered acceptable in brachytherapy [23]. However, in the skull base, the precision requirement is greater, and robot-assisted needle insertion technology will help surgeons perform high-accuracy seed implantation. Currently, most related research has focused primarily on applications in treating liver [24], prostate [25], lung [26], and cervical [27] cancer, with few studies addressing the application of such robots in treating maxillofacial tumours. In 2017, Cui [28] established a surgical robot for radioactive seed implantation capable of automating radioactive seed placement. The robot’s mechanical structure design and the seed storage and propulsion device have been completed. The positional error for each robotic arm joint is less than 0.1 mm. In 2019, Zhu [29] reported the use of a fully automatic robot for radioactive particle implantation in treating skull base tumours. This technology is currently in the in vitro validation stage using models and cadaveric skulls, with an implantation accuracy of 1.41 ± 0.38 mm and 2.48 ± 0.32 mm, respectively. The surgical robots developed by Zhu and Cui both attach to either side of the operating table and can achieve movement with six degrees of freedom (DOF). The Remebot used in this study, however, is mounted on an operating trolley and can be moved freely within the operating room. It also features a 6-DOF mechanical arm that can precisely reach any target point and navigate paths within the surgical field, offering enhanced flexibility. As mentioned in several other studies on surgical robots, the sources of error of the Remebot mainly consist of the following components: the monitored accuracy, large orientation workspace, real-time tissue resistance monitoring, haptic control, maintenance program, sterilizable needle insertion device, CT scan compliance, automated safety control and fast manual control. Any component of the system can contribute to the overall systematic error. However, our preliminary work has verified the accuracy and safety of the Remebot device for performing skull base tumour punctures [30].

An eligible phantom could provide the possibility of overall quality assurance and specific quality control of the brachytherapy process. To simulate an authentic surgical scenario and mimic real tissue mechanics, we added fresh animal skin to the skull model to closely replicate the surgical puncture process. Previous research has reported that the total needle insertion force is the total of dissimilar forces spread sideways relative to the shaft of the insertion needle, for example, the cutting force, stiffness force, and frictional force [31,32,33]. Diverse types of needle tips may also affect skin deviation during puncture [34]. On the basis of our preliminary research and clinical experience, we found that placement error is significantly associated with deviations at the skin puncture point.

The results indicated that puncture errors varied across different skin areas. In the submandibular region near the mandible, that is, the outer edge of the skin, both the actual puncture point and the needle angulation exhibited minimal deviation. Conversely, larger deviations were observed in the central and posterior skin areas near the foramen magnum, which may be attributable to the varying skin tension in these areas. The skin at the outer edge was sutured to the inferior border of the mandible, resulting in greater tension and a reduced likelihood of shifting during puncture. In contrast, the more relaxed skin near the centre and foramen magnum was prone to deviation from puncture forces upon needle contact. Thus, the process of puncture needle insertion or retraction through the skin has an effect on accuracy. However, the mechanism of these needle-tissue interactions has not been elucidated. Consequently, predicting and compensating for needle deflection through calculations was not feasible, necessitating further research into the mechanisms of interaction between various needle tip shapes and skin.

The end effector for radioactive particle implantation currently under investigation is notably bulky [35]. The distance from the mechanical arm’s centre point to the puncture needle tip is approximately 31.6 mm, comprising a 10.8 mm puncture needle and a 20.8 mm device body. This design accommodates the need for needle insertion beneath the jaw or through the sigmoid notch of the mandible for most skull base tumours but necessitates a relatively long puncture trajectory. If entry is gained from beneath the jaw, the mechanical arm must manoeuvre the puncturing apparatus to the patient’s side. During the operation, there is a risk of colliding with the patient or failing to reach the intended position. Consequently, refining the end-effector to reduce its size and enhance its practicality is a necessary next step.

Conclusion

This study presents an innovative, fully automated radioactive particle implantation system utilizing the Remebot device, which can successfully complete automated localization, needle insertion, and radioactive particle implantation procedures for skull base tumours. The phantom experiments showed the robotic system to be reliable, stable, efficient and safe. Currently, the precision of radioactive particle implantation robots can preliminarily meet the requirements of clinical surgery. Subsequent research will aim to further elucidate the mechanisms of interaction and deformational forces between the puncture needle and soft tissues.

Data availability

All data are calculated by the software itself. The datasets used and/or analyzed during the current study available from the corresponding author on reasonable request.

References

  1. Skowronek J. Current status of brachytherapy in cancer treatment - short overview. J Contemp Brachytherapy. 2017;9(6):581–9.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Sato M, Saito Y, Takayama T, Omata T, Watanabe H, Yoshimura R, Miura M. Remote radioactive seed-loading device for permanent brachytherapy of oral cancer with Au-198 grains. ROBOMECH J 2017, 4(1).

  3. Aumuller P, Rothfuss A, Polednik M, Abo-Madyan Y, Ehmann M, Giordano FA, Clausen S. Multiple direction needle-path planning and inverse dose optimization for robotic low-dose rate brachytherapy. Z Med Phys. 2022;32(2):173–87.

    Article  PubMed  Google Scholar 

  4. Rodin J, Bar-Ad V, Cognetti D, Curry J, Johnson J, Zender C, Doyle L, Kutler D, Leiby B, Keane W, et al. A systematic review of treating recurrent head and neck cancer: a reintroduction of brachytherapy with or without surgery. J Contemp Brachytherapy. 2018;10(5):454–62.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Buzurovic IM, Salinic S, Orio PF, Nguyen PL, Cormack RA. A novel approach to an automated needle insertion in brachytherapy procedures. Med Biol Eng Comput. 2018;56(2):273–87.

    Article  PubMed  Google Scholar 

  6. Ma X, Yang Z, Jiang S, Zhang G, Chai S. A novel auto-positioning method in Iodine-125 seed brachytherapy driven by preoperative planning. J Appl Clin Med Phys. 2019;20(6):23–30.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Huang MW, Zhang JG, Zheng L, Liu SM, Yu GY. Accuracy evaluation of a 3D-printed individual template for needle guidance in head and neck brachytherapy. J Radiat Res. 2016;57(6):662–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Afrah A, Aldelaimi HHE, Tahrir N, Aldelaimi. Khalil Abdulla Mohammed: tumors of Craniofacial Region in Iraq (Clinicopathological Study). J Res Med Dent Sci. 2021;9(1):66–71.

    Google Scholar 

  9. Aldelaimi TN, Khalil AA. Reconstruction of Facial defect using Deltopectoral Flap. J Craniofac Surg. 2015;26(8):e786–788.

    Article  PubMed  Google Scholar 

  10. Ma X, Yang PZ, Jiang PS, Huo PB, Cao Q, Chai PS, Wang PH. Effectiveness and safety of a robot-assisted 3D personalized template in (125)I seed brachytherapy of thoracoabdominal tumors. J Contemp Brachytherapy. 2018;10(4):368–79.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Cunha JAM, Flynn R, Belanger C, Callaghan C, Kim Y, Jia X, Chen Z, Beaulieu L. Brachytherapy Future directions. Semin Radiat Oncol. 2020;30(1):94–106.

    Article  PubMed  Google Scholar 

  12. Bosschieter J, Vis AN, van der Poel HG, Moonen LM, Horenblas S, van Rhijn BWG, Pieters BR, Nieuwenhuijzen JA, Hendricksen K. Robot-assisted laparoscopic implantation of Brachytherapy catheters in bladder Cancer. Eur Urol. 2018;74(3):369–75.

    Article  PubMed  Google Scholar 

  13. Aleong AM, Looi T, Luo K, Zou Z, Waspe A, Singh S, Drake JM, Weersink RA. Preliminary study of a modular MR-Compatible Robot for image-guided insertion of multiple needles. Front Oncol. 2022;12:829369.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Liu W, Yang Z, Fang P, Jiang S. Deflection simulation for a needle adjusted by the insertion orientation angle and axial rotation during insertion in the muscle-contained double-layered tissue. Med Biol Eng Comput. 2020;58(10):2291–304.

    Article  PubMed  Google Scholar 

  15. Fanhao M, Xiaodong X, Bo Q, Lejun X, Yu S, Liang Z, Tingting J, Rui Z, Depeng Z, Ran A, et al. A new multimodal, image-guided, robot-assisted, interstitial brachytherapy for the treatment of head and neck tumors-A preliminary study. Int J Med Robot. 2020;16(5):1–5.

    Article  PubMed  Google Scholar 

  16. Tu Y, Wang J, Chen Y, Chen Y, Tang Q. Clinical implementation of three-dimensional standardized template-guided brachytherapy for patients with locally advanced cervical cancer. J Contemp Brachytherapy. 2023;15(6):405–13.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Fahimian BP, Liu W, Skinner L, Yu AS, Phillips T, Steers JM, DeMarco J, Fraass BA, Kamrava M. 3D printing in brachytherapy: a systematic review of gynecological applications. Brachytherapy. 2023;22(4):446–60.

    Article  PubMed  Google Scholar 

  18. Padasdao B, Konh B. A model to Predict Deflection of an active Tendon-Driven notched needle inside soft tissue. J Eng Sci Med Diagn Ther. 2024;7(1):011006.

    PubMed  Google Scholar 

  19. Zhu JH, Wang J, Wang YG, Li M, Guo YX, Liu XJ, Guo CB. Performance of robotic assistance for Skull Base Biopsy: a Phantom Study. J Neurol Surg B Skull Base. 2017;78(5):385–92.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Kinnear N, Kucheria A, Ogbechie C, Adam S, Haidar O, Cotter Fonseca P, Brodie A, Pullar B, Adshead J. Concordance between renal tumour biopsy and robotic-assisted partial and radical nephrectomy histology: a 10-year experience. J Robot Surg. 2024;18(1):45.

    Article  PubMed  Google Scholar 

  21. Fu X, Ma Y, Hou Y, Liu Y, Zheng L. Comparison of endoscopic bilateral areolar and robotic-assisted bilateral axillo-breast approach thyroidectomy in differentiated thyroid carcinoma: a propensity-matched retrospective cohort study. BMC Surg. 2023;23(1):338.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Popescu T, Kacso AC, Pisla D, Kacso G. Brachytherapy next generation: robotic systems. J Contemp Brachytherapy. 2015;7(6):510–4.

    Article  PubMed  PubMed Central  Google Scholar 

  23. de Jong TL, van de Berg NJ, Tas L, Moelker A, Dankelman J, van den Dobbelsteen JJ. Needle placement errors: do we need steerable needles in interventional radiology? Med Devices (Auckl). 2018;11:259–65.

    PubMed  Google Scholar 

  24. Lin X, Zhou S, Wen T, Jiang S, Wang C, Chen J. A novel multi-DoF surgical robotic system for brachytherapy on liver tumor: design and control. Int J Comput Assist Radiol Surg. 2021;16(6):1003–14.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Dai X, Zhang Y, Jiang J, Li B. Image-guided robots for low dose rate prostate brachytherapy: perspectives on safety in design and use. Int J Med Robot. 2021;17(3):e2239.

    Article  PubMed  Google Scholar 

  26. Dou H, Jiang S, Yang Z, Sun L, Ma X, Huo B. Design and validation of a CT-guided robotic system for lung cancer brachytherapy. Med Phys. 2017;44(9):4828–37.

    Article  PubMed  Google Scholar 

  27. Podder TK, Beaulieu L, Caldwell B, Cormack RA, Crass JB, Dicker AP, Fenster A, Fichtinger G, Meltsner MA, Moerland MA, et al. AAPM and GEC-ESTRO guidelines for image-guided robotic brachytherapy: report of Task Group 192. Med Phys. 2014;41(10):101501.

    Article  PubMed  Google Scholar 

  28. Cui T, Wang Y, Duan X, Ma X. Control Strategy and experiments for Robot assisted craniomaxillofacial surgery system. Math Probl Eng. 2019;2019:1–12.

    Google Scholar 

  29. Zhu JH, Wang J, Wang YG, Li M, Liu XJ, Guo CB. Prospect of robotic assistance for fully automated brachytherapy seed placement into skull base: experimental validation in phantom and cadaver. Radiother Oncol. 2019;131:160–5.

    Article  PubMed  Google Scholar 

  30. Meng FH, Song Y, Qiao B, Jin NH, Zhu YM, Liang BF, Gao DF, Zhang HZ. Image-guided, surgical robot-assisted percutaneous puncture of the foramen ovale and foramina stylomastoideum: a cadaveric study. Chin Med J (Engl). 2021;134(19):2362–4.

    Article  PubMed  Google Scholar 

  31. Barua R, Datta S, RoyChowdhury A, Datta P. Study of the surgical needle and biological soft tissue interaction phenomenon during insertion process for medical application: a Survey. Proc Inst Mech Eng H. 2022;236(10):1465–77.

    Article  PubMed  Google Scholar 

  32. Tucan P, Vaida C, Horvath D, Caprariu A, Burz A, Gherman B, Iakab S, Pisla D. Design and Experimental Setup of a Robotic Medical Instrument for Brachytherapy in Non-Resectable Liver Tumors. Cancers (Basel) 2022, 14(23).

  33. Gherman B, Hajjar NA, Tucan P, Radu C, Vaida C, Mois E, Burz A, Pisla D. Risk Assessment-oriented design of a needle insertion robotic system for non-resectable liver tumors. Healthc (Basel) 2022, 10(2).

  34. Muzzammil HM, Zhang YD, Ejaz H, Yuan Q, Muddassir M. A review on tissue-needle interaction and path planning models for bevel tip type flexible needle minimal intervention. Math Biosci Eng. 2024;21(1):523–61.

    Article  PubMed  Google Scholar 

  35. Duan XG, Chen NN, Wang Y, Kong XZ. Design and implementation of Radioactive seeds implantation surgery Robot for Cranio-maxillofacial tumors. Mech Sci Technol Aerosp Eng. 2017;36(3):341–6.

    Google Scholar 

Download references

Acknowledgements

None.

Funding

This study was supported by National Natural Science Foundation of China (62073043) and National Key Research and Development Plan (No. 2022YFC2405404).

Author information

Authors and Affiliations

Authors

Contributions

Conceptualization: Meng Fanhao, Zhang Haizhong, Duan Xingguang Performed the experimental and analytical part: Meng Fanhao, Liang Bofu, Jin Nenghao, Xie Dongsheng Writing – original draft preparation: Meng Fanhao, Qiao Bo, Xie Dongsheng, Tian Huanyu, Zhang Ning Writing - review and editing: Song Yu, Chang Shimin, Gao Runtao, Zhang Haizhong, Duan Xingguang. Funding acquisition: Meng Fanhao, Duan Xingguang Supervision: Duan Xingguang, Zhang Haizhong.

Corresponding authors

Correspondence to Duan Xingguang or Zhang Haizhong.

Ethics declarations

Competing interests

The authors declare no competing interests.

Ethical approval and consent to participate

This article does not report on any studies with human participants or living animals performed by any of the authors.

Consent for publication

Not applicable.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fanhao, M., Dongsheng, X., Nenghao, J. et al. Phantom study of a fully automatic radioactive seed placement robot for the treatment of skull base tumours. BMC Oral Health 24, 420 (2024). https://doi.org/10.1186/s12903-024-04089-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-024-04089-w

Keywords