Accuracy evaluation of dental CBCT and scanned model registration method based on pulp horn mapping surface: an in vitro proof-of-concept

Abstract

Background and aim

3D fusion model of cone-beam computed tomography (CBCT) and oral scanned data can be used for the accurate design of root canal access and guide plates in root canal therapy (RCT). However, the pose accuracy of the dental pulp and crown in data registration has not been investigated, which affects the precise implementation of clinical planning goals. We aimed to establish a novel registration method based on pulp horn mapping surface (PHMSR), to evaluate the accuracy of PHMSR versus traditional methods for crown-pulp registration of CBCT and oral scan data.

Materials and methods

This vitro study collected 8 groups of oral scanned and CBCT data in which the left mandibular teeth were not missing, No. 35 and No. 36 teeth were selected as the target teeth. The CBCT and scanned model were processed to generate equivalent point clouds. For the PHMSR method, the similarity between the feature directions of the pulp horn and the surface normal vectors of the crown were used to determine the mapping points in the CBCT point cloud that have a great influence on the pulp pose. The small surface with adjustable parameters is reconstructed near the mapping point of the crown, and the new matching point pairs between the point and the mapping surface are searched. The sparse iterative closest point (ICP) algorithm is used to solve the new matching point pairs. Then, in the C +  + programming environment with a point cloud library (PCL), the PHMSR, the traditional sparse ICP, ICP, and coherent point drift (CPD) algorithms are used to register the point clouds under two different initial deviations. The root square mean error (RSME) of the crown, crown-pulp orientation deviation (CPOD), and position deviation (CPPD) were calculated to evaluate the registration accuracy. The significance between the groups was tested by a two-tailed paired t-test (p < 0.05).

Results

The crown RSME values of the sparse ICP method (0.257), the ICP method (0.217), and the CPD method (0.209) were not significantly different from the PHMSR method (0.250). The CPOD and CPPD values of the sparse ICP method (4.089 and 0.133), the ICP method (1.787 and 0.700), and the CPD method (1.665 and 0.718) than for the PHMSR method, which suggests that the accuracy of crown-pulp registration is higher with the PHMSR method.

Conclusion

Compared with the traditional method, the PHMSR method has a smaller crown-pulp registration accuracy and a clinically acceptable deviation range, these results support the use of PHMSR method instead of the traditional method for clinical planning of root canal therapy.

Peer Review reports

Introduction

RCT controls the microbial infection in the root canal by proper shaping and disinfection of the root canal system to promote the healing of periapical lesions and tissue regeneration [1]. Physicians are required to have rich clinical experience, with the help of familiar anatomical knowledge, accurate 3D images, and stable operation, the exploration and preparation are carried out along the expected root canal access [2]. CBCT can provide 3D structural information on teeth, bones, main root canals, and periapical tissues [3]. Due to limited image resolution and various artifacts (including metal-induced artifacts), dental CBCT alone may not be able to describe the precise details of the tooth surface [4]. The CBCT and the optical surface scanned data were registered and integrated into the guided surgery software, which can express the crown and internal anatomical details at the same time [5]. Researchers have adopted the surgical method of “inlay-guided endodontics” by integrating CBCT and 3D scanned data, designing the best virtual pulp opening access, and accurately placing the file into the planned position with the help of a 3D printing guide plate [6,7,8,9]. Torres et al. have used 3D models for intraoperative dynamic navigation of RCT [10]. Compared with free hand operation, the use of the above digital methods can reduce the difference between the planned and actual instrument positions, accurately visualize its path and surrounding oral anatomy, and achieve low-risk, minimally invasive, and precise treatment. The registration process greatly affects the deviation of the designed virtual access or guide plate.

Commonly used clinical registration can be divided into two categories, marker-based registration is to fix the existing markers around the surgical area in a certain way, such as bone implant nails [11], occlusal splints [12], and Lego bricks [8], and the markers of the two scans can be aligned. However, this method takes up a certain amount of oral space and invasive operation will cause discomfort to patients. For unmarked registration, on the one hand, the anatomical features in the surgical area can be used as special markers, such as cusps [13], etc., but the method is not robust when tooth features are missing; On the other hand, using 3D point set registration algorithms, including point-based (e.g. ICP [14]) and surface-based registration, which are the computationally fast and simple method in oral clinical practice [15,16,17,18,19].

Currently, traditional algorithms embedded in Mimics, Rapidform, and Geomagic Studio software are mostly used for clinical registration. Kim et al. used manual surface-based registration, and the mean registration accuracy of mandibular teeth was 0.13 ± 0.11 mm [15]. Lim et al. used the three-point method for manual coarse registration, and then used the ICP algorithm for fine registration. The mean registration accuracy of the mandibular was 0.21 ± 0.04 mm [16]. Lee et al. developed a registration method for orthodontic treatment, using surface-based fine automatic registration. The difference between the expected and the actual maxillary root position was 0.1678 ± 0.3178 mm [17]. Ye et al. used the same method and reported that the registration errors of CBCT pixels at 0.2 mm and 0.4 mm were 0.163 ~ 0.231 mm and 0.202 ~ 0.345 mm, respectively [18]. Guo et al. used a manually operated three points structure reference plane for point-based registration, and the registration accuracy of the lower jaw was 0.151 ± 0.0337 mm [19]. Sun et al. used surface-based regional automatic registration with a mean registration error of 0.33 mm [20]. Yi et al. used the ICP algorithm to align the virtual registration unit and optical scanned data to evaluate the dental implant accuracy [21]. Rangel et al. placed titanium markers on the gingiva to match dental casts into CT data, and the mean deviation of the lower jaw were and 0.30 mm [22].

Some scholars have proposed other registration methods. Bai proposed to generate a multi-layer spherical point set inside the marker ball, and used the weighted ICP algorithm to perform multi-layer point set registration, which solved the problem of surface height deformity of the laser scanning marker ball, and the mean square error was 0.301 mm [23]. Yau et al. adopted the three-point location and ICP algorithm to achieve the crown registration [24]. Zhang et al. [25] used the principal component analysis and the ICP method to register the two models. Dai et al. [26] proposed a multi-point registration method to achieve rapid registration of crown optical scanned data and root CBCT data. Wang et al. [27] used the sparse ICP method to achieve accurate registration of different dentition coordinate systems. Lin et al. [28] proposed an improved iterative ICP algorithm and convex hull selection method for the automatic segmentation of dentition models. Lee et al. [5] proposed a novel depth map-based registration method to register 3D surface models, and performed the ICP method for the subsequent fine registration. Lee et al. proposed deep-pose regression neural networks and optimal clusters for automatic registration [29]. However, these recent studies only considered the registration deviation from crown to crown, and did not introduce pulp pose information into the registration. Moreover, the effectiveness of traditional methods for pulp to crown registration has not been studied.

The study aims to propose a registration method based on the pulp horn mapping surface (PHMSR). Four methods (PHMSR, sparse ICP [30], ICP [14], and CPD [31]) were used to register the CBCT and optical scanned point cloud of the left mandibular teeth. The RSME and crown-pulp pose deviations after registration were compared. The null hypothesis was that no difference would be found in the precision of the crown-pulp registration between the PHMSR and conventional methods.

Materials and method

Study design

This study was approved by the Medical Ethics Committee of the 2nd Affiliated Hospital of Harbin Medical University (YJSKY2022-115) and informed consents sought from the subjects whose scans were used as a study material. The sample size was calculated by a software program (G*power 3.1.9.7, Heinrich-Heine-Universität, Germany), assuming that the alpha error of 5%, a study power of 80%, and an effect size of 0.92, a minimum sample size of n = 8 per group was needed [32]. All oral scanned and CBCT data were selected from the 2nd Affiliated Hospital of Harbin Medical University, the subjects were aged between 25 and 35 years old, 4 males and 4 females. The inclusion criteria were no loss of left mandible teeth, no artifacts, and tooth crowns were not severely damaged. Exclusion criteria were no history of endodontic treatment or no anatomical variations to avoid artifacts hampering the registration of the real main anatomical configurations. To reduce the calibration error of the anatomical variation, intraobserver and interobserver calibrations were previously performed while examining selected sample images with diverse morphology. Each CBCT was evaluated and interpreted by two radiologists twice with a one-week interval between assessments. The reliability data was analyzed using the weighted Kappa test, and the Kappa coefficients of all samples were greater than 0.91. The study images were assessed independently. The research flow is shown in Fig. 1.

Fig. 1
figure 1

Research flow

Experimental equipment and software

Each CBCT was collected by an oral X-ray computed tomography system (SBR3D, SOREDEX PaloDEx Group Oy, Finland). The dimensions of CBCT are 580 × 580 × 300, and the resolution is 0.25 × 0.25 × 0.25 mm. All optical scanned data were collected by a digital intraoral scanner (iTero Element, Align Technology, USA), and the scanning accuracy was \(20\mu m\). The number of point clouds is about 150000, forming 300000 triangular meshes.

We used Matlab 2020b (MathWorks, USA) for image filtering, segmentation, 3D reconstruction, and registration error calculation, and used reverse engineering software Geomagic warp 2017 (3D System, USA, Morrisville) for extraction of target teeth and processing of 3D point cloud. The PHMSR, ICP [14], sparse ICP [31], and coherent point drift (CPD) [32] algorithms were based on the C +  + programming language with the PCL. For experiment evaluation, the experiment environment is Ubuntu 20.04 with X64 processor (AMD Ryzen 5 3600 6-Core Processor). Statistical analyses were performed using SPSS software (version 22.0; IBM Corp., Armonk, NY, USA).

Data processing

All CBCT data were exported to the Digital Imaging and Communications in Medicine (DICOM) format. All scanned data are exported to STL format. No. 35 and No. 36 teeth were selected as the target teeth.

In general, the anatomical region boundary of the pulp cavity in the actual CBCT images are generally low in brightness, and blurred in the boundary. In the process of medical image acquisition, the unknown mixed noise will affect image quality, so the pulp cavity is difficult to capture. To minimize the effects of noise and enhance the strength of teeth and pulp contours, the wavelet [33] and NL-means [34] hybrid filtering algorithm is used to perform filtering operations on CBCT. We set the block size to 2, the search window to 5, and the decay parameter to 10 to obtain the filtered result shown in Fig. 2, the edges of the pulp and enamel were clearer in CBCT, and the contrast between enamel and alveolar bone was enhanced.

Fig. 2
figure 2

The Wavelet + NL-means hybrid filtering results

The PCNN has many unique properties, including pulse coupling, pulse synchronization, multiplicative modulation, and variable thresholds, which has been used for the segmentation of dental images. Since there are too many parameters and the network is miscellaneous [35], we applied the spiking cortical model (SCM) [36] for individual tooth and pulp segmentation for CBCT, as shown in Fig. 3. Compared with the PCNN model, the SCM model simplifies the internal activity term and feedback input part, but it still maintains the feedback input and connection input. The internal activity in the SCM model is more closely related to the external excitation, which better captures the intensity changes in the pulp cavity. However, CBCT images with different spatial resolution, voxel size, and sharpness make it difficult to determine the size of their parameter values. To improve the segmentation accuracy and exclude the deviation of subjective factors, the adaptive parameter rules for \(\alpha_{e}\), \(V_{E}\)\(\beta\) and \(\alpha_{u}\) are designed as follows:

Fig. 3
figure 3

Working principle of the SCM model, \(S_{ij}\) is external excitation, \(F_{ij} \left( n \right)\) is the feedback input, \(L_{ij} \left( n \right)\) is the coupling connection input, \(U_{ij} \left( n \right)\) is the internal activity, \(E_{ij} (n)\) is the dynamic threshold, \(Y_{ij} (n)\) is pulse output, \(\alpha_{e}\) is the amplitude of the \(E_{ij} (n)\), \(\beta\) is the connection strength, \(V_{E}\) is the exponential decay time constant, \(W_{ijkl}\) is the weight matrix for coupled connected domains, \(V_{L}\) is the amplification factor of the coupled connection domain

The value of \(V_{E}\) affects the width of each segmentation area, and the value of \(\alpha_{e}\) affects the attenuation rate of \(E_{ij} (n)\). The smaller the \(\alpha_{e}\), the higher the segmentation accuracy is. The two values can be adjusted by Eqs. (1) and (2).

$$\alpha_{e} = {\raise0.7ex\hbox{$C$} \!\mathord{\left/ {\vphantom {C \mu }}\right.\kern-0pt} \!\lower0.7ex\hbox{$\mu $}}$$
(1)
$$V_{E} = 2\mu$$
(2)

where C is a constant, and \(\mu\) is the average intensity value of the given slice.

The connection strength \(\beta\) between the neighboring neurons: the value of \(\beta\) indicates the degree of interaction between neurons in the neighborhood. The larger the value of \(\beta\) is, makes the \(U_{ij} \left( n \right)\) fluctuate more violently. According to Weber-Fechner law, when the stimulus rises, changes, or the intensity of the stimulus corresponds to the stimulus itself, the sensation will be triggered, and the amount of sensation of the stimulus follows a simple logarithmic rule [37]. The internal mechanism of the synchronous pulse phenomenon is derived from the influence of two different inputs on the internal activity items. The main factor is the connection strength \(\beta\). The subjective sensation of the human eye is roughly logarithmically related to the intensity of the input signal, the \(\beta\) also plays a similar role and can be simulated by (3).

$$\beta = a\ln (\mu )$$
(3)

where \(a\) is an experimental constant.

The exponential decay time constant \(\alpha_{u}\) of the \(U_{ij}\) affects the distribution interval of the \(U_{ij}\). The variance of the image reflects the overall structural characteristics of the image. \(\alpha_{u}\) represents the degree of neuronal memory, and \(\alpha_{u}\) should be different for different images. If the whole image is relatively flat, the variance is small, and the neuron should be easier to maintain the previous state. If the image contains many different blocks, the variance is large, and the neurons should be more difficult to maintain the previous state. The smaller the value of the \(\alpha_{u}\), the wider the distribution range of the internal activity \(U_{ij}\). Therefore, \(\alpha_{u}\) is inversely proportional to the variance of the image, which can be described by the logarithmic function of the reciprocal of the variance of the input image:

$$\alpha_{u} = \log \left[ {1/{\text{var}} (I)} \right]$$
(4)

where \({\text{var}}\) represents the function of calculating variance, \(I\) represents the input image pixel matrix.

The initial matrix \(E(0)\) is usually set to the maximum value of the input pixel matrix. In general, the highest intensity part of the tooth CBCT image is the enamel, and the \(U_{ij}\) must be compared with the maximum value of the dynamic threshold.

$$E(0) = \max (I)$$
(5)

Since the registration using a single tooth easily led to the local optimum, the segmentation of the mandibular molar (No. 36) as well as its neighboring teeth was performed using the adaptive SCM algorithm. The segmentation results are shown in Fig. 4. The segmented pulp and teeth were reconstructed using the marching cube method. Since the roughness of the model surface, we applied the Laplace smoothing method to the CBCT model and transformed it into point cloud data.

Fig. 4
figure 4

CBCT segmentation and pre-processing of target and neighboring teeth

As shown in Fig. 5a), the triangular mesh of the dentition was collected by an oral scanning device and imported into the “Geomagic wrap” software. We adjusted the curve generation coefficients according to the curvature distribution to extract the gingival margin curve. The gingival margin curve is used to segment the target crown mesh by “plane cutting”, “curve extraction”, “curve adjustment” and “curve cutting”. Considering the morphological differences between the two models, the Oriented Bounding Box (OBB) of the crown scanned model is established to intercept the equivalent crown CBCT model. Assuming that the coordinate of the point in the point cloud is \(P_{i} (x_{i} ,y_{i} ,z_{i} )\), the center \(O_{{{\text{op}}}}\) of the OBB is

$$O_{{{\text{op}}}} = \frac{1}{n}\sum\limits_{i = 1}^{n} {P_{i} }$$
(6)
Fig. 5
figure 5

Model preprocessing. a Gingival margin line extraction and crown segmentation of optical scanned crown; b Equivalent processing of crown point cloud

Move the OBB center \({\text{O}}^{op}\)\((o_{x}^{{{\text{op}}}} ,o_{y}^{{{\text{op}}}} ,o_{z}^{{{\text{op}}}} )\) of scanned model to \((o_{x}^{{{\text{ct}}}} ,o_{y}^{{{\text{ct}}}} ,o_{z}^{{{\text{op}}}} )\), where \(o_{x}^{{{\text{ct}}}}\)\(o_{x}^{{{\text{ct}}}}\)\(o_{y}^{{{\text{op}}}}\)\(o_{y}^{{{\text{ct}}}}\)\(o_{z}^{{{\text{op}}}}\) are OBB center coordinates of the scanned model and the CBCT model, respectively. The prongs \({\text{A}}_{{{\text{op}}}} {\text{D}}_{{{\text{op}}}}\) were placed parallel to \({\text{A}}_{{{\text{ct}}}} {\text{D}}_{{{\text{ct}}}}\) and \({\text{A}}_{{{\text{op}}}} {\text{B}}_{{{\text{op}}}}\) parallel to \({\text{A}}_{{{\text{ct}}}} {\text{B}}_{{{\text{ct}}}}\).The OBB of CBCT model was intercepted using the bottom surface \(A{'}_{op}B{'}_{op}C{'}_{op}D{'}_{op}\) We retained the crown of tooth CBCT model, as shown in Fig. 5b).

Registration method

Group 1: registration using the PHMSR method

Since the crown morphology of CBCT and scanned model is not consistent, and a lack of reasonable matching rules, the two types of point clouds cannot be directly matched. Hence, we identified the point set in the crown CBCT point cloud that affects the pulp pose by feature mapping algorithm, increasing the weight of them as well as the neighboring points in the crown registration, and the crown-pulp matching performed by the two crowns registration. The pulp horn is part of the intramedullary extension of the tooth tip into a cusp, and its shape and position are highly similar to the tooth cusp. It is opposite the top of the pulp chamber, which is the closest part to the occlusal surface. The pulp horn is the key feature to determine the pose of the whole pulp cavity. We were inspired by the normal space sampling algorithm [38], and the principle of the feature mapping algorithm is described as follows:

Calculating the feature direction of each pulp horn, estimating the normal vector of the crown surface. The angle between the normal vector and the feature direction of the pulp horn is taken as the mapping indicator, and retaining the points in the crown point cloud with greater correlation with the pulp horn feature. Record the indexes of these points, and surface reconstruction is performed around these points during the registration process. The pulp horn feature mapping (PHFM) algorithm is shown in Algorithm 1. Under different initial deviations between crown models, we can adjust the upper limit \(\gamma_{{{\text{max}}}}\) to change the number of mapping points. The mapping diagram is shown in Fig. 6.

Fig. 6
figure 6

Feature directions of different pulp horns and their mapping results

figure a

Algorithm 1. The PHFM for crown CBCT point cloud

Most of the teeth requiring RCT have caries, and the occlusal surfaces are usually of incomplete and non-smooth morphology. Since the noise and artifacts exist in the actual CBCT images, the slices in the occlusal region are discontinuous [39], and CBCT reconstructed point clouds of the tooth crown surface may produce discontinuous or outlying points with less surface accuracy than optical scanned models. The fine registration should not only be robust to point cloud registration with noise, outliers, and error points, but also use the feature surface as the reference unit of registration, and take into account the pulp pose.

We introduced the point-surface registration method. The basic principle is to calculate the surface patches formed by the corresponding neighborhoods of mapped points on the crown CBCT point cloud, and transform the point-point matching into point-surface matching. To robustly find the matching point pair between the point and surface, the surface is downsampled to obtain dense points. The influence of outliers and noise is reduced by adjusting the surface reconstruction parameters, and the sparse ICP algorithm is used to solve the transformation parameters of the newly established matching point pairs to ensure the final registration accuracy.

To obtain more efficient matching point pairs, we used the quadric surface based on moving least squares (MLS) [40] as a reference surface (Fig. 7a). Compared with implicit surfaces such as the Poisson surface and Bessel surface, the quadric surface reconstruction speed is faster, and the MLS method can improve the local accuracy, and avoid the process of block fitting and smoothing.

Fig. 7
figure 7

Registration algorithm. a MLS-based surface reconstruction principle; b Point-surface registration rules

The matching rules are as follows: As shown in Fig. 7b), Using crown CBCT point set \({\text{C}}\) as a fixed object and optical scanned point set Y as a free object, the nearest point \(c_{h}\) of \({\text{y}}_{h}\) in \({\text{C}}\) is determined. Given that the \(k_{y}\)-nearest neighbor point set of \({\text{y}}_{h}\) is \({\text{N(y}}_{h} {)}\), the \(k_{c}\)\((k_{c} \ge 3k_{y} )\)-nearest neighbor point set of \(c_{h}\) is \({\text{N(y}}_{h} {)}\), which is set to the sample of local surface \(\Lambda\) reconstruction. We can obtain the nearest point \(c^{\prime}_{h}\) from point \(y_{h}\) to surface \(\Lambda\). Taking point \(c^{\prime}_{h}\) as the new matching point of \({\text{y}}_{h}\), that is, \({(}c_{h}^{\prime } {\text{, y}}_{h} {)}\) is a new matching point pair. The algorithm is shown in Algorithm 2.

figure b

Algorithm 2. Search for matching point pairs between points and mapping surfaces

In this algorithm, the \(r_{{_{h} }}^{\Lambda }\) can adjust the range of the point cloud used for the surface fitting, \(r_{h}^{s}\) can adjust the range of dense sampling points on the surface \(\Lambda_{h}\), specifically as follows:

  1. a)

    When the initial deviation between the free and the fixed point cloud is large, the \(r_{{_{h} }}^{\Lambda }\) and the \(r_{h}^{s}\) should be increased, expanding the matching range of the surface \(\Lambda\) to ensure that \(y_{h}\) finds the correct matching point.

  2. b)

    When the initial deviation between them is small, the \(r_{{_{h} }}^{\Lambda }\) and \(r_{h}^{s}\) can be reduced, then the point-surface matching degenerates to the point-point matching, to improve the calculation efficiency.

  3. c)

    When there are noises in the point clouds to be registered, we can select a suitable surface fitting radius \(r_{{_{h} }}^{\Lambda }\) to filter the noise points and generate surfaces on the target point cloud, which improves the robustness of the alignment algorithm. In theory, the sampling span \(d_{h}^{s}\) depends on the average span of the target point cloud and the accuracy requirements. Since the \(d_{h}^{s}\) is too small will lead to a sudden increase in the amount of computation. Therefore, the selection of the parameter should take into account the hardware conditions and select the maximum value under the premise of satisfying registration accuracy.

Based on the point-surface registration rules, the sparse ICP [31] is introduced into the registration process to enhance the robustness of the registration. The sparse ICP algorithm can be expressed by Eq. (7):

$$\left\{ {{\mathbf{R}}, \, {\mathbf{t}}} \right\} = \mathop {\arg \min }\limits_{{{\mathbf{R,t}}}} \sum\limits_{h = 1}^{n} {w^{p - 2} \left\| {{\mathbf{R}}c_{h}^{\prime } + {\mathbf{t}} - y_{h} } \right\|_{2}^{2} } + I_{SO(k)} ({\mathbf{R}})$$
(7)

where \({\mathbf{R}}\) and \({\mathbf{t}}\) are rotation and translation matrices respectively. The intermediate variable \({\mathbf{Z}} = \left\{ {{\mathbf{z}}_{h} \in R^{3} , \, h = 1, \, ..., \, n} \right\}\) is introduced to ensure a stable solution:

$$\left\{ {{\mathbf{R}},{\mathbf{t}}} \right\} = \mathop {\arg \min }\limits_{{{\mathbf{R}},{\mathbf{t}},Z}} \sum\limits_{h = 1}^{n} {\left\| {{\mathbf{z}}_{h} } \right\|_{2}^{p} } + I_{SO(k)} ({\mathbf{R}}){\text{ s}}{\text{.t }}{{\varvec{\updelta}}}_{h} = {\mathbf{0}}$$
(8)

where \({{\varvec{\updelta}}}_{h} = {\mathbf{R}}c_{h} + {\mathbf{t}} - {\text{y}}_{h} - {\mathbf{z}}_{h}\), the augmented Lagrangian method is an effective tool for solving the above-constrained optimization problems. The Eq. 9 can be defined as:

$$L_{A} ({\mathbf{R}},{\mathbf{t}},{\mathbf{Z}},{{\varvec{\Lambda}}}) = \sum\limits_{h = 1}^{n} {\left\| {{\mathbf{z}}_{h} } \right\|_{2}^{p} } + {{\varvec{\uplambda}}}_{h}^{T} {{\varvec{\updelta}}}_{h} + \frac{\mu }{2}\left\| {{{\varvec{\updelta}}}_{h} } \right\|_{2}^{2} + I_{SO(k)} ({\mathbf{R}})$$
(9)

where \({{\varvec{\Lambda}}} = \left\{ {{{\varvec{\uplambda}}}_{h} \in {\mathbb{R}}^{k} ,h = 1...n} \right\}\) is a set of Lagrange multipliers and \(\mu (\mu > 0)\) is the penalty weight. Equation (9) contains four unknowns: rotation matrix \({\mathbf{R}}\), translation matrix \({\mathbf{t}}\), intermediate variable \({\mathbf{z}}\) and coefficient vector \({{\varvec{\Lambda}}}\). The alternating multiplier method (ADMM) is used to optimize this function. The PHMSR registration process is shown in Fig. 8. During the solution process, the matching target point in Step 2 is changed to the intermediate point \(\tilde{y}_{i}\). Through the point-surface matching rule, the closer matching target point and intermediate point are found, which reduces the number of iterations to a certain extent and increases the accuracy.

Fig. 8
figure 8

Flow chart of PHMSR registration algorithm

The above process is programmed in a C +  + environment. In this paper, the \(\gamma_{{{\text{max}}}}\) is set to 20° for small initial deviations, and 25° for large initial deviations. The \(r_{{_{h} }}^{\Lambda }\) \((r_{{_{h} }}^{\Lambda } = r_{h}^{s} )\) is set to 0.1 for small initial deviations and 0.3 for large initial deviations. The parameters of sparse ICP are \(p = 0.5\), \(\mu = 10\), the iteration times of Step 2 is \(n_{2} = 2\), and the number of convergences for step (2.1) is \(n_{s} = 3\).

Group 2: registration using the Sparse ICP method

The source code of Sparse ICP method is from the Reference [31] and is implemented programmatically in a C++ environment. The parameters are set to \(p=0.5\), \(\mu=10\), the iteration times of Step 2 is \(n_{2}=2\) , and the number of convergences for step (2.1) is \(n_{s}=3\).

Group 3: registration using the ICP method

The source code of ICP method is from the Reference [14] and is implemented programmatically in a C++ environment. The maximum number of iterations is set to 100 and the allowable error is 0.1 mm.

Group 4: registration using the CPD method

The source code of the CPD method is from the Reference [32] and is implemented programmatically in a C++ environment.

For each method, two different initial poses (large and small deviation) were set for the same group of crown point clouds for registration, as shown in Fig. 9, the crown CBCT point cloud as a fixed object and the optical scanned point set as a free object, twice registration results for one subject were averaged for evaluation.

Fig. 9
figure 9

Relative pose of the point cloud. a Small initial deviation. b Large initial deviation

Measurement evaluation

The accuracy of the registration was measured using three metrics: Root mean square error (RMSE) between the crown scanned and CBCT crown point cloud, crown-pulp orientation deviation (CPOD), and crown-pulp position deviation (CPPD), and are calculated using Matlab. When the RMSE is smaller, it indicates that the registration accuracy of the crown surface point cloud is higher; when the CPOD and CPPD are smaller, it indicates that the registration accuracy of the crown-pulp point cloud is higher.

The Root mean square (RSME) is commonly used to evaluate the point cloud alignment accuracy [41]. The measurement is formulated by:

$${\text{RSME}} = \sqrt {\frac{{\sum\limits_{h = 1}^{n} {(c_{h} - y_{h} )^{2} } }}{n}}$$
(10)

where \(c_{h}\) denotes the point coordinates in the target point cloud (CBCT), and \(y_{h}\) denotes the \(k_{c}\)-nearest neighbor point of \(c_{h}\) in the source point cloud (optical scanned).

Since there are no relevant metrics regarding crown-pulp registration, we used the relative pose deviation of the optical scanned crown from the pulp CBCT point cloud as an evaluation measurement. The theoretical crown-pulp registration poses as a ground truth, then the crown-pulp Orientation deviation (CPOD) and position deviation (CPPD) were presented. As shown in Fig. 10, we used the PCA algorithm to calculate the centers of the correct pose crown and pulp point clouds, i.e., \({\text{O}}_{{\text{p}}} (x_{{\text{p}}} ,y_{{\text{p}}} ,z_{{\text{p}}} ) \,\) and \({\text{O}}_{{\text{c}}} (x_{{\text{c}}} ,y_{{\text{c}}} ,z_{{\text{c}}} ) \,\), then used them as the origin to establish theoretical coordinate systems, \({\text{O}}_{{\text{c}}} {\text{ - X}}_{{\text{c}}} {\text{Y}}_{{\text{c}}} {\text{Z}}_{{\text{c}}}\) and \({\text{O}}_{{\text{p}}} {\text{ - X}}_{{\text{p}}} {\text{Y}}_{{\text{p}}} {\text{Z}}_{{\text{p}}}\). Established the actual origin \(O{'}_{c}(x{'}_{c},\,y{'}_{c},\,z{'}_{c})\) and coordinate system \(O{'}_{c}-X{'}_{c},\,Y{'}_{c},\,Z{'}_{c}\) of the aligned post-crown. The measurements are formulated by:

$${\text{CPPD}} { = }\sqrt {(x_{a} - x_{t} )^{2} + (y_{a} - y_{t} )^{2} + (z_{a} - z_{t} )^{2} }$$
(11)
$${\text{CPOD}} { = }\sqrt {(\alpha_{a} - \alpha_{t} )^{2} + (\beta_{a} - \beta_{t} )^{2} + (\gamma_{a} - \gamma_{t} )^{2} }$$
(12)

where \(x_{a} = x_{{\text{p}}} - x_{{\text{c}}}\), \(y_{a} = y_{{\text{p}}} - y_{{\text{c}}}\), \(z_{a} = z_{{\text{p}}} - z_{{\text{c}}}\), \(\alpha_{a}\), \(\beta_{a}\), and \(\gamma_{a}\) are the position and orientation deviations of the registered crown scanned point cloud and the pulp CBCT point cloud in the xyz direction. \(x_{t} = x_{{\text{p}}} - x^{\prime}_{{\text{c}}}\), \(y_{t} = y_{{\text{p}}} - y^{\prime}_{{\text{c}}}\), \(z_{t} = z_{{\text{p}}} - z^{\prime}_{{\text{c}}}\), \(\alpha_{t}\), \(\beta_{t}\), and \(\gamma_{t}\) are the theoretical position and orientation deviations of the crown scanned point cloud and the pulp CBCT point cloud.

Fig. 10
figure 10

Evaluation of crown-pulp pose deviation

Statistical analyses

Statistical analyses were performed using SPSS software. Mean, standard deviation, and 95% confidence intervals were calculated for the errors of registration for the eight groups data. A two-tailed paired t-test was used to assess significant differences between each method at p < 0.05.

Results

In the vitro experiments, CBCT and oral scanned models of eight groups (No. 35 and 36 teeth) were registered. As shown in Fig. 11, small surfaces were generated near the mapping points of the CBCT crown point cloud during the PHMSR registration process. When the initial deviation is small (Case I), fewer mapping surfaces and smaller radius are used to reduce the calculation time to reconstruct the surface. When the initial deviation is large (Case II), the number of mapping surfaces and the radius are enlarged to avoid local optimality of the matching and ensure the convergence of the crown registration.

Fig. 11
figure 11

Surface reconstruction results of the PHMSR method in Case I and Case II

As shown in Fig. 12, the crown surface deviation of the four methods did not differ much, and the range of error is approximated. With the pulp attitude unchanged, it is evident that the scanned crowns have a significantly different attitude and position. As shown in Table 1, according to the PHMSR method, the mean CPOD and CPPD deviations were 0.536° ± 0.295° and 0.064 ± 0.055 mm, respectively, which showed that the crown-pulp registration accuracies were significantly superior to the other conventional method. The difference of CPOD error between PHMSR group and other groups was statistically significant (p < 0.05), and the difference of CPPD error between PHMSR group, CPD group and ICP group was statistically significant.

Fig. 12
figure 12

The registration results of four methods. A darker color in the color deviation chart indicates a larger registration error in these areas

Table 1 Statistical analysis of registration accuracy of different methods

As shown in Fig. 13a), the CPOD of the PHMSR, Sparse ICP, ICP, and CPD method are distributed in the ranges of 0.289 to 0.783°, 2.535 to 5.630°, 2.535 to 5.630°, and 1.107 to 2.223°. The CPOD distribution interval of Sparse ICP is the largest, and the maximum error reaches 6.992°, indicating the poor stability of the method. The PHMSR method had the smallest error distribution interval and the maximum error was 1.017°. The maximum CPOD errors of ICP and CPD were close to each other with 3.092° and 2.7146°, respectively. As shown in Fig. 13b), the CPPD of PHMSR, Sparse ICP, ICP, and CPD were distributed in 0.018 ~ 0.110 mm, 0.038 ~ 0.228 mm, 0.434 ~ 0.966 mm and 0.569 ~ 0.867 mm. The distribution interval of the CPPD for ICP was the largest, and the maximum deviation of the ICP and CPD were 0.806 mm and 0.725 mm, respectively. In summary, the crown-pulp registration error distribution of PHMSR method is more centralized.

Fig. 13
figure 13

Comparison of registration errors of four methods. a CPOD registration error. b CPPD registration error. c RSME registration error

As shown in Fig. 13b) and Table 1, The PHMSR, Sparse ICP, ICP, and CPD method with error ranges of 0.238 ~ 0.261 mm, 0.243 ~ 0.271 mm, 0.178 ~ 0.256 mm, and 0.175 ~ 0.243 mm, respectively. The RSME of PHMSR and Sparse ICP is similar, and the RSME error of PHMSR was slightly larger than that of the ICP and CPD methods, but there was no significant difference in the accuracy of the crown registration with Sparse ICP (p = 0.3411) and ICP (p = 0.7985).

In terms of computational time, the registration time of PHMSR is more than that of the sparse ICP and ICP methods, but less than that of the CPD method.

Discussion

PHMSR algorithm is an important innovation in this paper. By finding the mapping relationship between pulp horn and crown, small surfaces are reconstructed near the mapping point of the crown CBCT point cloud to increase the accuracy of crown-pulp registration, and propose the corresponding accuracy indexes, which solves the problem of poor applicability of the generalized method in dental crown-pulp registration. Since the point cloud of the optical scanned and the CBCT reconstruction point cloud are not exactly the same in morphology, number, and density, the optimal registration is probably not point-to-point, and the traditional methods use the original corresponding points for the registration, thus the registration error of the crown surface is small, and the crown-pulp pose deviation is large. For our proposed method, thanks to the reconstruction of the mapping surface, the target points matching the points in the source point cloud are newly created point pairs, and thus the registration error of the crown surface is slightly larger than the other three methods, but the crown-pulp relative pose is adequately taken into account. In addition, the stability of the PHMSR algorithm (SD(CPOD) = 0.295° and SD(CPPD) = 0.055 mm) was significantly higher than that of Sparse ICP, ICP, and CPD algorithms.

The registration is the digital preparation before pulp opening, the registration deviation directly affects the final positioning result. From the clinical studies [15,16,17,18,19,20,21,22], it can be found that the surface registration errors in the maxillary and mandibular ranged from 0.13 to 0.345 mm, which means that the RSME deviations (0.1434 to 0.2295 mm) of the four methods in our study are acceptable in the clinic, and the reasons for the discrepancies may include the differences in the quality of the CBCT, the accuracy of the segmentation, and the accuracy of the oral scanned model. Taking the mean root deviation (0.1678 ± 0.3178 mm) before and after registration reported by Lee et al. [17] as a reference, it can be seen that the deviation of the crown-pulp position (0.064 ± 0.055 mm) of the PHMSR method is perfectly acceptable. At present, there is no research on crown-pulp registration, therefore, the relationship between the use of digital technology to assist root canal opening and registration accuracy is discussed. Zehnder et al. studied the digital guide-assisted root canal treatment of upper anterior teeth. The distance deviation was 0 ~ 1.59 mm, and the angle deviation was 0 ~ 5.60° [42]. The mean apical deviation obtained by Jacobs et al. [8] was 0.10 ~ 1.88 mm, and the angle deviation was 0.18 ~ 5.87°. The distance deviation of the drill needle obtained by Zhang et al. [9] is 0.183 ~ 0.537 mm, and the angle deviation is 1.770° ~ 3.260°. The mean distance deviation of the apical obtained by Torres et al. [10] was 0.63 ± 0.35 mm, and the mean angle deviation was 2.81° ± 1.53°. Taking study [10] as an example, the mean crown-pulp registration deviation (0.536° and 0.064 mm) of PHMSR accounted for 10.16% of the distance deviation and 19.07% of the angle deviation. The crown-pulp orientation deviation (4.089°) of Sparse ICP exceeded the surgical deviation range, and the position deviation (0.133 mm) accounted for 21.11% of the surgical deviation. The crown-pulp deviation of ICP (1.787°) accounted for 63.59% of the surgical deviation, and the crown-pulp deviation of CPD (1.665°) accounted for 59.25% of the surgical angle deviation. Their position deviation exceeded the range of surgical distance deviation. It can be seen that our PHMSR method can have the least impact on surgical planning, thus ensuring the accuracy of surgery.

During the registration, the PHMSR method needs to traverse the neighboring points of the mapped point cloud to fit the new surface and then densely downsample the surface, this process is computationally intensive, resulting in a longer time compared to sparse ICP and ICP, but the time cost is still much smaller than the CPD method.

Although it is the first implementation that the pulp pose is indirectly introduced into the crown registration, there are some limitations to our study: First, the sample size of the study was only 8 cases, and all of them were young people in China, whose dental pulp had not been reduced or calcified. In addition, we mainly used molar teeth for the study, which have four pulp horns. For the teeth with anatomical variation and pulp injury, the feature direction of the pulp horn may be difficult to reflect the information of pulp pose, so the clinical applicability of the algorithm needs to be further studied. Secondly, PHMSR method can be regarded as an improved version of sparse ICP, including parameters such as the number of mapping points and the radius of surface reconstruction. Parameter tuning and the selection of pulp horn still requires manual interaction. Therefore, the automatic parameter setting method will be further studied in the future.

Conclusions

Within the limitations of this vitro study, it was concluded that the PHMSR method has a smaller crown-pulp registration accuracy and a clinically acceptable deviation range than conventional methods, which may be used for clinical virtual planning of inlay-guided root canal therapy, although further clinical studies are needed to confirm our results.

Availability of data and materials

The datasets used and analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CBCT:

Cone Beam Computed Tomography

3D:

Three-Dimensional

PHMSR:

Registration method based on pulp horn mapping surface

2D:

Two-Dimensional

STL:

Standard tessellation language

DICOM:

Digital imaging and communications in medicine

SCM:

Spiking cortical model

RMSE:

Root mean square error

CPD:

Coherent point drift

MLS:

Moving least squares

ICP:

Iterative closest point

CPOD:

Crown-pulp orientation deviation

CPPD:

Crown-pulp position deviation

PCL:

Point cloud library

References

  1. Hildebrand H, Leontiev W, Krastl G, Weiger R, Dagassan-Berndt D, Bürklein S, Connert T. Guided endodontics versus conventional access cavity preparation: an ex vivo comparative study of substance loss. BMC Oral Health. 2023;23(1):713.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Shao TF, Guan R, Zhang C, Hou BX. Influence of operator’s experience on complications of root canal treatment using contemporary techniques: a retrospective study. BMC Oral Health. 2024;24(1):96.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Kfir A, Telishevsky-Strauss Y, Leitner A, Metzger Z. The diagnosis and conservative treatment of a complex type 3 dens invaginatus using cone beam computed tomography (CBCT) and 3D plastic models. Int Endod J. 2013;46(3):275–88.

    Article  CAS  PubMed  Google Scholar 

  4. Jang TJ, Yun HS, Hyun CM, Kim JE, Lee SH, Seo JK. Fully automatic integration of dental CBCT images and full-arch intraoral impressions with stitching error correction via individual tooth segmentation and identification. Med Image Anal. 2024;93:103096.

    Article  PubMed  Google Scholar 

  5. Woo S, Lee S, Chae J, Rim J, Lee J, Seo J, Lee C. Automatic matching of computed tomography and stereolithography data. Comput Meth Prog Bio. 2019;175:215–22.

    Article  CAS  Google Scholar 

  6. Bernardo BF, Stephane V, Eduarda HLN, Marcus F, Israel C. Guided endodontic access in a calcified central incisor: a conservative alternative for endodontic therapy. Iran Endod J. 2021;16(1):56–9.

    Google Scholar 

  7. Yan YQ, Wang HL, Liu Y, Zheng TJ, Tang YP, Liu R. Three-dimensional inlay-guided endodontics applied in variant root canals: a case report and review of literature. World J Clin Cases. 2021;9(36):11425–36.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Torres A, Dierickx M, Coucke W, Pedano MS, Lambrechts P, Jacobs R. In vitro study on the accuracy of sleeveless guided endodontics and treatment of a complex upper lateral incisor. J Dent. 2023;131:104466.

    Article  CAS  PubMed  Google Scholar 

  9. Zhang T, Chen D, Miao LY, Xie SJ, Tang XN. Guided endodontic access of calcified root canal by laser melting templates. Hua Xi Kou Qiang Yi Xue Za Zhi. 2020;38(5):525–31.

    PubMed  Google Scholar 

  10. Torres A, Boelen GJ, Lambrechts P, Pedano MS, Jacobs R. Dynamic navigation: a laboratory study on the accuracy and potential use of guided root canal treatment. Int Endod J. 2021;54(9):1659–67.

    Article  PubMed  Google Scholar 

  11. Mai HY, Lee D. Time efficiency and operator convenience of using a micro-screw in image registration for guided implant surgery. J Korean Acad Prosthodont. 2019;57(3):219–24.

    Article  Google Scholar 

  12. Wu BZ, Sun F. The impacts of registration-and-fixation device positioning on the performance of implant placement assisted by dynamic computer-aided surgery: a randomized controlled trial. Clin Oral Implants Res. 2024;35(4):386-95.

  13. Wei T, Ma FF, Sun F, Ma Y. Assessment of the accuracy of two different dynamic navigation system registration methods for dental implant placement in the posterior area: an in vitro study. J Pers Med. 2023;13(1):139.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Besl PJ, McKay ND. A method for registration of 3-D shapes. IEEE Trans Pattern Anal Mach Intell. 1992;14(2):239–56.

    Article  Google Scholar 

  15. Kim BC, Lee CE, Park W, Kang SH, Zhengguo P, Yi CK, et al. Integration accuracy of digital dental models and 3-dimensional computerized tomography images by sequential point- and surface-based markerless registration. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;110:370–8.

    Article  PubMed  Google Scholar 

  16. Lim SW, Hwang HS, Cho IS, Baek SH, Cho JH. Registration accuracy between intraoral-scanned and cone-beam computed tomography-scanned crowns in various registration methods. Am J Orthod Dentofac. 2020;157(3):348–56.

    Article  Google Scholar 

  17. Lee RJ, Pham J, Choy M, Weissheimer A, Dougherty HL, Sameshima GT, Tong HS. Monitoring of typodont root movement via crown superimposition of single cone-beam computed tomography and consecutive intraoral scans. Am J Orthod Dentofac. 2014;145(3):399–409.

    Article  Google Scholar 

  18. Ye N, Long H, Xue J, Wang S, Yang X, Lai W. Integration accuracy of laser-scanned dental models into maxillofacial cone beam computed tomography images of different voxel sizes with different segmentation threshold settings. Oral Surg Oral Med Oral Pathol Oral Radiol. 2014;117:780–6.

    Article  PubMed  Google Scholar 

  19. Guo H, Zhou J, Bai Y, Li S. A three-dimensional setup model with dental roots. J Clin Orthod. 2011;45(4):209–16.

    PubMed  Google Scholar 

  20. Sun L, Hwang HS, Lee KM. Registration area and accuracy when integrating laser-scanned and maxillofacial cone-beam computed tomography images. Am J Orthod Dentofacial Orthop. 2018;153:355–61.

    Article  PubMed  Google Scholar 

  21. Yi C, Li S, Wen AN, Wang Y, Zhao YJ, Zhang Y. Digital versus radiographic accuracy evaluation of guided implant surgery: an in vitro study. BMC Oral Health. 2022;22(1):540.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Rangel FA, Maal TJJ, de Koning MJJ, Bronkhorst EM, Berge SJ, Kuijpers-Jagtman AM. Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study. Clin Oral Investig. 2018;22(3):1215–22.

    Article  PubMed  Google Scholar 

  23. Bai SF, Liao SH, Han FC, Pox XY, Li ZL, Yang WP, Wu RZ. Automatic real-time space registration application for simulating dental and maxillofacial surgery. J Carniofac Surg. 2022;33(6):1698–704.

    Article  Google Scholar 

  24. Yau HT, Yang TJ, Chen YC. Tooth model reconstruction based upon data fusion for orthodontic treatment simulation. Comput Biol Med. 2014;48:8–16.

    Article  PubMed  Google Scholar 

  25. Zhang DX, Gan YZ, Xiong J, Xia ZY. Three-dimensional tooth model reconstruction based on fusion of dental computed tomography images and laser-scanned images. J Biomed Eng. 2017;34(1):7–14.

    CAS  Google Scholar 

  26. Dai N, Li DW, Yang W, Cheng C, Sun YC. Research and primary evaluation of an automatic fusion method for multisource tooth crown data. Int J Numer Meth Biomed. 2017;33(11):e2878.

    Article  Google Scholar 

  27. Wang G, Feng Y. A New Approach for Tooth Fusion between Cone Beam Computed Tomography Images and Crown Models. ISICDM 2018. In: Proceedings of the 2nd International Symposium on Image Computing and Digital Medicine. 2018. p. 148–54.

    Google Scholar 

  28. Qian JH, Lu SJ, Gao Y, Tao YB, Lin J, Lin H. An automatic tooth reconstruction method based on multimodal data. J Vis. 2021;24(1):205–21.

    Article  Google Scholar 

  29. Chung MY, Lee J, Song W, Song Y, Yang IH, Lee J, et al. Automatic registration between cone-beam CT and scanned surface via deep-pose regression neural networks and clustered similarities. IEEE Trans Med Imaging. 2020;39(12):3900–9.

    Article  PubMed  Google Scholar 

  30. Bouaziz S, Tagliasacchi A, Pauly M. Sparse iterative closest point. Comput Graph Forum. 2013;32(5):113–23.

    Article  Google Scholar 

  31. Myronenko A, Song XB. Point set registration: coherent point drift. IEEE Trans Pattern Anal Mach Intell. 2010;32(12):2262–75.

    Article  PubMed  Google Scholar 

  32. Faul F, Erdfelder E, Buchner A, Lang AG. Statistical power analyses using G*Power 3.1: tests for correlation and regression analyses. Behav Res Methods. 2009;41:1149–60.

    Article  PubMed  Google Scholar 

  33. Chen ZK, Ning KL. Breast volume denoising and noise characterization by 3D wavelet transform. Comput Med Imag Grap. 2004;28(5):235–46.

    Article  Google Scholar 

  34. Dabov K, Foi A, Katkovnik V, Egiazarian K. Image denoising by sparse 3D transform-domain collaborative filtering. IEEE Trans Signal Process. 2007;16(8):2080–95.

    Google Scholar 

  35. Dong Z, Lai CS, Qi DL, Xu Z, Li CY, Duan SK. A general memristor-based pulse coupled neural network with variable linking coefficient for multi-focus image fusion. Neurocomputing. 2018;2018(308):172–83.

    Article  Google Scholar 

  36. Zhan K, Zhang HJ, Ma YD. New spiking cortical model for invariant texture retrieval and image processing. IEEE Trans Neural Netw. 2009;20(12):1980–6.

    Article  PubMed  Google Scholar 

  37. Chen YL, Park SK, Ma Y, Ala R. A new automatic parameter setting method of a simplified PCNN for image segmentation. IEEE Trans Neural Netw. 2011;22(6):880–92.

    Article  PubMed  Google Scholar 

  38. Kwok TH. DNSS: Dual-normal-space sampling for 3-D ICP registration. IEEE Trans Autom Sci Eng. 2019;16(1):241–52.

    Article  Google Scholar 

  39. Yau HT, Yang TJ, Chen YC. Tooth model reconstruction based upon data fusion for orthodontic treatment simulation. Comput Biol Med. 2014;48:8–16.

    Article  PubMed  Google Scholar 

  40. Zhang L, Gu TQ, Zhao J, Ji SJ, Hu M, Li XB. An improved moving least squares method for curve and surface fitting. Math Probl Eng. 2013;2013:159694.

    Article  Google Scholar 

  41. Ge XM. Non-rigid registration of 3D point clouds under isometric deformation. Isprs J Photogramm. 2016;121:192–202.

    Article  Google Scholar 

  42. Zehnder MS, Connert T, Weiger R, et al. Guided endodontics: accuracy of a novel method for guided access cavity preparation and root canal location. Int Endod J. 2016;49(10):966–72.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors thank dental technician Jingchao Wang from the 2nd Affiliated Hospital of Harbin Medical University for their support with digital techniques.

Funding

This work was supported in part by the Natural Science Foundation of Heilongjiang Province under Grant LH2021E081, in part by the Research Launch Foundation for Heilongjiang Province Postdoctoral, and in part by the China Postdoctoral Science Foundation Special Funded Project under Grant 2018T110313.

Author information

Authors and Affiliations

Authors

Contributions

All authors made substantial contributions to the present study. WDH contributed to validation, writing-original draft, software practice and visualization of the manuscript; JJG contributed to conceptualization, methodology, supervision and funding acquisition of the manuscript; WJK contributed to Investigation, review and editing of the manuscript; ZS contributed to data curation, supervision and validation of the manuscript; QK contributed to supervision of the manuscript; All authors have read and approved the manuscript.

Corresponding author

Correspondence to Jingang Jiang.

Ethics declarations

Ethics approval and consent to participate

The research was authorized by the Medical Ethics Committee of the 2nd Affiliated Hospital of Harbin Medical University (YJSKY2022-115), and informed consent was obtained from all subjects whose scans were used as a study material.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wu, D., Jiang, J., Wang, J. et al. Accuracy evaluation of dental CBCT and scanned model registration method based on pulp horn mapping surface: an in vitro proof-of-concept. BMC Oral Health 24, 827 (2024). https://doi.org/10.1186/s12903-024-04565-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-024-04565-3

Keywords