Evaluation of micromotion in multirooted root analogue implants embedded in synthetic bone blocks: an in vitro study

Abstract

Background

While conventional threaded implants (TI) have proven to be effective for replacing missing teeth, they have certain limitations in terms of diameter, length, and emergence profile when compared to customised root analogue implants (RAI). To further investigate the potential benefits of RAIs, the aim of this study was to experimentally evaluate the micromotion of RAIs compared to TIs.

Methods

A 3D model of tooth 47 (mandibular right second molar) was segmented from an existing cone beam computed tomography (CBCT), and a RAI was designed based on this model. Four RAI subgroups were fabricated as follows: 3D-printed titanium (PT), 3D-printed zirconia (PZ), milled titanium (MT), milled zirconia (MZ), each with a sample size of nā€‰=ā€‰5. Additionally, two TI subgroups (B11 and C11) were used as control, each with a sample size of nā€‰=ā€‰5. All samples were embedded in polyurethane foam artificial bone blocks and subjected to load application using a self-developed biomechanical Hexapod Measurement System. Micromotion was quantified by analysing the load/displacement curves.

Results

There were no statistically significant differences in displacement in Z-axis (the loading direction) between the RAI group and the TI group. However, within the RAI subgroups, PZ exhibited significantly higher displacement values compared to the other subgroups (pā€‰<ā€‰0.05). In terms of the overall total displacement, the RAI group showed a statistically significant higher displacement than the TI group, with mean displacement values of 96.5Ā Āµm and 55.8Ā Āµm for the RAI and TI groups, respectively.

Conclusions

The RAI demonstrated promising biomechanical behaviour, with micromotion values falling within the physiological limits. However, their performance is less predictable due to varying anatomical designs.

Peer Review reports

Background

The average survival rate of dental implants after ten years of clinical use is almost 95%, establishing them the best treatment choice for replacing missing and severely decayed teeth [1]. The International Team for Implantology (ITI) consensus conference has classified dental implants based on the insertion protocol as follows: a) immediate implant placement on the day of tooth extraction, b) early implant placement with soft tissue healing typically occurring after 4 to 8Ā weeks, c) early implant placement with partial bone healing taking place approximately 12 to 16Ā weeks later, and d) late implant placement after complete bone healing after of least 6Ā months [2].

Immediate implant placement has many advantages, including shortening the overall treatment time, reducing costs, and decreasing the number of surgical interventions [3]. Additionally, it helps preserving the height and width of the alveolar bone while minimising marginal bone loss following extraction [3,4,5]. However, the decreased primary stability due to the incongruence between the implant and the alveolus, as well as the difficult implant placement can be a real surgical challenge [6,7,8].

One of the treatment alternatives to conventional threaded implants (TIs) is the use of fully customised root-analogue implants (RAIs) [9]. The concept of RAI was initially introduced by Hodosh et al. in 1969 as a heat-processed methyl methacrylate implant. However, it was deemed unsuccessful after failure to achieve osseointegration [10]. In 1992, the technique was reintroduced using pure titanium instead of polymer, leading to successful osseointegration [7].

Such RAIs are the product of the combined technologies of computer-aided design/computer-aided manufacturing (CAD/CAM) and cone beam computed tomography (CBCT) [6, 11]. The idea of the RAI is to replace a tooth scheduled for extraction through an immediate implant placement by designing the RAI with similar dimensions to the original root anatomy based on CBCT. Thus, a perfect congruence between the implant and the empty socket could be achieved, unlike TI [12]. The expected benefits include a reduced number of surgeries, simple and straightforward placement, improved primary stability and immediate soft tissue support [13,14,15].

RAIs are produced using either subtractive or additive manufacturing techniques [16]. Subtractive manufacturing employs a milling process facilitated by computer numerical control (CNC) milling machines. These machines are categorized based on the number of axes they operate on, ranging from 3-axis to 5-axis machines [17, 18]. On the other hand, additive manufacturing involves 3D printing, a method that transforms a digital model into a physical object by depositing materials layer by layer [19, 20].

A notable advancement in additive manufacturing is the use of lithography-based ceramic manufacturing (LCM) for 3D printing advanced ceramics. In this process, a ceramic slurry coated with a photosensitive resin is hardened layer by layer using a light-emitting diode device [21]. Another significant technique in additive manufacturing is selective laser melting (SLM), which involves fusing powdered metal using a high-power fibre laser in an inert environment. The laser precisely melts each layer onto the preceding one, creating a solid object through the accumulation of thousands of micro-welds [22, 23].

The stability of the implant in the alveolar bone is of crucial importance for successful osseointegration [24]. The term micromotion in dental implants refers to the subtle displacement of an implant in relation to the surrounding tissue, which cannot be observed with the naked eye [25]. Studies have suggested that for successful osseointegration, the micromotion between the implant and bone should not exceed a threshold value of 150Ā Āµm [26, 27]. Implant stability can be classified into primary and secondary stability. Primary stability is achieved by the mechanical retention of the implant during initial insertion, whereas secondary stability is reached after consecutive bone remodelling processes and complete healing. Consequently, primary stability is considered a mechanical phenomenon, while secondary stability is a biological phenomenon influenced by osseointegration [28]. Many factors influence the primary stability of the implant, such as the quality and quantity of the surrounding bone and the implant geometry; changing the implant-bone contact area by increasing the length or width of the implant could enhance the primary stability [9, 29].

The biomechanical behaviour of dental implants has been extensively investigated in various studies [26, 30, 31]. However, there is a notable gap in research regarding the specific biomechanical behaviour of RAIs. Hence, the aim of this study was to experimentally evaluate the micromotion of multi-rooted titanium and zirconia RAIs by analysing their load/displacement curves. The null hypothesis stated that there would be no statistically significant difference in micromotion among the examined groups or subgroups (pā€‰>ā€‰0.05).

Methods

Study design

Two implant designs were used in this study: a custom-designed RAI and a traditional TI as a control. The RAI was designed based on a CBCT scan of a dentate mandible using the following scanning parameters: beam accelerating voltage of 90Ā kV, X-ray current of 12Ā mA, voxel dimension of 75Ā Āµm, and total scanning time of 15Ā s. The total number of slices was 668. The CBCT scan was processed using a 3D medical image processing software (Mimics 22; Materialise, Leuven, Belgium) and the right mandibular second molar (tooth 47) was segmented based on histogram analysis. The segmented tooth was subsequently imported into a 3D modelling software (3-matic 15; Materialise, Leuven, Belgium), to finalise the design of the RAI based on the anatomy of the tooth. The coronal portion was designed as an idealised cube with a side length of 5Ā mm to facilitate further biomechanical investigations.

The RAIs were produced using two methods: milling and 3D printing, with both titanium and zirconia as materials of choice. Milling was performed with 5-axis CNC milling machines, whereas LCM and SLM technologies were used for the 3D printing of zirconia and titanium RAIs, respectively. During 3D printing, the printing supports were placed on the overhanging coronal surface around the idealised cube, and the layer thickness was set to 25Ā Āµm. A total of 20 RAIs were fabricated and categorized into four subgroups (nā€‰=ā€‰5) based on the manufacturing method: 3D printed titanium (PT), 3D printed zirconia (PZ), milled titanium (MT), and milled zirconia (MZ).

As a control group, conventional TI were included (Ankylos; Dentsply-Friadent, Mannheim, Germany). The TI group was divided into two subgroups (nā€‰=ā€‰5) based on implant size: B11 subgroup with a diameter of 4.5Ā mm and length of 11Ā mm, and C11 subgroup with a diameter of 5.5Ā mm and length of 11Ā mm (Fig.Ā 1).

Fig. 1
figure 1

Overview of the examined implant groups. A Root analogue implant (RAI) group, including PT, PZ, MT, and MZ subgroups. B Threaded Implant (TI) group, comprising C11 and B11 subgroups. Ī¦, Diameter; L, Length

Biomechanical analysis

For the biomechanical testing, the implants were inserted into test blocks made of polyurethane foam artificial bone (Sawbones; Pacific Research Laboratories, Vashon, USA). These test blocks were comprised of two layers: a 2Ā mm-thick layer of epoxy filled with short glass fibres (type #3401ā€“01, density 1.64Ā g/cm3), which simulated cortical bone, and a 40Ā mm-thick layer of rigid polyurethane foam (type #1522ā€“01, density 0.16Ā g/cm3), which simulated cancellous bone. The TIs were screwed into Sawbones following the surgical protocol provided by the manufacturer (Fig.Ā 2A). As for the RAIs, a socket-shaped cavity, resembling the negative replica of the root-shaped RAI, was initially drilled in the Sawbones (Fig.Ā 2B). Prior to insertion into the Sawbones, the surface of the RAIs was coated with a thin layer of resin (PalaXpress; Heraeus Kulzer GmbH, Wasserburg, Germany) to secure a tight attachment between the RAI and the Sawbones. Each specimen was then firmly fastened to the base of the specimen holder using PalaXpress resin (Fig.Ā 2C).

Fig. 2
figure 2

Preparation of specimens for biomechanical testing. AĀ Preparation of the osteotomy of the polyurethane foam block to insert the TI. B Preparation of the osteotomy of the polyurethane foam block to insert the RAI. C Secure fixation of the specimen to the specimen holder using resin. The metal structure on top of the specimen holder holds the aluminium cube in place during preparation, and is removed before measuring the specimen

The samples were inserted into a custom-developed biomechanical Hexapod Measurement System (HexMeS) [32]. HexMeS is specifically designed to apply various forces on small objects like dental implants. It consists of three main components: a high-precision hexapod robot (PI M-850.50; Physik Instrumente, Karlsruhe, Germany) capable of precise translations and rotations with a resolution of less than 1Ā Āµm and 5 Āµrad. Additionally, the system incorporates a high-precision 3D force/torque sensor (ATI FTSGamma 130/10; SCHUNK GmbH & Co. KG, Lauffen/Neckar, Germany) for force and torque measurements, as well as an optical system for precise position detection, consisting of an aluminium cube with three pinholes, each 2Ā Āµm in diameter (Melles-Griot, Bensheim, Germany) illuminated by a laser beam (35 mW, 658Ā nm; Laser 2000, Wessling, Germany), and three video cameras with macro zoom optics (JAI CV-M1; Stemmer-Imaging, Puchheim, Germany). This setup enables the accurate tracking of micromotions in the specimens under load application by monitoring the pinholes through the video cameras (Fig.Ā 3).

Fig. 3
figure 3

Schematic representation of the Hexapod Measurement System (HexMeS)

The samples were mounted on the HexMeS with the implant aligned parallel to the Z-axis. The laser-illuminated aluminium cube was securely attached to the top of the samples, and a spoon-shaped lever arm was connected to the implant. This configuration allowed any movement of the Hexapod to be transmitted as a force to the implant (Fig.Ā 4).

Fig. 4
figure 4

Experimental setup for biomechanical testing. The specimen was securely mounted on the HexMeS system with the implant aligned parallel to the Z axis. An aluminium cube, illuminated by a laser, was attached to the specimen. A loading cycle of 1.5Ā mm translation in the negative Z direction was applied, followed by a releasing cycle. The resultant force and torque were accurately recorded by the force/torque sensor

The samples were indirectly loaded by programming the Hexapod to perform a loading cycle of 1.5Ā mm translation in the negative Z direction, followed by a release cycle in the positive Z direction until reaching the zero position (150 steps of 0.02Ā mm each). The applied force and torque were recorded by the force/torque system, and simultaneously the displacement of the implant (translation and rotation) was recorded by tracking the laser-illuminated pinholes through the video cameras. The collected data were exported in CSV format (comma separated values) for further data analysis.

Statistical analysis

The numerical data are represented as mean values and standard deviations. The normality of the data was assessed using Shapiroā€“Wilkā€™s test, while Levene's test was employed to test for homogeneity of variances. The data showed a parametric distribution, homogeneity of variances, and were analysed using nested ANOVA. Estimated marginal means were compared using t-test with p-value adjustment using Tukeyā€™s method. A significance level of pā€‰ā‰¤ā€‰0.05 was chosen for all statistical tests. The statistical analysis was performed using R statistical analysis software, version 4.1.3 for Windows Footnote 1.

Results

The maximum magnitude of forces recorded by the force sensor ranged from 64 to 96 N, whereas the produced maximum displacements ranged from 40Ā Āµm to 178Ā Āµm. Owing to the different magnitudes of forces, a maximum force of 50 N was chosen for the different specimens to include all specimens (Fig.Ā 5).

Fig. 5
figure 5

Line chart showing load/displacement curves for all specimens. The dashed line represents the maximum force (50 N) chosen to include all the specimens

The HexMeS allows for the precise measurement of micromovements in each sample, by tracking both the translations and rotations in the three spatial directions. Particular attention will be given to the displacement in the loading direction (Z-axis), as well as the total displacement. The mean values and the standard deviation for all measured displacements in loading direction at a force of 50 N are shown in TableĀ 1.

Table 1 Displacement (Z) and total displacement values for both groups and subgroups

The nested ANOVA model for displacement (Z) revealed that there was no statistically significant difference between the two parent groups. The mean displacement for the TI group was 44Ā Āµm compared to 48Ā Āµm for the RAI group. However, a significant interaction within the nested subgroup variable (pā€‰=ā€‰0.002) was observed. Notably, among the RAI subgroups, PZ exhibited the highest displacement value of 71Ā Āµm, while the differences in displacement between the PT, MT, and MZ subgroups were statistically insignificant (Fig.Ā 6 and TablesĀ 1 and 2).

Fig. 6
figure 6

Boxplot diagrams illustrating displacement (Z) and total displacement values across groups and subgroups

Table 2 Nested ANOVA model for displacement (Z) and total displacement

Comparisons of estimated marginal means presented in TableĀ 3 and Fig.Ā 7 indicated that there were no statistically significant differences among the parent groups (pā€‰=ā€‰0.325) or the two subgroups within the TI group (pā€‰=ā€‰0.964). However, within the RAI group, PZ had significantly higher values than the other subgroups (pā€‰<ā€‰0.05).

Table 3 Comparison of estimated marginal mean for displacement (Z)
Fig. 7
figure 7

Interval plot showing the variation in estimated marginal means for displacement along the Z-axis

Regarding total displacement, the nested ANOVA model showed that the RAI group had a statistically significantly higher displacement compared to the TI group (pā€‰=ā€‰0.013), with mean displacement values of 96Ā Āµm and 56Ā Āµm for the RAI and TI groups, respectively. However, the effect of the nested subgroup variable was not statistically significant (pā€‰=ā€‰0.133) (Fig.Ā 6 and TablesĀ 1 and 2).

Discussion

The objective of this study was to compare the biomechanical behaviour of multi-rooted RAIs and TIs. In vitro load/displacement curves were analysed to assess the micromotion of both implant designs. Based on the study findings, the null hypothesis, which implies no statistically significant difference in micromotion among the examined groups or subgroups, was partially rejected.

One of the determining key factors for successful implant placement is implant stability, whether it is assessed immediately after implant placement or after osseointegration. The stability of dental implants has been evaluated in literature by various methods including PeriotestĀ® and resonance frequency analysis [29]. In the present in vitro study, we have used micromotion as an indicator of implant stability by comparing the produced displacements of TI and RAI under specific loading conditions [31, 33].

Our objective was to replicate the clinical process of placing RAI by utilizing CBCT data acquisition and CAD software to design the RAI according to the natural root shape of tooth 47 [34]. We have specifically chosen a multirooted tooth to explore the intricate mechanical characteristics of such teeth, which have not been thoroughly examined in prior research. The two most common materials used in the manufacturing of RAIs are titanium alloy and zirconia [16, 35]. Owing to the biocompatibility and the remarkable mechanical and physical properties of titanium, it has been widely used for dental implants [36]. Nonetheless, the increasing emphasis on aesthetics has led to the emergence of zirconia as a viable alternative [37]. Zirconia exhibits high biocompatibility, superior flexural strength, reduced bacterial affinity, and the advantage of adjustable white colour [36]. In our study, both titanium alloy and zirconia were selected as materials for the RAIs, using both additive and subtractive manufacturing methods.

Sawbones artificial bone blocks were used instead of cadaver bone to take advantage of their uniform and standardised physical properties. This reduced variability and eliminated the special handling requirements associated with cadaver bone. PalaXpress resin was chosen to fix the RAIs in the Sawbones, owing to its appropriate working time, stability, and radiopacity, as previously reported in the literature [31]. In contrast, the TIs were firmly inserted into the drilled Sawbones without requiring any resin application.

The results of this study revealed that there was no statistically significant difference in displacements along the loading direction (Z-axis) between the RAIs and the TIs, suggesting comparable stability between the two implant types. These findings are consistent with a study by Gattinger et al. [38], where they compared by finite element analysis the micromotion of RAI and standard implant and reported that the RAI was as good as the standard threaded implant in terms of micromotion. A similar conclusion was reached out by Chen et al. [39] who studied the biomechanical performance of RAI for both the immediate and the delayed loading protocols. They observed increased micromotion in the RAI during immediate loading, but reduced micromotion during the osseointegrated phase with bonded contact simulation, indicating reliable long-term stability.

Based on the findings of a previous study conducted by Aldesoki et al. [34], it was observed that the manufacturing method had a slight impact on the dimensions of the produced RAIs. Taking this into consideration, our study incorporated four RAI subgroups that comprised different combinations of manufacturing techniques and materials. The analysis of estimated marginal means revealed a statistically significant higher displacement in the PZ subgroup compared to the other RAI subgroups (pā€‰>ā€‰0.5). This observation aligns with the aforementioned study, which reported noticeable warpage at the apical part of the RAI during the manufacturing process specifically in the PZ group [34]. Such warpage may contribute to increased susceptibility of the RAI to displacement or movement under loading conditions.

Regarding total displacement, the RAI group exhibited a statistically significant increase in displacement compared to the TI group (pā€‰<ā€‰0.05). Specifically, the RAI demonstrated increased micromotion along the X and Y axes, while the TI group primarily experienced micromotion along the loading Z axis. This eccentric micromotion behaviour of the RAI can be attributed to its anatomical shape, characterised by asymmetric mesial and distal roots in terms of form and length. Additionally, the experimental loading conditions of the HexMeS setup, where the specimens are indirectly loaded through the spoon-shaped attachment, indirectly contribute to this behaviour. From a biomechanical perspective, the spoon-shaped attachment acts as a lever arm, generating torque on the implant and resulting in rotation around the Y axis, thereby increasing displacement along the X axis.

Noteworthy, the mean micromotion values observed in the RAI were approximately 48Ā Āµm for displacement along the Z-axis and approximately 96Ā Āµm for total displacement. These values remain below the maximum threshold value of micromotion crucial for successful osseointegration, which is estimated to be around 150Ā Āµm [26, 27]. These findings indicate that the range of micromotion exhibited by the RAI is unlikely to impede the osseointegration process.

We presume that this study has effectively investigated the stability of RAI by closely adhering to the clinical workflow and utilizing the latest technologies in RAI preparation. Nevertheless, certain limitations should be considered. Firstly, this is an in vitro study, thus the histological examination of osseointegration was not feasible. Secondly, the fixation of RAI in the artificial bone involved the use of a thin resin layer, which was a crucial step owing to the anatomical shape of RAI and the non-drilling surgical protocol. Finally, the loading of the HexMeS setup could not be done directly on the specimens, and the spoon-shaped load applicator might have introduced an additional torque.

In view of the results of the present study, RAIs showed biomechanical behaviour in terms of stability and micromotion comparable to that of TIs. Moreover, based on previous studies [14, 34], RAIs fabricated by milling or 3D printing showed promising results in terms of dimensional accuracy. Collectively, these findings propose that RAI could serve as a feasible alternative to TI, particularly in immediate implant cases, provided a well-prepared preoperative treatment plan and access to a capable CBCT device. However, it's crucial to note that the tooth to be replaced should lack sharp undercuts that might impede the insertion of the RAI or compromise its proper fit. Nevertheless, further clinical trials and studies are necessary to validate its clinical application.

Conclusions

After acknowledging the limitations of this study, we drew the following conclusions:

  1. 1.

    The RAI exhibited promising biomechanical behaviour, as indicated by micromotion values within physiological limits.

  2. 2.

    The stability of the RAI could be influenced by the manufacturing technique.

  3. 3.

    Compared to the TI, the biomechanical behaviour of the RAI is less predictable due to its irregular anatomical design.

  4. 4.

    Precise definition of the implant geometry is essential to ensure a precise fit and a seamless insertion.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

Notes

  1. R Core Team (2022). R: A language and environment for statistical computing. R Foundation for Statistical Computing, Vienna, Austria. URL https://www.R-project.org/

Abbreviations

TI:

Threaded implant

RAI:

Root analogue implant

CBCT:

Cone beam computed tomography

ITI:

International team for implantology

CAD/CAM:

Computer-aided design/Computer-aided manufacturing

CNC:

Computer numerical control

LCM:

Lithography-based ceramic manufacturing

SLM:

Selective laser melting

HexMeS:

Hexapod measurement system

CSV:

Comma separated values

References

  1. Moraschini V, Poubel LADC, Ferreira VF, Barboza EDSP. Evaluation of survival and success rates of dental implants reported in longitudinal studies with a follow-up period of at least 10 years: A systematic review. Int J Oral Maxillofac Surg. 2015;44(3):377ā€“88.Ā https://www.sciencedirect.com/science/article/pii/S0901502714004251

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  2. Gallucci GO, Hamilton A, Zhou W, Buser D, Chen S. Implant placement and loading protocols in partially edentulous patients: A systematic reviewĀ  Clin Oral Implants ResĀ Blackwell Munksgaard. 2018;29:106ā€“34.

    ArticleĀ  Google ScholarĀ 

  3. Beagle JR. The Immediate Placement of Endosseous Dental Implants in Fresh Extraction Sites. Dent Clin North Am. 2006;50(3):375ā€“89.Ā https://www.sciencedirect.com/science/article/pii/S0011853206000346

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  4. Bhola M, Neely AL, Kolhatkar S. Immediate implant placement: Clinical decisions, advantages, and disadvantages. J Prosthodont. 2008;17(7):576ā€“81. https://doi.org/10.1111/j.1532-849X.2008.00359.x.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  5. Koh RU, Rudek I, Wang HL. Immediate implant placement: Positives and negatives. Implant Dent. 2010;19(2):98ā€“108.Ā Ā https://pubmed.ncbi.nlm.nih.gov/20386212/. Cited 2021 Nov 10

  6. Regish KM, Sharma D, Prithviraj DR. An overview of immediate root analogue zirconia implants. J Oral Implantology. 2013;39:225ā€“33.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  7. Saeidi Pour R, Freitas Rafael C, Engler MLPD, Edelhoff D, Klaus G, Prandtner O, et al. Historical development of root analogue implants: a review of published papers. Br J Oral Maxillofac Surg. 2019;57(6):496ā€“504. https://doi.org/10.1016/j.bjoms.2019.01.021.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  8. Yong LT. Single stage immediate implant placements in the esthetic zone. J Oral Implantol. 2012;38(6):738ā€“46.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  9. Figliuzzi M, Mangano F, Mangano C. A novel root analogue dental implant using CT scan and CAD/CAM: Selective laser melting technology. Int J Oral Maxillofac Surg. 2012;41(7):858ā€“62. https://doi.org/10.1016/j.ijom.2012.01.014

  10. Hodosh M, Povar M, Shklar G. The dental polymer implant concept. J Prosthet Dent. 1969;22(3):371ā€“80.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  11. Ghuneim WA. In situ tooth replica custom implant: A 3-dimensional finite element stress and strain analysis. J Oral Implantol. 2013;39(5):559ā€“73.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  12. Moin D, Hassan B, Wismeijer D. Immediate Nonsubmerged Custom Root Analog Implants: A Prospective Pilot Clinical Study. Int J Oral Maxillofac Implants. 2018;33(2):e37-44.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  13. Anssari Moin D, Hassan B, Wismeijer D. A novel approach for custom three-dimensional printing of a zirconia root analogue implant by digital light processing. Clin Oral Implants Res. 2017;28(6):668ā€“70.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  14. Chen J, Zhang Z, Chen X, Zhang C, Zhang G, Xu Z. Design and manufacture of customized dental implants by using reverse engineering and selective laser melting technology. J Prosthet Dent. 2014;112(5):1088ā€“1095.E1. https://doi.org/10.1016/j.prosdent.2014.04.026

  15. Mangano FG, De Franco M, Caprioglio A, MacChi A, Piattelli A, Mangano C. Immediate, non-submerged, root-analogue direct laser metal sintering (DLMS) implants: A 1-year prospective study on 15 patients. Lasers Med Sci. 2014;29(4):1321ā€“8.

    PubMedĀ  Google ScholarĀ 

  16. Dantas T, Madeira S, Gasik M, Vaz P, Silva F. Customized root-analogue implants: A review on outcomes from clinical trials and case reports. Materials (Basel). 2021;14(9):1ā€“14.

    ArticleĀ  Google ScholarĀ 

  17. Beuer F, Schweiger J, Edelhoff D. Digital dentistry: An overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008;204(9):505ā€“11.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  18. Kanazawa M, Inokoshi M, Minakuchi S, Ohbayashi N. Trial of a CAD/CAM system for fabricating complete dentures. Dent Mater J. 2011;30(1):93ā€“6.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  19. Etemad-Shahidi Y, Qallandar OB, Evenden J, Alifui-Segbaya F, Ahmed KE. Accuracy of 3-dimensionally printed full-arch dental models: A systematic review. J Clinical Med MDPI. 2020;9:1ā€“18.

    Google ScholarĀ 

  20. Revilla-LeĆ³n M, Ɩzcan M. Additive Manufacturing Technologies Used for Processing Polymers: Current Status and Potential Application in Prosthetic Dentistry. J Prosthodont. 2019;28(2):146ā€“58.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  21. Griffith ML, Halloran JW. Freeform fabrication of ceramics via stereolithography. J Am Ceram Soc. 1996;79(10):2601ā€“8. https://doi.org/10.1111/j.1151-2916.1996.tb09022.x.

    ArticleĀ  Google ScholarĀ 

  22. Moin DA, Hassan B, Mercelis P, Wismeijer D. Designing a novel dental root analogue implant using cone beam computed tomography and CAD/CAM technology. Clin Oral Implants Res. 2013;24(A100):25ā€“7.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  23. Revilla-LeĆ³n M, Sadeghpour M, Ɩzcan M. A Review of the Applications of Additive Manufacturing Technologies Used to Fabricate Metals in Implant Dentistry. J Prosthodont. 2020;29(7):579ā€“93.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  24. Vayron R, Nguyen VH, Lecuelle B, Lomami HA, Meningaud JP, Bosc R, et al. Comparison of resonance frequency analysis and of quantitative ultrasound to assess dental implant osseointegration. Sensors (Switzerland). 2018;18(5). Available from: www.mdpi.com/journal/sensors. Cited 2022 Oct 24.

  25. Laney. Glossary of Oral and Maxillofacial Implants. Int J Oral Maxillofac Implants. 2017;32(4):Gi-G200.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  26. Winter W, Klein D, Karl M. Micromotion of Dental Implants: Basic Mechanical Considerations. J Med Eng. 2013;2013:1ā€“9.

    ArticleĀ  Google ScholarĀ 

  27. Szmukler-Moncler S, Piattelli A, Favero GA, Dubruille JH. Considerations preliminary to the application of early and immediate loading protocols in dental implantology. Clin Oral Implants Res. 2000;11(1):12ā€“25.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  28. Miri R, Shirzadeh A, Kermani H, Khajavi A. Relationship and changes of primary and secondary stability in dental implants: A review. Int J Contemp Dent Med Rev. 2017;6. https://ijcdmr.com/index.php/ijcdmr/article/viewFile/258/232

  29. Ivanova V, Chenchev I, Zlatev S, Mijiritsky E. Correlation between primary, secondary stability, bone density, percentage of vital bone formation and implant size. Int J Environ Res Public Health. 2021;18(13):6994.

  30. Tobar-Reyes J, Andueza-Castro L, JimĆ©nez-Silva A, Bustamante-Plaza R, Carvajal-Herrera J. Micromotion analysis of immediately loaded implants with Titanium and Cobalt-Chrome superstructures. 3D finite element analysis. Clin Exp Dent Res. 2021;7(4):581ā€“90.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  31. Wilhelm R, Hasan I, Keilig L, Heinemann F, Stark H, Bourauel C. Biomechanical investigations of the secondary stability of commercial short dental implants in porcine ribs. Biomed Tech. 2014;59(6):507ā€“13.

    ArticleĀ  Google ScholarĀ 

  32. Keilig L, Bourauel C, GrĆ¼ner M, HĆ¼ltenschmidt R, Bayer S, Utz KH, et al. Aufbau und Erprobung eines neuartigen MeƟsystems fĆ¼r die dentale Biomechanik - MeƟprinzip und Beispielmessungen des Hexapod-MeƟ-Systems. Biomed Tech. 2004;49(7ā€“8):208ā€“15.

    ArticleĀ  Google ScholarĀ 

  33. Hasan I, Heinemann F, Schwegmann M, Keilig L, Stark H, Bourauel C. Experimental investigation of commercial small diameter dental implants in porcine mandibular segments. Biomed Tech. 2017;62(1):103ā€“8.

    ArticleĀ  Google ScholarĀ 

  34. Aldesoki M, Keilig L, Dƶrsam I, Evers-Dietze B, Elshazly TM, Bourauel C. Trueness and precision of milled and 3D printed root-analogue implants: A comparative in vitro study. J Dent. 2023;130:104425(December 2022). https://doi.org/10.1016/j.jdent.2023.104425.

  35. Tribst JPM, Dal Piva AM de O, Blom EJ, Kleverlaan CJ, Feilzer AJ. Dental biomechanics of root-analog implants in different bone types. J Prosthet Dent. 2022;1ā€“11. https://doi.org/10.1016/j.prosdent.2022.10.005

  36. Niinomi M. Biologically and mechanically biocompatible titanium alloys. Mater Trans. 2008;49(10):2170ā€“8.

    ArticleĀ  Google ScholarĀ 

  37. Van Oers RFM, Feilzer AJ. Abutment-to-fixture load transfer and peri-implant bone stress. Am J Dent. 2015;28(5):247ā€“50.

    PubMedĀ  Google ScholarĀ 

  38. Gattinger J, Bullemer CN, Harrysson OLA. Patient specific root-analogue dental implants ā€“ Additive manufacturing and finite element analysis. Curr Dir Biomed Eng. 2016;2(1):101ā€“4.

    ArticleĀ  Google ScholarĀ 

  39. Chen J, Zhang Z, Chen X, Zhang X. Influence of custom-made implant designs on the biomechanical performance for the case of immediate post-extraction placement in the maxillary esthetic zone: a finite element analysis. Comput Methods Biomech Biomed Engin. 2017;20(6):636ā€“44. https://doi.org/10.1080/10255842.2017.1283406.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

Download references

Acknowledgements

The authors gratefully acknowledge Dr. Bassam A. Abulnoor for his contribution to the statistical analysis of this research.

Funding

Open Access funding enabled and organized by Projekt DEAL. This work was supported by a grant from the German academic exchange service (DAAD, 2019/20, 57440921).

Author information

Authors and Affiliations

Authors

Contributions

MA: Conceptualization, Methodology, Software, Formal analysis, Data curation, Writing, Original draft preparation, Visualisation. CB: Project administration, Validation, Writing, Reviewing and Editing, Supervision. TME: Investigation, Data curation, Writing, Reviewing and Editing. ES: 3D printing, Reviewing and Editing, Visualisation. LK: Methodology, Investigation, Data curation, Software, Writing, Reviewing and Editing, Supervision. All authors have read and agreed to the published version of the manuscript.

Corresponding author

Correspondence to Mostafa Aldesoki.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Aldesoki, M., Bourauel, C., Elshazly, T.M. et al. Evaluation of micromotion in multirooted root analogue implants embedded in synthetic bone blocks: an in vitro study. BMC Oral Health 24, 99 (2024). https://doi.org/10.1186/s12903-024-03854-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12903-024-03854-1

Keywords